Medical device recalls: December 2021
The FDA posted 22 medical device recall records in December 2021, from 6 recall events by 6 companies. 19 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Covidien Llc · 1 event, 16 records, 16 Class I
- Datascope Corp. · 1 event, 2 records, 2 Class I
- Cardiovascular Systems Inc · 1 event, 1 record, 1 Class I
- NuVasive Specialized Orthopedics, Inc. · 1 event, 1 record
- Olympus Corporation of the Americas · 1 event, 1 record
- ROi CPS LLC · 1 event, 1 record
FDA root cause
- Reprocessing Controls · 16
- Device Design · 4
- Under Investigation by firm · 1
- Process control · 1
Class I recalls (19)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0336-2022 · Datascope Corp. · posted Dec 30, 2021 · Open, Classified
Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.
- Z-0337-2022 · Datascope Corp. · posted Dec 30, 2021 · Open, Classified
Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.
- Z-0394-2022 · Cardiovascular Systems Inc · posted Dec 21, 2021 · Terminated
Under certain circumstances, the WIRION filter assembly may become difficult to withdraw. Under such circumstances, alternative means may be necessary for withdrawal and the device may be prone to tearing or separating which may result in a risk for…
- Z-0338-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0339-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0340-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0341-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0342-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0343-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0344-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0345-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0346-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0347-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0348-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0349-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0350-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0351-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0352-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Z-0353-2022 · Covidien Llc · posted Dec 9, 2021 · Terminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Class II recalls (3)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0411-2022 · ROi CPS LLC · posted Dec 22, 2021 · Terminated
The kit contains a non-sterile needle.
- Z-0379-2022 · Olympus Corporation of the Americas · posted Dec 10, 2021 · Open, Classified
Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to patients.
- Z-0319-2022 · NuVasive Specialized Orthopedics, Inc. · posted Dec 1, 2021 · Open, Classified
The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.