Alinity m recallZ-0710-2022 · Abbott Molecular, Inc. · initiated Feb 11, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Molecular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is potential for the Reaction Vessels (RVs) detaching from the RV cap during transport by the pipettor robot.

FDA root cause category: Under Investigation by firm

What the company is doing

On February 11, 2022, Abbott Molecular issued a Product Recall letter to all customers who received Integrated Reaction Unit (IRU) List Number 09N26-010 Lot Numbers 766233, 770433, 771200, 771201, and 773642 used with the Alinity m System. Consignees must complete and return the included customer reply form. Customers are instructed to follow the Necessary Actions outlined in the Customer Letter: Discontinue use of and discard all remaining inventory of IRU Lots 766233, 770433, 771200, 771201, and 773642. To coordinate replacement, contact your local Molecular Diagnostics at Abbott representative. If RVs are found inside the instrument with no cap and there are associated 1998 errors, please contact an Abbott representative to inspect the instrument. Results generated between the onset of the associated 1998 errors and the identification of the RV with no cap inside the instrument should be considered presumptive. Retesting of samples that yielded positive results during this timeframe should be considered.

Recall details

Recall number
Z-0710-2022
Status
Terminated
Initiated by the company
Feb 11, 2022
Recalling firm
Abbott Molecular, Inc.
Quantity
3063 kits
Product code
OOI
Distribution
US Nationwide distribution in the states of AZ, CA, CO, CT, DC, FL, GA, IL, KY, LA, MA, MI, MO, NC, NJ, NY, OK, RI, SD, TX, WA, and WI.

Product description

Alinity m Integrated Reaction Units (IRU)

Code information (models, lots, serial numbers)

List Number 09N26-010, Lot Numbers/UDI: 1) 766233 / (01)20884999047864(10)766233(11)210614(240)09N26-010 2) 770433 / (01)20884999047864(10)770433(11)210621(240)09N26-010 3) 771200 / (01)20884999047864(10)771200(11)210712(240)09N26-010 4) 771201 / (01)20884999047864(10)771201(11)210719(240)09N26-010 5) 773642 / (01)20884999047864(10)773642(11)210723(240)09N26-010

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines