MEDPOR TITAN recallZ-0715-2022 · Stryker Leibinger GmbH & Co. KG · initiated Dec 29, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Leibinger GmbH & Co. KG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Inside the packaging of one reported item a non-conforming product was found. Contrary to the design, the barrier layer was manufactured on the inferior side of the implant versus the superior side.

FDA root cause category: Process control

What the company is doing

An URGENT MEDICAL DEVICE Recall notification letter dated 12/29/21 was sent to customers. Actions needed 1. Immediately check your internal inventory to locate the product listed on the attached business reply form. 2. Return the enclosed business reply form by email to [email protected] to confirm receipt of this notification. 3. Upon receipt of the completed business reply form, we will contact you to arrange for the return and replacement of your affected product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. Inform a Stryker representative if any of the subject devices have been distributed to other organizations. If so, provide contact details so that we may inform the recipients appropriately. If you have any questions or concerns, please contact Customer Service +1 800 962-6558.

Recall details

Recall number
Z-0715-2022
Status
Terminated
Initiated by the company
Dec 29, 2021
Recalling firm
Stryker Leibinger GmbH & Co. KG
Quantity
74
Product code
FTM
FDA submissions
K040364
Distribution
Worldwide distribution - US Nationwide.

Product description

MEDPOR Titan OFW - MTB - Right - 1.0mm, Catalog number 81032

Code information (models, lots, serial numbers)

UDI: 07613252084334; Lot Number: 3R7YHD

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K040364 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page