Medical device recalls: April 2022
The FDA posted 44 medical device recall records in April 2022, from 19 recall events by 17 companies. 6 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- GE Healthcare, LLC · 3 events, 7 records
- Siemens Medical Solutions USA, Inc · 1 event, 14 records
- Respironics California, LLC · 1 event, 3 records, 3 Class I
- restor3d Inc. · 1 event, 3 records
- ROi CPS LLC · 1 event, 3 records
- Abbott Molecular, Inc. · 1 event, 2 records
- Philips North America LLC · 1 event, 2 records
- Baxter Healthcare Corporation · 1 event, 1 record
FDA root cause
- Device Design · 22
- Under Investigation by firm · 8
- Nonconforming Material/Component · 4
- Process change control · 3
- Radiation Control for Health and Safety Act · 2
- Process control · 2
- Software Design Change · 1
- Other · 1
Class I recalls (6)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0953-2022 · Celltrion USA Inc · posted Apr 27, 2022 · Terminated
Point of Care (PoC) rapid test products were distributed to customers who did not have a valid CLIA ID.
- Z-0966-2022 · Medtronic formerly Covidien · posted Apr 25, 2022 · Open, Classified
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac…
- Z-0907-2022 · Respironics California, LLC · posted Apr 18, 2022 · Open, Classified
Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
- Z-0908-2022 · Respironics California, LLC · posted Apr 18, 2022 · Open, Classified
Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
- Z-0909-2022 · Respironics California, LLC · posted Apr 18, 2022 · Open, Classified
Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
- Z-0927-2022 · Medtronic Inc · posted Apr 18, 2022 · Open, Classified
There is potential for the capsule bond to break during the procedure.
Class II recalls (38)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1019-2022 · Baxter Healthcare Corporation · posted Apr 29, 2022 · Terminated
There is a potential for foreign matter.
- Z-1026-2022 · restor3d Inc. · posted Apr 29, 2022 · Terminated
Product labeled incorrectly. The lot was labeled with Small footprint when the product was Medium.
- Z-1027-2022 · restor3d Inc. · posted Apr 29, 2022 · Terminated
During re-work of disposable instruments to relabel for 4-year shelf-life of disposable instruments, parts got mixed into other lots.
- Z-1028-2022 · restor3d Inc. · posted Apr 29, 2022 · Terminated
During re-work of disposable instruments to relabel for 4-year shelf-life of disposable instruments, parts got mixed into other lots.
- Z-1014-2022 · Instrumentation Laboratory · posted Apr 28, 2022 · Open, Classified
Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid…
- Z-1010-2022 · Siemens Medical Solutions USA, Inc. · posted Apr 27, 2022 · Open, Classified
Transducers (component of ultrasound system) may exceed recommended thermal index limits
- Z-1004-2022 · ROi CPS LLC · posted Apr 26, 2022 · Terminated
A drug kit component was recalled by its manufacturer due to the potential for inadequacy of the validation of test methods used in its manufacture.
- Z-1005-2022 · ROi CPS LLC · posted Apr 26, 2022 · Terminated
A drug kit component was recalled by its manufacturer due to the potential for inadequacy of the validation of test methods used in its manufacture.
- Z-1006-2022 · ROi CPS LLC · posted Apr 26, 2022 · Terminated
A drug kit component was recalled by its manufacturer due to the potential for inadequacy of the validation of test methods used in its manufacture.
- Z-0998-2022 · Medtronic Neuromodulation · posted Apr 22, 2022 · Terminated
There is a software anomaly with the InterStim X Clinician software application with results in the data entered in the Patient Information fields not e retained and a " Data Lost" notification being displayed.
- Z-0999-2022 · Intuitive Surgical, Inc. · posted Apr 22, 2022 · Open, Classified
As a result of complaint data review, it identified tissue pushout events occurring during firing of the SureForm stapler when tissue is pushed distally within the jaws, rather than being held in place and transected. Overall rate of harm associated with…
- Z-0967-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0968-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0969-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0970-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0971-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0972-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0973-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0974-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0975-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0976-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0977-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0978-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0979-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0980-2022 · Siemens Medical Solutions USA, Inc · posted Apr 21, 2022 · Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the…
- Z-0983-2022 · GE Healthcare, LLC · posted Apr 21, 2022 · Open, Classified
During the de-installation of an MR system magnet, if all rigging and hardware that attach to the magnet for transportation are not properly installed and secured, the magnet may fall, leading to potential injury.
- Z-0894-2022 · Ventana Medical Systems Inc · posted Apr 8, 2022 · Open, Classified
Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.
- Z-0877-2022 · Hologic, Inc · posted Apr 6, 2022 · Open, Classified
Samples, run on a GBS Assay, that have analyte concentrations that are at or near the limit of detection may produce false negative results.
- Z-0881-2022 · Philips North America LLC · posted Apr 6, 2022 · Open, Classified
There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may separate almost completely leaving…
- Z-0882-2022 · Philips North America LLC · posted Apr 6, 2022 · Open, Classified
There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may separate almost completely leaving…
- Z-0883-2022 · Abbott Molecular, Inc. · posted Apr 6, 2022 · Open, Classified
There is a potential for misquantitation high results of negative samples.
- Z-0884-2022 · Abbott Molecular, Inc. · posted Apr 6, 2022 · Open, Classified
There is a potential for misquantitation high results for negative samples.
- Z-0815-2022 · GE Healthcare, LLC · posted Apr 5, 2022 · Open, Classified
The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.
- Revolution CT, Revolution CT ESClass IIZ-0816-2022 · GE Healthcare, LLC · posted Apr 5, 2022 · Open, Classified
The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.
- Z-0857-2022 · GE Healthcare, LLC · posted Apr 4, 2022 · Open, Classified
There is potential for the images to be flipped left to right.
- Z-0858-2022 · GE Healthcare, LLC · posted Apr 4, 2022 · Open, Classified
There is potential for the images to be flipped left to right.
- Z-0859-2022 · GE Healthcare, LLC · posted Apr 4, 2022 · Open, Classified
There is potential for the images to be flipped left to right.
- Z-0860-2022 · GE Healthcare, LLC · posted Apr 4, 2022 · Open, Classified
There is potential for the images to be flipped left to right.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.