Alinity m recallZ-0007-2022 · Abbott Molecular, Inc. · initiated Sep 2, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Molecular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for false positive results.
FDA root cause category: Device Design
What the company is doing
The firm issued an Urgent Field Safety Notice/Field Correction Recall 09/02/2021 to all impacted customers. Per the customer letter, all positives are to be considered presumptive until the Alinity m SARS-CoV-2 Application Specification file version 5 or the updated Alinity m Resp-4-Plex Application Specification file is implemented at the customer site. On 03/30/2022, Abbott Molecular issued a follow-up Urgent Field Safety Notice/Field Correction Recall letter to all customers who ordered Alinity m SARS-CoV-2 AMP Kit Emergency Use Authorization (EUA), List Number 09N78-095 and/or Alinity m Resp-4-Plex AMP Kit EUA, List Number 09N79-096 used with the Alinity m System. Also to customers who ordered Alinity m SARS-CoV-2 AMP Kit (CE), List Number 09N78-090 or 09N78-091 and/or Alinity m Resp-4-Plex AMP Kit (CE), List Number 09N79-090 used with the Alinity m System. The Customer Letter explains Abbott has updated the existing Application Specification files for Alinity m SARS-CoV-2 and Resp-4-Plex to further reduce the potential for carryover. FDA has not yet authorized the Alinity m SARS-CoV-2 Application Specification file version 6 update or the Alinity m Resp-4-Plex Application Specification file version 5 update. Implementation of the modification for Alinity m SARS-CoV-2 and Resp-4-Plex will occur when the update to the Application Specification file has been authorized. A Molecular Diagnostics Abbott Representative will be contacting you once the updated Application Specification files are available. If you have any questions, please contact your local Molecular Diagnostics Abbott representative or call Abbott Technical Support at 1-800-553-7042 Option 2.
Recall details
- Recall number
- Z-0007-2022
- Status
- Terminated
- Initiated by the company
- Sep 2, 2021
- Posted by the FDA
- Oct 14, 2021
- Recalling firm
- Abbott Molecular, Inc.
- Quantity
- 11408 units
- Product code
- QJR
- Distribution
- Worldwide Distribution. US Nationwide, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, COLOMBIA, CZECH REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, HONG KONG, IRELAND, ISRAEL, ITALY, JAPAN, MALAYSIA, MONTENEGRO, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, UNITED KINGDOM, and VIETNAM.
Product description
Alinity m SARS-CoV-2 Amp Kit (Non-US Assay) and Alinity m SARS-CoV-2 Application Specification File. in vitro diagnostic
Code information (models, lots, serial numbers)
List Number: 09N78-091, UDI 00884999049222; Lot/Serial, Expiration: 520596, 06/18/2022; 520952, 07/09/2022; 524054, 10/18/2022; 524965, 10/25/2022; 525462, 12/10/2022; 525768, 12/16/2022; 526708, 01/18/2023; 527842, 02/08/2023; List Number 09N78-03E, UDI (01)00884999050211(240)09N78-03E(8012)5.oo
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Alinity m 09N78-095Abbott Molecular Inc. · Reagents, 2019-novel coronavirus nucleic acid3 recalls357 reports
- Alinity m 09N78-03EAbbott Molecular Inc. · Reagents, 2019-novel coronavirus nucleic acid3 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 88626).
- Z-0004-2022 · TerminatedAlinity m SARS-CoV-2 Amp Kit (Emergency Use Authorization, US Assay) and Alinity m SARS-CoV-2 Application Specification File. in vitro diagnostic
- Z-0005-2022 · TerminatedAlinity m Resp-4-Plex AMP Kit (Emergency Use Authorization, US Assay) and Alinity m Resp-4-Plex Application Specification File. in vitro diagnostic
- Z-0006-2022 · TerminatedAlinity m SARS-CoV-2 Amp Kit (Non-US Assay) and Alinity m SARS-CoV-2 Application Specification File. in vitro diagnostic
- Z-0008-2022 · TerminatedAlinity m Resp-4-Plex AMP Kit (Non-US Assay) and Alinity m Resp-4-Plex Application Specification File. in vitro diagnostic
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page