Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers:… recallZ-0337-2022 · Datascope Corp. · initiated Nov 15, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.
FDA root cause category: Device Design
What the company is doing
On November 15, 2021, the firm issued an Urgent Medical Device Correction letter to all affected consignees which informed them that the exterior of the Cardiosave Hybrid and Rescue IABP may be susceptible to fluid ingress at specific locations on the device. IABPs contain various electronic circuit boards. Liquid spills, such as saline, can create bridges of resistance between the circuit components; causing the circuit to not function as intended. This can impact initiation or continuation of counterpulsation therapy. Datascope/Getinge had previously issued an Urgent Medical Device Correction letter on April 26, 2018 to install a Top Protective Cover for the Cardiosave Hybrid IABP to help reduce the potential for fluid ingress. However, in some instances and depending on the volume of spill, this Top Protective Cover can overflow and fluid can enter the device in other susceptible areas. Consignees were requested to immediately examine inventory to determine if there is affected product on hand and that a correction should be available in 2022. Consignees were advised that in the meantime, continue to follow the IFU and adhere to the instructions when using the Cardiosave Hybrid and/or Rescue IABP In the unlikely event that a sudden interruption of therapy occurs, transfer the patient to an alternative IABP. If an alternative IABP is unavailable; manually inflate the IAB with air or helium and immediately aspirate, repeat every 5 minutes until either an alternate IABP is available or alternatively, the intra-aortic balloon catheter should be removed from the patient. The Intra-Aortic Balloon (IAB) Catheter Instructions for Use reinforces that the IAB catheter should not remain inactive (i.e. not inflating and deflating) for more than 30 minutes because of the potential for thrombus formation. Consignees were requested to forward this information to all current and potential affected. Distributors were also requested to forward the letter to customer atte
Recall details
- Recall number
- Z-0337-2022
- Status
- Open, Classified
- Initiated by the company
- Nov 15, 2021
- Posted by the FDA
- Dec 30, 2021
- Recalling firm
- Datascope Corp.
- Quantity
- 44 (US); 5 (OUS)
- Product code
- DSP
- FDA submissions
- K112372, K151254, K163542
- Distribution
- Worldwide distribution - US Nationwide.
Product description
Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83, 0998-00-0800-85, 0998-UC-0800-85.
Code information (models, lots, serial numbers)
All Serial Numbers. May also be designated as part number D998 etc. Model Number, UDI-DI: 0998-00-0800-75, 10607567112312; 0998-UC-0800-75, 10607567112312; 0998-00-0800-83, 10607567108407; 0998-UC-0800-83, 10607567108407; 0998-00-0800-85, 10607567113449; 0998-UC-0800-85, 10607567113449.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE RESCUE, CHINESE 0998-00-0800-75Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls12 reports
- CARDIOSAVE RESCUE, 3.1 EDITION 0998-00-0800-85Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 88983).
- Z-0336-2022 · Open, ClassifiedCardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 09…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page