Skip to content

TurboHawk Plus recallZ-0695-2022 · ev3 Inc. · initiated Feb 4, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ev3 Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The device has similarities in design to another device that the firm recently issued a field correction on due to instances of tip detachment and embolization.

FDA root cause category: Labeling design

What the company is doing

The recalling firm began issuing the URGENT MEDICAL DEVICE NOTICE dated February 2022 on 2/4/2022 via 2-day UPS delivery. The letter explained the issue and instructed the consignee to adhere to the existing IFU instructions and warnings and precautions listed in the letter to reduce the risk of guidewire prolapse. The Customer Instructions included the notice was to be shared with all who needed to be aware within the consignee's organization or to any other organization where the product has been transferred; to review the IFU included with the product prior to use, noting the warnings and precautions listed in the notice; and to complete the Customer Confirmation Form to acknowledge receipt of the notice and that the appropriate action has been taken. It was to be emailed to the recalling firm.

Recall details

Recall number
Z-0695-2022
Status
Terminated
Initiated by the company
Feb 4, 2022
Posted by the FDA
Mar 7, 2022
Recalling firm
ev3 Inc.
Quantity
245 devices
Product code
MCW
FDA submissions
K212027
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IA, IL, KS, LA, MI, MS, NC, NJ, NY, OR, SC, SD, TN, TX, and VA.

Product description

Medtronic TurboHawk Plus Directional Atherectomy System 6F, Small Vessel, Catalog number THP-S. For use in peripheral vasculature.

Code information (models, lots, serial numbers)

GTIN 00763000402419

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K212027 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 89470).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page