Autolith Touch recallZ-0333-2022 · Northgate Technologies, Inc. · initiated Oct 25, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Northgate Technologies, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The languages for Russian, Romanian, Slovak, and Czech have the power settings for high and low switched in the device.
FDA root cause category: Software design (manufacturing process)
What the company is doing
The recalling firm, Northgate Technologies Inc., issued a recall letter dated 10/25/2021 to their customer explaining the reason for recall. The customer was requested to determine the location of the affected devices. If the device is/could be used in one of the affected languages, the device is to be returned so the software can be updated to address the text mix-up. If the device is not or will not be operated in the affected languages, the device does not need to be returned immediately and the software will be upgraded during the next servicing to address the mix-up. Complete the enclosed Recall Acknowledgment Form indicating the serial numbers, distribution dates, and whether the device needs correction now or later and email it to: [email protected]. In addition, if you have further distributed this product, please identify your customers and notify them at once of this product recall. If you have any questions, please contact me at (224) 856-2250 or by e-mail, [email protected].
Recall details
- Recall number
- Z-0333-2022
- Status
- Open, Classified
- Initiated by the company
- Oct 25, 2021
- Recalling firm
- Northgate Technologies, Inc.
- Quantity
- 2,277 devices
- Product code
- FFK
- FDA submissions
- K130368
- Distribution
- US Distribution to state of MA.
Product description
Boston Scientific AUTOLITH TOUCH Bipolar Electrohydraulic Lithotripter, REF M005466800. The device is designed to be used with Northgate Technologies Inc. bipolar disposable EHL probes for the fragmentation of calculi.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
All serial number revisions for the following bracketed base serial numbers distributed between 2/14/2017 and 10/7/2021: 21578LAA to 21627LAA 21628BBA to 21662BBA 21667CBA to 21689CBA; 21691CBA to 21697CBA; 21699CBA to 21716CBA 21690CBA, 21698CBA, 21721DBA to 21732DBA; 21734DBA to 21747DBA; 21749DBA to 21756DBA; 21758DBA to 21763DBA; 21765DBA to 21770DBA 21733DBA, 21748DBA, 21757DBA, 21764DBA, 21771EBA to 21781EBA; 21783EBA to 21787EBA; 21789EBA, 21790EBA, 21792EBA to 21800EBA; 21804EBA to 21820EBA 21782EBA, 21788EBA, 21791EBA, 21801EBA to 21803EBA; 21821FBA to 21874FBA 21883HBB to 21942HBB 21954IBB to 22025IBB 22026IBB to 22073IBB 22074IBB to 22130KBB; 22133KBB 22131KBB 22134LBB to 22157LBB 22158ACB to 22229ACB 22230ACB to 22249ACB 22250BCB to 22321BCB 22322CCB to 22393CCB 22394CCB to 22441CCB 22453ECB to 22476ECB 22477FCB to 22488FCB 22489GCB to 22500GCB 22511HCB to 22534HCB 22535ICB to 22558ICB 22559JCB to 22564JCB; 22570JCB to 22575JCB 22566JCB to 22569JCB; 22576JCB, 22577JCB, 22579JCB, 22581JCB to 22585JCB 22565JCB, 22578JCB, 22580JCB, 22586JCB to 22588JCB; 22590JCB to 22593JCB; 22595JCB, 22598JCB 22597JCB, 22600JCB to 22603JCB; 22607JCB to 22609JCB; 22611JCB, 22614JCB, 22615JCB, 22623JCB 22589JCB, 22594JCB, 22596JCB, 22599JCB, 22604JCB, 22606JCB, 22610JCB, 22612JCB, 22613JCB, 22616JCB to 22618JCB 22605JCB, 22619JCB to 22622JCB; 22624JCB to 22630JCB 22637KCB, 22638KCB, 22640KCB to 22666KCB; 22669KCB, 22671KCB to 22637KCB; 22677KCB, 22678KCB, 22680KCB 22636KCB, 22639KCB, 22639KCB to 22668KCB; 22670KCB to 22676KCB; 22679KCB, 22681KCB to 22707KCB 22632KCC to 22635KCC; 22708LCC to 22751LCC 22752ADC to 22799ADC 22800BDC to 22895BDC 22896CDC to 22943CDC 22944EDC to 22991EDC 22992FDC to 23039FDC 23040GDC to 23087GDC 23088IDD to 23135IDD 23136JDD to 23183JDD 23184KDD to 23231KDD 23232LDD to 23279LDD 23280AED to 23327AED 23328CED to 23375CED 23376DED to 23406DED; 23408DED to 23413DED; 23415DED, 23416DED, 23419DED, 23420DED, 23432DED, 23436DED to 23441DED 23407DED, 23414DED, 23417DED, 23418DED; 23421DED to 23431DED; 23433DED to 23435DED; 23442DED to 23471DED 23472HED to 23519HED 23520IED to 23567IED 23568KED to 23615KED 23616AFD to 23627AFD 23628AFD to 23663AFD 23664CFE to 23711CFE 23712DFE to 23759DFE 23760EFE to 23807EFE 23808GFE to 23855GFE 23856HFE to 23867HFE 23868IFE to 23879IFE 23880IFE to 23894IFE; 23896IFE to 23899IFE; 23902IFE to 23904IFE; 23907IFE, 23908IFE GTIN 00817183020493, UDI 00817183020493.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Autolith Touch M005466800Northgate Technologies Inc. · Lithotriptor, Electro-Hydraulic1 recall12 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K130368 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Lithotripsy Probe 1.9FR 375cm DS M00546620Northgate Technologies Inc. · Lithotriptor, Electro-Hydraulic1 recall1 death19 reports
- Autolith Touch M0054668R0Northgate Technologies Inc. · Lithotriptor, Electro-Hydraulic1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page