Centricity recallZ-2503-2021 · GE Healthcare, LLC · initiated Aug 6, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Image acquisition failures and synchronization failure with the Centricity Enterprise Archive

FDA root cause category: Software design

What the company is doing

This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected software versions on August 06, 2021. This letter informs customers of the issues and provides safety instructions. GE Healthcare will correct the issue through a product change. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-2503-2021
Status
Open, Classified
Initiated by the company
Aug 6, 2021
Posted by the FDA
Sep 22, 2021
Recalling firm
GE Healthcare, LLC
Quantity
462 (279 US; 183 OUS)
Product code
LLZ
FDA submissions
K150420
Distribution
AK, AZ, CA, CO, DE, FL, GA, IA, IN, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TX, UT, WA, WI, WV and WY. Austria, Brazil, Canada, France, Germany, Ghana, India, Ireland, Israel, Italy, Kuwait, Netherlands, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, Turkey, United Kingdom and Uruguay.

Product description

Centricity Universal Viewer and Viewer 6.0 - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources. Includes the following Model Numbers: 1. Model Number: 2090255-001; 2. Model Number: 2088026-723; 3. Model Number: 2102676-001; 4. Model Number: 2104867-044.

Code information (models, lots, serial numbers)

Show the full code information (8k characters)

1. Model Number: 2090255-001 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1) System ID (UDI-DI if applicable): PMIO14194768, PMIO20084199, PACS-IW-231, CITRUSVAL20884363, CITRUSVAL20884364, CITRUSVAL20884365, PACS-IW-155, PRIMEHEAL02315681, PRIMEHEAL20501355, PARKVIEWC21200827 (840682103800), PARKVIEWC21200830 (840682103800), PACS-IWUVN2-441, QUEEN19630914, QUEEN19630915, QUEEN19630916T, QUEEN20553950, 203268UVSW, DELAWAREC01589220, PACS-IW-462UV, PACS-IW-462UVN1, PACS-IW-126, PACS-IW-1062, OFMC01819402, OFMC20791988, HGS01766912, HGS20584004 (840682103800), GA001PACTRL01, TRGI02180958, DRY01810490, DRY19670358, PACS-IW-101, RML20502096, PACS-IWUV-239, PACS-IW-00813, PHS01374810, 01722652SEAS, 02031921SEASN1, MARYLANDF17985489, MYRTLEHIL17713895, MYRTLEHIL01444483, MYRTLEHIL17713895, VT01766830, VT01977629, PACS-IW-2637, SRMC01982629, PACS-IWUV-568, 1-142022244, PACS-IWUV-1131, OGAA02373799NOD1, ALLG00558571 (840682103800), EXC02189547, EXC01584099, PACS-IW-574, PRIME HEAL19508102, PRIME HEAL19050922, PRIME HEAL19050921, PACS-IW-219, ANCL00558570, ANCL02157634, NSW01452088, NSW20080545, MARYLANDF17985489, PACS-IW-518UV (840682103800), MTGD02393053, MTGD00558454, PACS-IW-44, QUENTIN01970717, LAKESIDEM02256808, BMC12P00558644P1, BMC12T02267839, BMC12N01271417N1, PACS-IW-58, SMCO02039946PC (840682103800), BWH01301357, PACS-IW-349, PACS-IW-255, 1-357059471, CAROCOMMU02205222, CAROCOMMU01459013, CSPA01291589 (840682103800), CSPA02097167 (840682103800), PACS-IW-79, PACS-IW-93, 1-276423210, MFCI01448329, LSFH02032334, CORONAREG01300927, 951737IWUWWN1, CORONAREG02288447, 1-382753509, CITRUSVAL19618406, MGMC02288391, MGMC02398531, 1-473998727, BARRETTHO00558704, BARRETTHO17737529, FORTDUNCA01298556 (840682103800), FORTDUNCA02395799 (840682103800), 1-283597082, TTU01973509, AVHL02396030, PACS-IW-511, PACSIWTEST, PACS-IW-386UV (840682103800), PACS-IW-486-UV (840682103800), PACS-IW-486-UVN1 (840682103800), PACS-IW-390 (840682103800), PACS-IW-202TEST, PACS-IW-202, HZC14961014, AVHL02396026, AVHL02396027, AVHL02396028, AVHL02396029, PACS-IW-419, PMRM14768587UVProxy, HANNIBALR02179353 (840682103800), TESTHANNIBALR02255613 (840682103800), HANN16174584UVProxy (840682103800), PACS-IW-189 (840682103800), HARTSDALE14768195 (840682103800), HOA01451973 (840682103800), RML01448365 (840682103800), ICA01367198 (840682103800), ICA14776604 (840682103800), JKS01452133 (840682103800), JKS14776447UVSFNode (840682103800), JKS14776448UVProxy (840682103800), JKS14776449UVTest (840682103800), PACS-IW-181, GTEC01477448, GTEC01477447, GTEC15993728UVProxy3, 1-592001096 (840682103800), UHSD14767800 (840682103800), PACS-IW-551UVN1 (840682103800), PACS-IW-551UVN2 (840682103800), UV1-154550926, DHOL19603374, DHOL19603375, PACS-IW-551UV (840682103800), PACS-IW-424CTLR, PACS-IW-424UVN1, PACS-IW-424UVN2, SFCS19667347, SFCS15961473, SFCS15961473, MHOG01806096, PACS-IW-166, PACS-IW-166T, MCT02277843, MCT02278145, PACS-IW-565, PACS-IWT-565, WGH00558334CTRL, WGH02372377PROX, WGH02372375BACK, 1-464783615, PACS-IW-389, BVH00558643, PACS-IW-348UV (840682103800), NRVM19609993UVT (840682103800), NRVM19611189UVPROXY (840682103800), PACS-IW-441, PACS-IW-441TUVSW, PACS-IW-402, PACS-IW-318, MINNETOHE, PACS-IW-253, PACS-IW-581UV, PACS-IW-370, PACS-IW-UVT370, VISOU02345023, VISOU02288781, 1-218291079, 1-103108035, MFHO14767332, PALMDALEM00558700, PALMDALEM01810788, PACS-IW-454, PACS-IW-406, 20101378UBSW, ILCC01373579 (840682103800), ILCC14580319 (840682103800), 516473UVSW (840682103800), JOHNTMA19732280T (840682103800), 1-182669905, NCDC20279342, CENTERFOR01627896, CENTERFOR16023960UVProxy, PACS-IW-176, PACS-IW-176UVTEST, PACS-IW-405, PACS-IW-336, NASONHOSP14898835, RRA01814001, MYRTLEHIL01444483, PSPC01815255, 1-283596904, HOOO20091140, GZMG02182469CNTRLR (840682103800), GZMG02287985 (840682103800), GZMG02288300 (840682103800), PACS-IW-556UV, URSA20025013, PACS-IW-431, 1-598891598, PACS-IW-230, INTERCOAST17594527UVProxy, MDHC16204735UVProxy, 201392UVSW, 505368UVBSW (840682103800), NNAR14767799 (840682103800), PACS-IW-479, PACS-IW-557 (840682103800), URS20583582 (840682103800), PACS-IWUV-481, PACS-IW-476, PACSIW476UVTEST, GGMC02349137, 1-486091448, 1-486091475, 1-750710672, SUC00558667, 1-802428413, 954688UVBSW100, PACS-IW-1511, KDH217192544, ACP00558109, PACS-IW-31Test1, PACS-IW-31Test, ACP01613226, 1-219095008, OKOS14767330, 