Puritan Bennett recallZ-0341-2022 · Covidien Llc · initiated Nov 4, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
FDA root cause category: Reprocessing Controls
What the company is doing
On November 4, 2021, the firm began to notify customers through Urgent Medical Device Correction letters. Customers were informed of the device defect that may lead to the ventilator becoming inoperable during use. Actions you should take Immediately discontinue use of the affected ventilators. Remove the affected ventilators from clinical service and quarantine the affected ventilators until such time that a Medtronic Technical Service Engineer inspects and replaces the affected printed circuit board assemblies. Notify all personnel in all care environments in which the affected PB980 series ventilators are used about this medical device correction. If your facility has distributed affected PB980 series ventilators to other persons or facilities, please promptly forward a copy of this letter to those recipients. Complete the attached form and return it as directed to confirm your receipt and understanding of this information. If you are aware of any incidents related to this issue, please contact our Technical Support Department immediately at 1-800-255-6774, option 4, and then option 1 to provide information regarding those events so regulatory reporting obligations can be fulfilled. Work with the Medtronic Technical Support Department if you require assistance finding alternative ventilation devices. Beginning November 5, 2021, Medtronic Technical Service will contact all affected customers to schedule a service appointment for all potentially affected ventilators. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.
Recall details
- Recall number
- Z-0341-2022
- Status
- Terminated
- Initiated by the company
- Nov 4, 2021
- Posted by the FDA
- Dec 9, 2021
- Recalling firm
- Covidien Llc
- Quantity
- 278 total
- Product code
- CBK
- FDA submissions
- K193056
- Distribution
- Worldwide distribution - US Nationwide.
Product description
Puritan Bennett 980 Series Ventilator, 980S3ENDICU
Code information (models, lots, serial numbers)
GTIN/UDI - 10884521171244 Serial Numbers: 35B1700406 35B1700449 35B1700467 35B1700469 35B1700471 35B1700474 35B1700481 35B1700526 35B1700558
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Puritan Bennett 980S3ENDICUCovidien LP · VENTILATOR, CONTINUOUS, FACILITY USE3 recalls4 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 88987).
- Z-0338-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980A1ENNISB
- Z-0339-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980A3ENNISB
- Z-0340-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980S1ENDICU
- Z-0342-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980U1ENASAA
- Z-0343-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980U1ESDIUC
- Z-0344-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980U3ESDIUU
- Z-0345-2022 · TerminatedPuritan Bennett 980 Series Ventilator,980X1ENAIQC
- Z-0346-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980X1ENDIPC
- Z-0347-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980X1ENDIUU
- Z-0348-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980X1ENNISB
- Z-0349-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980X1JADIJJ
- Z-0350-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980X1PLDIPC
- Z-0352-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980X3ENDIUU
- Z-0353-2022 · TerminatedPuritan Bennett 980 Series Ventilator, 980X3JADIJJ
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page