Philips Ultrasound, Inc. Sparq, Diagnostic Ultrasound… recallZ-0071-2022 · Philips Ultrasound Inc · initiated Aug 19, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Ultrasound Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Battery system data issue with the ultrasound system can intermittently cause a system shutdown, regardless of actual battery state or application of alternating current power.
FDA root cause category: Software design
What the company is doing
On August 24, 2021, Philips issued "Urgent Medical device Correction" notification to affected consignees. The customer communication was distributed via: USPS Certified Mail In addition to informing consignees about the recall, Philips requested consignees take the following actions: Instructions for how to determine the software version of your Ultrasound system: 1. Power up the system and allow it to complete the boot sequence, 2. Press the Wrench icon on the upper right side of the control panel, 3. Press the Service button on the bottom left corner, 4. Under System Management, Press System Information, 5. The software version is listed in the Software Information Section. 4. Describe the actions that should be taken by the customer / user in order to prevent risks for patients or users A. If this shutdown occurs, Philips recommends that the system be put back on AC power and restarted. The system will be fully functional upon restart. However, restarting the system will not prevent the possibility of the issue reoccurring. B. Please complete and return the attached form to Philips promptly and no later than 30 days from receipt (via email: [email protected], or fax: 1-833-512-7756). 5. Describe the actions planned by Philips Ultrasound to correct the problem A Philips Field Service Engineer will contact you to schedule a software update to permanently resolve the issue (reference FCO79500550). 6. This notice has been reported to the appropriate Regulatory Agencies. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax. 7. Please be assured that maintaining a high level of safety and quality is our highest priority. If you need any further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377 and reference FCO79500550.
Recall details
- Recall number
- Z-0071-2022
- Status
- Terminated
- Initiated by the company
- Aug 19, 2021
- Recalling firm
- Philips Ultrasound Inc
- Quantity
- 1 unit
- Product code
- IYN
- FDA submissions
- K201012
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, MA, MD, MI, MS, NC, NY, OH, VA, WA and the countries of Australia, Portugal, Netherlands, Canada, Netherlands, Germany.
Product description
Philips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795116, 100-240V -, 50/60 Hz, 500 VA, CE2797, Rx Only,
Code information (models, lots, serial numbers)
Affected Serial Number/UDI numbers: US92010594 (01)00884838098015(21)US92010594
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Sparq Diamond Select Ultrasound System 795116Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 88625).
- Z-0070-2022 · TerminatedPhilips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795090, 100-240V -, 50/60 Hz, 500 VA, CE2797, Rx Only,
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page