918277CVUWBSW, CCA20759776SF, PACS-IW-487, QTHBL14767810UVPROXY, QTHBL14767809UVSFBK, QTHBL14767808UVSF1KO, QTHBL14767807UVSFSIM, QTHBL14767806UVSFVAL, QTHBL02256595, VCM02255583T, VCM01819618, VCM02255586TSP, VMC02255587NSP, BOJC00558520, PACS-IW-56, PACS-IW-225, 1-375705893, DDNI14776603, PACS-IW-129, DISSFHW01, 1-644600677, PACS-IW-154UV, FLMC14777071, NEWYORK01291738, OANT02035061, PACS-IW-424CTLR, PACS-IW-424UVN1, PACS-IW-424UVN2, VBHCP20497227, PACS-IW-2635, PREMI02182462, PREMI02182462, NORTHWEST01804127, HCIT4271649UW, AH4325SV01, AH5291PC02, AP6264PC01, AP6469PC01, AH0116SV03, AH0106SV03, HCD4440494UV, HCIT4258559UV, HCIT7109403UV, HCIT4244609UV, HCIT4708563UV, HCIT4470563UV-CTRL, HCIT4261024UVCTRL, HCIT7113573UV, 519264PACSAW1 (840682103800), 416813UV (840682103800), 416813TESTRPACSUV (840682103800), HCIT4266470UW, HCIT4238540UV, HCIT4486723UV, E4167658, X314119802, X141385617, A5485262 / A5485266 / A5485267 / A5485269 / A5485270 / A5485271, A9190177, M41980107, M4487535, M44997011, HC0399UV01, CC4105PC02, PC1695SV01, 786787SV02, HBON01SV02, HC4105PC01, 509940UV01, HC0529PC01, HC0529PC01, 233638UV01, HC0245PC02, 600791PC02, 427990UV01, PC1961SV02, 531344PC01, 434487SV03, PX4900SV02, HC0565PC03, TP0575UV02, GH1077UV01, 083055820116113, 083055824999515, 083055861021012, 83055862088813, 083055800046112, 81111DBS01, IL1469IT05, IL1469IT16, A5992711, A5412010, A6197102, KW0000PP02, HCIT4175024UV, AMS3963, BON0164, ECT0265, HCIT4192962UV, HCIT4264206UW, PL3673RW01, PL1493RW02, PL2970RW01, PL1124RW01, PL1499RW03, PL3349PC01, PL1501RW03, PL2624RW01, PL3249RW02, PL3277RW01, PL1005RW01, 600062UV, PL3362RW01, PL0326RW01, PL2975RW02, PL1504RW02, PL2971RW01, RU9255ID01, RU6968ID01, RU8945ID01, RU7519ID01, RU1781RW01, RU2580ID01, RU4255ID01, RU0003ID01, RU0472ID03, RU9595ID02, RU9327ID02, RU1011ID05, RU8366ID01, RU7808ID02, RU1516ID03, RU1646ID01, RU1112ID02, 01977813, 01980259, RU3052ID01, RU1107RW01, RU1107RW01, RU7175ID01, RU1717ID04, RU9216ID02, RU3453RW03, 600101RW02, RU8002ID01, RU1544ID01, RU1516ID03, RU9940RW01, RU5388ID01, RU0994ID03, RU3794ID01, RU9226ID01, RU9697ID01, 1509505, RU0968ID04, RU0120ID01, RU6164ID01, RU9546ID01, RU7468ID01, RU3998ID01, 600105ID03, SA2402PP02, SA2110PP05, SA1339PP03, ZA2541UV01, ZA2414PAC1, ZA2415PAC1, ZA2474UV01, ZA2267PAC1, ZA1058UV01, ZA2551PAC1, ZA1058UV01, ZA2412PAC2, ZA2472PAC2, ZA000010, ZA2412PAC2, MPX35219, DP760316, FP502804, 802030PC01, 228493WS01, 124774CS01, 482896SV03, 06051PAC04, HCIT4133734PACS, HCIT4148106PEE, Others Tracked under Service Cloud product ID. 2. Model Number: 2088026-723 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: NYHOSPI16204605, NYHOSPI16204606, NYHOSPI16204607, NYHOSPI16204608, NYHOSPI16204609, NYHOSPI16204612, NYHOSPI16204613. 3. Model Number: 2102676-001 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: PACS-IW-119. 4. Model Number: 2104867-044 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: DDOC02339896.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page