Skip to content

Draeger Fabius MRI Anesthesia Machine, Catalog Number… recallZ-0501-2022 · Draeger Medical, Inc. · initiated Dec 16, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.

FDA root cause category: Process control

What the company is doing

On December 17, 2021, the firm notified customers via Urgent Medical Device Correction Notice. Customers were instructed to do the following: Please ensure that all users of the above-mentioned products and other concerned persons within your organization are made aware of this Urgent Medical Device Correction Notice. If you have provided the products to third parties, please forward a copy of this information. Please keep this information at least until the corrective measure has been completed. Please also complete and return the attached Customer Acknowledgment Card to confirm to us that you have received this information. A Drger Service representative will contact you to arrange a date for the correction to be performed free of charge. Until the repair has been concluded you may continue using the device under constant supervision as long as appropriate O2-monitoring is in place and a suitable alarm limit is set. Please consider a suitable buffer when adjusting the O2 flow. If you have any questions regarding the operation of your Fabius MRI Anesthesia Machine, please contact Drger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437- 2437 (press 2 at the prompt, then 2, then 2 again). If you have any questions regarding this recall communication, please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).

Recall details

Recall number
Z-0501-2022
Status
Terminated
Initiated by the company
Dec 16, 2021
Recalling firm
Draeger Medical, Inc.
Quantity
5
Product code
BSZ
FDA submissions
K072884
Distribution
US Nationwide distribution in the states of NY, PA, and WI.

Product description

Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300

Code information (models, lots, serial numbers)

UDI 04048675041498 Serial Numbers ASPJ-0019 ASPJ-0020 ASPJ-0021 ASPK-0001 ASPK-0002

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page