Dimension Vista 500- In-vitro diagnostic device intended to… recallZ-0331-2022 · Siemens Healthcare Diagnostics, Inc. · initiated Oct 15, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista Assays
FDA root cause category: Software design
What the company is doing
Siemens issued Urgent Medical Device Correction letter dated 10/27/21 to US Customers (VSW-22-01.A.US) and OUS Customers will be sent Urgent Field Safety Notice (VSW-22-01.A.OUS) to inform them of the issue and instruct them on setting the HIL index for the affected assays if applicable. The Dimension Vista software has been updated with the correct HIL index default settings recommended in this letter and included in software version 3.10.2. Letter states reason for recall, health risk and action to take: For the products listed in Table 1, please perform the following steps if the HIL feature is activated on your Dimension Vista System. If your laboratory is not using the HIL feature and it is not activated on your system no action is required: If the HIL indices for Dimension Vista FOL are activated, be sure to set the I index (icterus) to 6 which is aligned with the information in the Dimension Vista FOL IFU. If the HIL indices for Dimension Vista AMM are activated, be sure to set the L index (lipemia) to 2 which is aligned with the information in the Dimension Vista AMM IFU. Note: Instructions for setting HIL indices can be found in the Dimension System Operator s Guide, in Section Configuration, System Configuration. If the HIL indices for Dimension Vista CTNI are activated, be sure to set the I index (icterus) to 7 which is aligned with the information in the Dimension Vista CTNI IFU. If the HIL indices for Dimension Vista TPSA and/or FPSA are activated, be sure to set the H index (hemolysis) to 7 and the I index (icterus) to 6 which is aligned with the information in the Dimension Vista TPSA and FPSA IFUs. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days to Siemens Healthineers for reporting to the Authorities. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your lo
Recall details
- Recall number
- Z-0331-2022
- Status
- Terminated
- Initiated by the company
- Oct 15, 2021
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 1326 units
- Product code
- JJE
- FDA submissions
- K143720, K051087
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of DE, ES, FR, IT, CA, NL, JP, PT, AU, CZ, AT, DK, SI, BE, SK, QA, SA, KR, CH.
Product description
Dimension Vista 500- In-vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluid SMN: 10284473 (US) and 10488224 (OUS)
Code information (models, lots, serial numbers)
All software versions prior to 3.10.2. GTIN: 00630414949581 and 00630414989556 (GTIN + Instrument Serial Number = Unique Device Identification (UDI)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Dimension Vista® 500 10284473Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall524 reports
Other devices under the same FDA submission
4 devices cleared under K143720, K051087 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Dimension Vista® 500 Intelligent Lab System 10488217Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE13 reports
- Dimension Vista® V-LYTE® Integrated Multisensor Na+ / K+ / Cl- 10484429Siemens Healthcare Diagnostics Inc. · ELECTRODE, ION SPECIFIC, POTASSIUM10 reports
- Dimension Vista® 1500 Intelligent Lab System 10459469Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE9 reports
- Dimension Vista® Flex® reagent cartridge MMB 10445097Siemens Healthcare Diagnostics Inc. · COLORIMETRIC METHOD, CPK OR ISOENZYMES1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 89008).
- Z-0330-2022 · TerminatedDimension Vista 1500 In-vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determi…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page