uEXPLORER PET/CT System - Product Usage: intended to be… recallZ-2594-2021 · Shanghai United Imaging Healthcare Co., Ltd. · initiated Aug 2, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Shanghai United Imaging Healthcare Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The EXAM acquisition workflow could occasionally freeze after canceling the PET scan.
FDA root cause category: Device Design
What the company is doing
On 08/02/2021, the Firm emailed to its customer an "Advisory Notification" informing them of a potential safety-relevant issue in which the EXAM acquisition workflow could occasionally freeze after canceling the PET scan. The issue and the sequence of events leading to this occurrence are as follow: During a myocardial perfusion scan procedure, the follow steps could lead to the EXAM acquisition workflow to occasionally freeze- 1. Selecting the myocardial perfusion scan procedure and entering EXAM. 2. Finishing the CT scout and helical/cine scan. 3. Starting the PET scan. 4. Canceling the current PET scan -- in this case the EXAM acquisition workflow may shut down, resulting in an error message stating: "EXAM (Study ID: XXXXXXX) has some errors, trying to restart. Please wait and try again. 5. Reentering the EXAM acquisition workflow without restarting the system, and then attempting to continue the scan. If the above occurs, this may require rescanning of the patient, resulting in additional CT dose exposure. Customer are instructed to: 1. When the EXAM acquisition workflow is interrupted after cancellation of the scan, follow the steps to resume the procedure: a) Wait until CT reconstruction finishes. b) Perform a system restart. c) After restart, select the suspended procedure on the patient administration (PA) interface and enter EXAM. System will load and recover the status of this procedure. d) Duplicate or copy and paste the unfinished PET scan, and then continue. 2. Please notify all internal users of the PET/CT system, using this letter. 3. Complete and sign the Receipt of Advisory Notification Form and return it to the Recalling Firm via fax, email or mail. The Recalling Firm is working on releasing a software update to resolve this issue. Planned release of software update is Q4 of 2021. For questions, contact the Recalling Firm's service engineer or call the customer contact center's 24 hour hotline at (855)221-1552 (toll free) or (832)699-2799 (d
Recall details
- Recall number
- Z-2594-2021
- Status
- Open, Classified
- Initiated by the company
- Aug 2, 2021
- Recalling firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Quantity
- 1 system
- Product code
- KPS
- FDA submissions
- K182237, K193241
- Distribution
- U.S. Nationwide distribution in the states of CA, IL, IN, TX, and WI.
Product description
uEXPLORER PET/CT System - Product Usage: intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Code information (models, lots, serial numbers)
Model: 88000025 GTIN/UDI Code: 06971576832040 / 01 06971576832040 240 88000025 21 F00002 Serial Numbers: F00002
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- uEXPLORERShanghai United Imaging Healthcare Co., Ltd. · System, X-Ray, Tomography, Computed3 recalls0 reports
Other devices under the same FDA submission
2 devices cleared under K182237, K193241 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- uMI uMI 550Shanghai United Imaging Healthcare Co., Ltd. · System, Tomography, Computed, Emission6 recalls0 reports
- uMI uMI 550Shanghai United Imaging Healthcare Co., Ltd. · System, Tomography, Computed, Emission2 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 88521).
- Z-2593-2021 · Open, ClassifieduMI 550 System - Product Usage: intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment pla…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page