Skip to content

Nextra Hammertoe Correction System recallZ-0913-2022 · Nextremity Solutions · initiated Feb 10, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nextremity Solutions or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery

FDA root cause category: Nonconforming Material/Component

What the company is doing

Nextremity Solutions issued Urgent Medical Device Recall letter on 2/10/22 via mail and email to Distributors and Risk Mangers. Nextremity Solutions products with the referenced part and lot number should not be used and should be quarantined and returned to Zimmer Biomet for return to Nextremity Solutions. If you have affected product at your facility, locate and quarantine affected product in your inventory. Zimmer Biomet Sales representative will remove the affected product from your facility. Complete Medical Device Recall Return Response and send to [email protected], even if you do not have affected products at your facility. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Or email to [email protected].

Recall details

Recall number
Z-0913-2022
Status
Terminated
Initiated by the company
Feb 10, 2022
Recalling firm
Nextremity Solutions
Quantity
89 units
Product code
HWC
FDA submissions
K110445
Distribution
Distribution US Nationwide, Japan, and the Netherlands

Product description

Nextra Hammertoe Correction System 3.5 Middle and 3.2 Proximal Saw Cut SKU: NX-3532K-SC Orthopedic surgical instrument

Code information (models, lots, serial numbers)

Lot Number: 168125017B UDI: 00817701020059

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 89764).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page