BioPlex 2200 recallZ-0236-2022 · Bio-Rad Laboratories, Inc. · initiated Sep 17, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bio-Rad Laboratories, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 09/17/2021, the Recalling Firm sent an "URGENT PRODUCT NOTIFICATION" Letter to customer via overnight FedEx informing them of two issues their Syphilis Total and RPR (Rapid Plasma Reagin) kits: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure. Customer are instructed to: -Immediately discontinue use and dispose of any unused product of reagent packs and calibrator set according to local waste management procedures. -Complete the customer response form and return it to their local Recalling Firm Technical Support, or return this form to [email protected] or the Toll-Free Fax 1-888-228-0688 within 10 days of receiving this notice.. -If the affected products have been further distributed or transferred, notify those customers of the field action. For additional questions, contact local Bio-Rad Technical Support.

Recall details

Recall number
Z-0236-2022
Status
Open, Classified
Initiated by the company
Sep 17, 2021
Recalling firm
Bio-Rad Laboratories, Inc.
Quantity
791 units
Product code
LIP
FDA submissions
K170413
Distribution
Worldwide Distribution: U.S.(nationwide) including states of: AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WV; O.U.S.(foreign) countries of: France, Germany, Israel, Italy, Russian Federation, South Africa, Spain, and Taiwan.

Product description

BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL The BioPlex 2200 Syphilis Total & RPR Calibrator Set is intended for the calibration of the BioPlex 2200 Syphilis Total & RPR Reagent Pack.

Code information (models, lots, serial numbers)

Catalog Number: 12000651 UDI Code: 03610520652770 Lot Number: 52953

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 88874).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page