Skip to content

Lisendo 880 Ultrasound recallZ-0191-2022 · FujiFilm Healthcare Americas Corporation · initiated Oct 14, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from FujiFilm Healthcare Americas Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.

FDA root cause category: Software design

What the company is doing

Fujifi Film issued Medical Device Correction Notice letter dated 10/07/21 via courier on 10/14/21. Letter states reason for recall, health risk and action to take: Customers/Users are advised to accept the corrected software and manual as soon as available. Advisory Notification applies to all customers and distributors of the ARIETTA 850 systems with software version V1.0.0 through V4.1.3, See following page for complete list of affected systems. We will contact you to set up time to update the software and provide the revised manual. (7) Other Information: For questions, comments, or concerns, please contact: Aaron Pierce Director, QA/RA 1959 Summit Commerce Park Twinsburg, OH 44087 Phone (330) 425-1313 x2934, Fax (330) 963-0749 Email [email protected]

Recall details

Recall number
Z-0191-2022
Status
Open, Classified
Initiated by the company
Oct 14, 2021
Recalling firm
FujiFilm Healthcare Americas Corporation
Product code
IYN
FDA submissions
K162583
Distribution
Nationwide

Product description

Lisendo 880 Ultrasound Software Version: V1.0 through V4.1.3

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

Serial Numbers/UDI: 20514657 (01)04562122958868(11)161220(21)20514657; 20659147 (01)04562122958868(11)20181124(21)20659147; 20659148 (01)04562122958868(11)20181124(21)20659148; 20659149 (01)04562122958868(11)20181124(21)20659149; 20659150 (01)04562122958868(11)20181124(21)20659150; 20659153 (01)04562122958868(11)20181126(21)20659153; 20659154 (01)04562122958868(11)20181126(21)20659154; 20659155 (01)04562122958868(11)20181126(21)20659155; 20659156 (01)04562122958868(11)20181126(21)20659156; 205G0225 (01)04562122958868(11)20170418(21)205G0225; 205G0226 (01)04562122958868(11)20170424(21)205G0226; 205P2146 (01)04562122958868(11)20170615(21)205P2146; 205P2147 (01)04562122958868(11)20170615(21)205P2147; 205P2148 (01)04562122958868(11)20170615(21)205P2148; 205P2149 (01)04562122958868(11)20170615(21)205P2149; 205Q0809 (01)04562122958868(11)20170707(21)205Q0809; 205Q0810 (01)04562122958868(11)20170707(21)205Q0810; 205Q0812 (01)04562122958868(11)20170707(21)205Q0812; 205Q0815 (01)04562122958868(11)20171006(21)205Q0815; 205T7139 (01)04562122958868(11)20171006(21)205T7139; 205T7143 (01)04562122958868(11)20171026(21)205T7143; 205V5544 (01)04562122958868(11)20171026(21)205V5544; 205V5545 (01)04562122958868(11)20171030(21)205V5545; 205V5547 (01)04562122958868(11)20171031(21)205V5547; 205V5548 (01)04562122958868(11)20171031(21)205V5548; G3005239 (01)04562122958868(11)20190308(21)G3005239; G3005240 (01)04562122958868(11)20190308(21)G3005240; G3025041 (01)04562122958868(11)20190603(21)G3025041; G3025042 (01)04562122958868(11)20190604(21)G3025042; G3025044 (01)04562122958868(11)20190605(21)G3025044; G3025045 (01)04562122958868(11)20190605(21)G3025045; G3025048 (01)04562122958868(11)20190606(21)G3025048; G3025049 (01)04562122958868(11)20190606(21)G3025049; G3037183 (01)04562122958868(11)20190624(21)G3037183; G3037184 (01)04562122958868(11)20190624(21)G3037184; G3037192 (01)04562122958868(11)20190709(21)G3037192; G3043580 (01)04562122958868(11)20190820(21)G3043580; G3043581 (01)04562122958868(11)20190820(21)G3043581; G3043582 (01)04562122958868(11)20190820(21)G3043582; G3052217 (01)04562122958868(11)20191016(21)G3052217; G3052218 (01)04562122958868(11)20191016(21)G3052218; G3052225 (01)04562122958868(11)20191202(21)G3052225; G3052234 (01)04562122958868(11)20200128(21)G3052234; G3052236 (01)04562122958868(11)20200109(21)G3052236; G3054831 (01)04562122958868(11)20200115(21)G3054831; G3054833 (01)04562122958868(11)20191212(21)G3054833; G3054840 (01)04562122958868(11)20200115(21)G3054840; G3080616 (01)04562122958868(11)20201215(21)G3080616; G3080617 (01)04562122958868(11)20201215(21)G3080617; G3106256 (01)04562122958868(11)20210413(21)G3106256; G3106258 (01)04562122958868(11)20210413(21)G3106258; G3111572 (01)04562122958868(11)20210413(21)G3111572; G3111573 (01)04562122958868(11)20210421(21)G3111573

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 88840).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page