Atellica CH 2-Microglobulin (B2M) -M) Assay for in vitro… recallZ-0440-2022 · Siemens Healthcare Diagnostics, Inc. · initiated Dec 1, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens issued An Urgent Medical Device Correction (UMDC) letter to US customers by FedEx beginning on 2021-12-01. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities beginning on 2021-12-01. Letter states reason for recall, health risk and action to take: Discontinue use of and discard the kit lot listed in Table 1. Review your inventory of this product to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-0440-2022
- Status
- Terminated
- Initiated by the company
- Dec 1, 2021
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- US: 46 units OUS: 794 units
- Product code
- JZG
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, CO, MA, MD, MS, NC, NE, NH, OH, VA, WI and the countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Czech Republic, Denmark, Ecuador, France, Germany, Greece, India, Iran, Israel, Italy, Japan, Mexico, Morocco, Oman, P.R. China, Pakistan, Philippines, Portugal, Slovakia, South Africa, Spain, Switzerland, Turkey, United Kingdom, Uruguay.
Product description
Atellica CH 2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of 2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635
Code information (models, lots, serial numbers)
Lot Number: 211446 UDI: 00630414595566
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica CH B2M 11097635Siemens Healthcare Diagnostics Inc. · System, test, beta-2-microglobulin immunological2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.