1.5T and 3.0T SIGNA HDx, 1.5T and 3.0T SIGNA HDxt, and 1.5T… recallZ-0860-2022 · GE Healthcare, LLC · initiated Feb 18, 2022

Class II Open, Classified All recalls posted in April 2022

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is potential for the images to be flipped left to right.

FDA root cause category: Device Design

What the company is doing

This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions and link for customer to access and download the revised service manual.

Recall details

Recall number
Z-0860-2022
Recall class
Class II
Status
Open, Classified
Initiated by the company
Feb 18, 2022
Posted by the FDA
Apr 4, 2022
Recalling firm
GE Healthcare, LLC
Quantity
653 units
Product code
LNH
FDA submissions
K121676, K052293
Distribution
Worldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, China, Denmark, Ecuador, Ethiopia, France, Germany, Greece, Hong Kong, India, Iran, Ireland, Italy, Japan, Korea, Kuwait, Lebanon, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Turkey, United Arab Emirates, and United Kingdom.

Product description

1.5T and 3.0T SIGNA HDx, 1.5T and 3.0T SIGNA HDxt, and 1.5T SIGNA HDxt Mobile. Magnetic Resonance Imaging System

Code information (models, lots, serial numbers)

Show the full code information (8k characters)

System IDs: 907562P3T, 256734MR3T, 623879MR1, 623433MR1, 480584MR1, 602689MR1, 9287573TMR, 480324MR1, 480614MR1, 928317MR1, 714868MR1, RADNETMR615, 714619MR, 5107923TMR, 858657MR1, 323442MR3, 323442MR1, 213977MR2, 323932MR1, 323442MR2, RADNETMR612, 650259SM3T, RADNETMR1521, RADNETMR592, 626397MR2, RADNETMR620, 916DR3TMR, 415UCLPMR, 415600D3T, 415476UMR6, 415353MRC1, 415MTZMR, 408GS3TMR, RADNETMR1595, 408885MR3T, RADNETMR608, 805569MR, 415723SHMR8, 661AIMR2, RADNETMR594, 805CHLIMR1, 720848MR2, 303340MR, 030022MR01, 515282MR1, 859275MR3T, 2709263HD, 504873TMR3, 504DISWMR, 504DISMR3T, BEV3TMR, 9786823T, RADNETMR1659, RADNETMR1589, 207MMCSMR, 517353VH3, 269342MR3T, 269342MR3T2, 517374MR3T, 248465MR3T, 248849MR3T, 662MSU3T, 406751K3T, CONCORD3T, 201669MR2, 716859MR3T, NYP347652, 212305MR83T, 5857853T, 631686MR3T, 5853413TLAC, 419383MR3T, 503280MR2, 6109883TMR, 423844MR2, 865541MR3T, 615777MR3T, 936441MR1, 817882H3TMR, 513556MR, 713441MR5, 512TSIMR2, 210617TCAMR2, 210617TCAMR1, 801221MR, 509758TMR, 425288CMR, 2063203TMR, 414973MR2, 608785HDMR, 414649MR5, 414973MR1, 422SIGNA3T, 250374MRTS, 604682MRI, 604682MRI2, 604739MR, 082427120201, DK1030MR02, ET1009MR01, 493013MR01, HC1819MR01, 850270170, 850270218, MR281340MR0, DVT0127, PER19516, PER08626, RU1095MR03, ZA2222MR01, MPX82251, GP178328, FPG98423, FPG98415, MEX68900, DPN88605, ME051580, MPX10000, EPP00302, 725030MR01, 00149MRS01, 205939MR4, 205783TSMR, 2059393T, 928213MR3T, 480354MR1, 602588CMR1, 480464MR1, 4809643TMR, 954SIMON3T, 480478MR1, 602406MR6, 831423MR, 4802643TMR, 480425MR5, 480264MR2, 5204493T, RADNETMR590, 310623MR, 626359MR3, 760ENCMR, RADNETMR627, 213777MR, 323268MR2, 213387MR, 510MIA3T, 951SWMR, 805GI3T, 760323MR3T, 650FALK3TMR, 650853PAMR2, RADNETMR617, RADNETMR646, 909806MR, 559585CCICMR2, 916UCDMR3T, 