Ingenia 1.5T S 781347Recalls & reported problems

Philips Medical Systems Nederland B.V. · Magnetic Resonance Equipment · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

6 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
7

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838068421), received between Dec 3, 2020 and Dec 3, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

7 FDA recall records name this device's identifier, 6 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 28, 2026Class IIOpen, Classified

    When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

    Z-3308-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Aug 21, 2026Class IIOpen, Classified

    A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

    Z-3236-2026 · Root cause: Software design · Recall details · FDA record ↗

  3. Jun 30, 2025Class IIOpen, Classified

    The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

    Z-2205-2025 · Root cause: Device Design · Recall details · FDA record ↗

  4. May 13, 2025Class IIOpen, Classified

    Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.

    Z-1796-2025 · Root cause: Software design · Recall details · FDA record ↗

  5. May 3, 2024Class IIOpen, Classified

    The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

    Z-1847-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  6. Mar 12, 2024Class IIOpen, Classified

    Patient support table floor plate may be incorrectly installed.

    Z-1565-2024 · Root cause: Process design · Recall details · FDA record ↗

  7. Aug 11, 2021Class IICompleted

    When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment

    Z-2413-2021 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1

Patient problems

  • Partial thickness (Second Degree) Burn1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 3, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    PHILIPS RECEIVED A REPORT ON A HEATING INCIDENT ON AN INGENIA 1.5T MR SYSTEM. PATIENT SUFFERED A LARGE 2ND DEGREE BURN TO THE RIGHT BUTTOCK/HIP.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 10942035 ↗

Other System, nuclear magnetic resonance imaging devices

All System, nuclear magnetic resonance imaging devices →