Skip to content

EVIS EXERA II DUODENOVIDEOSCOPE recallZ-0379-2022 · Olympus Corporation of the Americas · initiated Nov 2, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to patients.

FDA root cause category: Under Investigation by firm

What the company is doing

Olympus issued Urgent Medical Device Corrective Action on 11/02/21. Letter states reason for correction, health risk and action to take: 1. Consider transitioning from TJF-Q180V to a next generation duodenoscope through the Olympus incentive program. 2. Increase preventative maintenance frequency from annual to 6 months. 3. Evaluate existing duodenoscopes: https://medical.olympusamerica.com/TJFQ180VChecklist 4. The Technical Assistance Center (TAC) can assist you in answering questions on the TJF-Q180V inspection steps. 1-800-848-9024, option 1. 5. Use in-service training, the Olympus Video Guide, e-learning and on demand library to support staff training. ***UPDATE: Olympus issued an Urgent Medical Device Corrective Action UPDATED INFORMATION on 1/17/22. This is an update to the 11/2/21 previously provided supplemental material for inspection. In alignment with FDA Safety Communication dated August 29, 2019, and to further expedite transition to newer-generation duodenoscopes, Olympus America will soon begin offering customers with the older generation TJF-Q180V2 scopes a direct, no-cost 1 for 1 replacement with the state-of-the-art TJF-Q190V. Your Olympus sales representative will contact you regarding this program in the near future. In the interim, please continue to utilize the TJF-Q180V Inspection Checklist that Olympus recently sent. Olympus America will also be increasing the frequency of proactive inspection from once every twelve (12) months to once every six (6) months. To help minimize any potential patient impact regarding access to critical, life-saving ERCP procedures, loaners are available to mitigate disruption to healthcare facilities who need to return their duodenoscope(s) to Olympus for repair. For any questions or concerns, please do not hesitate to contact me directly at [email protected].

Recall details

Recall number
Z-0379-2022
Status
Open, Classified
Initiated by the company
Nov 2, 2021
Posted by the FDA
Dec 10, 2021
Recalling firm
Olympus Corporation of the Americas
Quantity
14,447 units
Product code
FDT
FDA submissions
K143153
Distribution
US Nationwide distribution.

Product description

EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V

Code information (models, lots, serial numbers)

All serial numbers. UDI: 04953170339967

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page