916KM3TMR, RADNETMR609, 858453VMR2, 858822MR3, 4419, RADNETMR1706, 510939JMMR3T, 626962MR, 303SMIMR1, 303633MR, 970244MR3T, 719542MR, 719595MR, 860659MR, 203863MR3T, 860545MR2, 860231MR, 202865MR3T, 407BDIMR, 3173383T1, SIMONMED3TBRANDON, 850916A3T, 407OSCMR3T, 239307MR, NCHMBMR3T, 904276OPTS, PRMTMR3T, 850475IC3T, 727440MR1, 727898MR1, 412692MR1, 813251MR3T, 813972MR3T, 772VRAMR3T, 407SMWPMR1, 7067213TMR, 7705353TMR, 404772NMR3T, 808591MR, 3096653TMR, 773665MR3T, 847SAMC3T, 708383WS3T, 309672MR3T, 859313MR, 225763MR, RADNETMR1636, 301459MR, 410535MR, 410543MR, 3137453TMR, 269341MR3T, 313294MR, 248551MR3T, 2489643T, 9525413TMR, 612626MR3T, 2187493T, 573815BMR1, 636947JMR1, 601268MR3T, 910609CF3T, 402219MR3T, 908725MR, 7326133T, 201632MR, 732632MR3, 609652MR3T, 732223MR, 609497MR3T, 919684MR3, 7024230SM3THL, M48823, RADNETMR554, COR373219, RADNETMR1583, 631982MR, 405CDIMR, 503652MR1, 503216MR3, 503215MR1, 717975MR3, 6053383TMR, 6053223TMR, 731935MR3, 865218MR3T, 901572L3T, 615284MR3T, 615TANMR, 865481MR3T, 865545MR3T, 806NWTH3TMR, 214559SRHMR1, 214329PHIMR1, 214341PDGPMR1, 281333MR1, 713524BIMR1, ACB41121MR, 713704HMR4, 212288GE3T, ACB41109MR, 972665XBPMR1, 210804SUNMR, 801373MR3, 435688MR3, 801321LMR2, 70389MR3T, 703573MR6, 8047643T, 8043303T, 540776MR3, 425297RMR, 304636MR, 304526CMR2, 16683THDX, 2901847MR3T, 910272046, 910273033, 83160776416, AZ4013MR01, BA4080MR01, 2885417, 2854971, MRR8701, 2766544, 2806189, 2871974-1, MRR9017, MRR2396, MRWB0540, 604557MR1, 905522MR, 902678MR1, 416369MR3T, 519254MR, 082427090190, 082427040149, 082427110026, 082427090170, 082427080033, 082427100192, 082427170011, 082427170012, 082427130067, 082427070069, 082427070120, 082427310042, 082427130090, 082427130046, 082427130044, 082427130084, 082427130059, 082427130061, 082427130083, 082427130062, 082427130042, 082427130077, 082427160081, 082427160105, 082427080051, 082427100220, 082427100201, 082427100173, 082427030115, 082427270016, 082427200052, 082427270030, 082427070054, 082427090115, 082427290013, 082427040101, 082427040105, 082427040088, 082427040113, 082427040147, 082427040099, 082427040100, 082427040092, 082427120070, 082427120102, 082427150051, 082427200056, 082427200036, 082427120086, 082427010074, 082427110042, 082427110025, 082427110034, 082427020037, 082427090174, 082427060051, 082427060063, 082427060032, 082427060025, 082427010054, 082427090246, 082427090292, 082427080044, 082427010038, 082427250060, 082427030106, 082427040094, 082427050027, 082427080026, 082427100172, 082427190049, 082427140150, 082427090171, 082427090164, 082427240019, 082427210063, 082427270022, 082427160159, 082427160084, 082427160110, 082427160065, 082427160064, 082427160113, 082427070062, 082427020065, 082427070055, 082427070116, 082427120091, 082427120199, 082427160117, 082427230068, 082427090335, 082427090245, 082427120126, 082427120062, 082427120077, 082427100153, 082427250088, 082427250067, 082427250030, 082427270040, 082427270032, 082427140134, 082427140120, 082427140062, 082427020025, 082427210049, 082427210053, 082427210070, 082427210060, 082427210058, 