greiner bio-one VACUETTE TUBE, 5 ml LH Lithium Heparin Sep… recallZ-0502-2022 · Greiner Bio-One North America, Inc. · initiated Dec 9, 2021
Class II Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Greiner Bio-One North America, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some of the blood collection tubes show clotting due to variation of Lithium Heparin.
FDA root cause category: Under Investigation by firm
What the company is doing
The recalling firm initiated issuance of recall letters dated 12/9/2021 on the same date via email, phone, USPS, or UPS to their end-user customers. The letter explains the issue, the potential health hazard, and the necessary actions to be taken. The actions included to discontinue use of the affected product and isolate the products in their facility; assess risks and consequences for the use of the defective product in accordance with the consignee's procedures and take the appropriate action; and complete the attached Product Disposition Site Confirmation form and FAX it to the recalling firm. The form is a confirmation that the affected product has been discarded/destroyed. The product will be replaced by the recalling firm. The Customer Disposition form says if they have further distributed the product, they should provide the customer contact information to the recalling firm so that further follow-up can be made. For distributors. the letters were issued on 12/10/2021 via email. The email requests the distributor to provide a list of their customers to the recalling firm for further follow-up or to perform a subrecall themselves to the end-user customers. The distributor recall letter describes the problem and instructs the distributor to stop distributing the product and isolate it. A Product Disposition form was enclosed for completion to be returned via FAX or email. The recalling firm would then contact the distributor as to return or destruction for the product. The letter also informs the distributor it is a legal obligation to notify customers, therefore, they are requested to provide a list of their customers who have received the affected product, even though the cover email gives them the option of contacting the customers themselves.
Recall details
- Recall number
- Z-0502-2022
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 9, 2021
- Recalling firm
- Greiner Bio-One North America, Inc.
- Quantity
- 889,200 pcs.
- Product code
- JKA
- FDA submissions
- K960857
- Distribution
- Distribution was made to FL, IL, MO, NC, NM, NY, PA, TN, TX, and VA. There was no foreign/military/government distribution.
Product description
greiner bio-one VACUETTE TUBE, 5 ml LH Lithium Heparin Sep, 13x100 green cap-yellow ring, non-ridged, 24 racks of 50 pcs., 1200 pcs. in total, REF 456087P
Code information (models, lots, serial numbers)
Lot #B21053QG, Exp. 09/09/2022, UDI: Case: 39120017579046, Rack: 29120017579049.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Vacuette 456087PGreiner Bio-One GmbH · Tubes, vials, systems, serum separators, blood collection1 recall0 reports
Other devices under the same FDA submission
2 devices cleared under K960857 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Vacuette 454008PGreiner Bio-One GmbH · Tubes, vials, systems, serum separators, blood collection1 report
- Vacuette 456287PGreiner Bio-One GmbH · Tubes, vials, systems, serum separators, blood collection1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page