Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX… recallZ-0445-2022 · Invacare Corporation · initiated Nov 12, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Invacare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

FDA root cause category: Software design

What the company is doing

Invacare issued Urgent Medical Device Letter issued on 11/12/21 to consignees who received devices with LiNX Gyro modules containing firmware version 6.1.2. This letter also contains instructions for consignees who upgraded devices in the field to version 6.1.2. A Consumer Letter will be provided on the recall website for Providers to send to end users who are impacted. Letter states reason for recall, health risk and action to take: Visit www. Invacare.com/RC916 within five (5) business days of receiving this notice. Enter the requested identification information to access the list of power wheelchair serial numbers and/or LiNX Gyro Modules sold to you and instructions to conduct this field correction. You will be prompted to provide responses to information requested regarding this field correction and affected units. " The account number required to access your list of impacted devices appears directly at the end of your company name in the mailing address. For example, ABC COMPANY-XXXXX. " The mailing postal code will also be required to log into your account. 2. Stop using LiNX Gyro Module firmware version 6.1.2 on all power wheelchair programming tools and stored electronic media. 3. Immediately update any LiNX Access iOS and LiNX Access PC tools containing LiNX Gyro Module firmware version 6.1.2 to the latest LiNX Access iOS or LiNX Access PC tool version. 4. Locate the power wheelchairs with affected LiNX Gyro Modules per the date ranges above. 5. For power wheelchairs built prior to the date ranges above, review your customer records to identify whether you have installed LiNX firmware version 6.1.2 on any LiNX Gyro-equipped power wheelchairs since the January 14, 2021 release date of firmware version 6.1.2. Note: If you do not have record of the firmware version installed on the LiNX Gyro Modules you have purchased from Invacare, you must connect to the LiNX system on the wheelchair using the LiNX Access tool to determine the firmware version of the Gyro

Recall details

Recall number
Z-0445-2022
Status
Open, Classified
Initiated by the company
Nov 12, 2021
Recalling firm
Invacare Corporation
Quantity
1477 units US; 1226 units OUS
Product code
ITI
FDA submissions
K170507
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.

Product description

Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: TDX SP2: TDXSP2, TDXSP2-CG, TDXSP2X-CG, TDXSP2-MCG, TDXSP2X-MCG, TDXSP2V, TDXSP2V-HD, TDXSP2HD, TDXSP2HD-CG, TDXSP2HD-MCG

Code information (models, lots, serial numbers)

LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: TDXSP2 00841447100911 TDXSP2-CG 00841447100942 TDXSP2-MCG 00841447100966 TDXSP2HD 00841447108641 TDXSP2HD-CG 00841447111207 TDXSP2HD-MCG 00841447111214 TDXSP2V 00841447101512 TDXSP2V-HD 00841447102052 TDXSP2X-CG 00841447101000 TDXSP2X-MCG 00841447101017

Devices named in this recall

10 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 89062).