082427240048, 082427310046, 082427160103, 082427030042, 082427030116, 082427030050, 082427150056, 082427100228, 082427090288, 082427090124, 082427100126, 082427100161, 082427100176, 082427170008, 082427100102, 082427160203, 082427160072, 082427140067, 082427140051, 082427140077, 082427110047, 082427110044, 082427100164, 082427100198, 082427100185, 082427160070, 082427010077, 082427010093, 082427010095, 082427010069, 082427120093, 082427030071, 082427070063, 082427150103, 082427150040, 082427150048, 082427070101, 082427160272, 082427110027, 082427120072, 082427200072, 082427030090, 082427030064, 082427190066, 082427210127, 082427280002, 082427310108, 082427310032, 082427310029, 082427310048, 082427310034, 082427240041, 082427200039, 082427060028, 082427060038, 082427120083, 082427020041, 082427240030, 082427240070, 082427090139, 082427090208, 082427090229, 082427090267, 082427090227, 082427090263, 082427090238, 082427090134, 082427090141, 082427090226, 082427090289, 082427090232, 082427190078, 082427060034, 082427220018, 082427220041, 082427220016, 082427030056, 082427030086, 082427300010, 082427300014, 082427300018, 082427300012, 082427020026, 082427190075, 082427190069, 082427100119, 082427100309, 082427100182, 082427090173, 082427010141, 082427010047, 082427010150, 082427240034, 082427010065, DK1018MR01, ET1293MR02, S41600123, HC0074MR01, MC6946MR07, H10392MR03, H2193MR05, 0828270039, 083027828659711, 083027041002010, 083027601000510, 083027609389015, 083027605826311, 083027202000249, 083027024220111, 083027022101691, 083027202001698, 083027202101396, 083027207848338, 083027002101567, 083027006495511, 083027002101706, 083027009445910, 083027007755738, 083027002730811, 083027202101819, 083027243783711, 083027862100837, IE1111MR01, A469231301, A5632416, C5323526, YM1404, YM1261, YM1799, YM0867, YM3040, YM0873, YM1836, YM1839, YM2021, YM1851, YM1837, YM1831, YM1817, YM1832, YM1821, YM1401, YM3900, YM1805, YM1395, YM3883, YM1815, YM1841, YM3032, YM1797, YM1794, YM1881, YM1796, YM1788, YM1842, YM1834, YM1840, YM1800, YM1882, YM1813, YM1812, YM1850, YM1823, YM1843, YM1810, YM1845, YM1820, YM2018, YM1400, YM0796, YM0790, YM1870, YM0888, YM3042, YM1804, YM1880, YM1877, YM2003, YM1808, YM0802, YM1411, YM1798, YM1819, YM1809, YM1822, YM0801, YM1848, YM1828, YM1844, YM1792, YM1793, YM1402, YM1818, YM2020, YM1824, YM1846, YM0791, YM1879, YM1399, YM1826, YM1849, YM1847, YM3939, YM0872, YM2005, YM1827, YM1803, YM2012, YM1835, YM1816, 850270200, 850270202, 850270333, 850270220, LB1000MR02, 2716757SIGNAHX, 920279024, PL3800MR01, PL3545MR01, PL1050MR01, PL3040MR01, PL3546MR01, PER28007, 787641HDXT, RO1189MR01, RU1588MR01, RU1087MR01, RU1089MR04, RU1067MR02, RU1688MR01, RU9155MR02, RU1884MR01, SA1012MR01, SA1001MR03, RS4202MR01, SI4024MR02, ZA2169MR01, DEL26444, CEN03406, DPC79014, DEN51500, NPV84502, G002MR01, SY2012MR02, 0853270074, 0853270095, 0853270081, 0853270094, 0853270088, AE1190MR01, 70328MRS04

Other devices under the same FDA submission

1 device cleared under K121676, K052293 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page