GLOBAL UNITE recallZ-0483-2022 · DePuy Orthopaedics, Inc. · initiated Dec 7, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage
FDA root cause category: Equipment maintenance
What the company is doing
DePuy notified US hospital / government consignees on 12/7/21 letter Urgent Medical Device Recall (Removal) via the sales consult and product will be removed/notice delivered by hand or if not able to enter facility the form will be emailed and returns will be arranged. US sales/distributor offices will be notified via email. Affiliates outside of the US affiliates will be notified via the electronic system and recall activities will be carried out per local regulations. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately to determine if you have the select lots and quarantine the product. 2. Return the subject product(s) using the normal returns process. Work with your sales consultant to return subject product(s). To receive replacement product(s) or reimbursement, customers must return the products subject to this removal. 3. Review, complete all fields, sign, and return the attached business response form (BRF) on the last page of this letter to [email protected] within three (3) business days of receipt of this Removal Letter. Please include in the email subject: FA 2044436 - 2021 GLOBAL UNITE Stems with undersized Spigot Bore Diameter. 4. Forward this Removal Letter to any personnel in your facility who need to be informed. 5. If any of the product(s) subject to this Removal have been forwarded to another facility, contact that facility and provide a copy of this Removal Letter to the relevant personnel. 6. Post a copy of this Removal Letter in a visible area for awareness of this Removal. If you have any questions, please contact your local DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir.
Recall details
- Recall number
- Z-0483-2022
- Status
- Terminated
- Initiated by the company
- Dec 7, 2021
- Recalling firm
- DePuy Orthopaedics, Inc.
- Quantity
- 28 units
- Product code
- PHX
- FDA submissions
- K170748, K101996
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Netherlands, New Zealand, Norway, Poland, South Africa, Sweden, Switzerland, Thailand, United Kingdom.
Product description
GLOBAL UNITE STD STEM SZ 6 Part Number: 11000610
Code information (models, lots, serial numbers)
Lot #: 9792595, 9793039, 9802121, 9813506. 9818509, 9830922, 9840359,9858603 GTIN: 10603295004011
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- GLOBAL UNITE 1100-06-100Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 recall2 reports
Product lines
Other devices under the same FDA submission
64 devices cleared under K170748, K101996 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- GLOBAL UNITE 1100-08-100Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall29 reports
- GLOBAL UNITE 1100-10-100Depuy (Ireland) · shoulder prosthesis, reverse configuration1 recall29 reports
- GLOBAL UNITE 1100-12-100Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 recall29 reports
- GLOBAL UNITE 1100-14-100Depuy (Ireland) · shoulder prosthesis, reverse configuration1 recall13 reports
- GLOBAL UNITE 1100-16-100Depuy (Ireland) · shoulder prosthesis, reverse configuration1 recall5 reports
- GLOBAL UNITE 1100-08-600Depuy (Ireland) · shoulder prosthesis, reverse configuration1 recall4 reports
- GLOBAL UNITE 1100-10-600Depuy (Ireland) · shoulder prosthesis, reverse configuration1 recall1 report
- DELTA XTEND 2307-90-003Depuy (Ireland) · shoulder prosthesis, reverse configuration43 reports
- GLOBAL UNITE 1100-30-000Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED37 reports
- GLOBAL UNITE 1100-40-000Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED28 reports
- GLOBAL UNITE 1100-44-600Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED21 reports
- GLOBAL UNITE 1100-44-610Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED21 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 89240).
- Z-0484-2022 · TerminatedGLOBAL UNITE STD STEM SZ 8 Part Number: 110008100
- Z-0485-2022 · TerminatedGLOBAL UNITE REV STEM SZ 8 Part Number: 110008600
- Z-0486-2022 · TerminatedGLOBAL UNITE STD STEM SZ 10 Part Number: 110010100
- Z-0487-2022 · TerminatedGLOBAL UNITE REV STEM SZ 10 Product Code: 110010600
- Z-0488-2022 · TerminatedGLOBAL UNITE STD STEM SZ 12 Product Code: 110012100
- Z-0489-2022 · TerminatedGLOBAL UNITE STD STEM SZ 14 Product Code: 110014100
- Z-0490-2022 · TerminatedGLOBAL UNITE STD STEM SZ 16 Product Code: 110016100
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page