Nextra Hammertoe Correction System NX-3532K-SCRecalls & reported problems
Medartis Inc. · 3.5 Middle 3.2 Proximal Saw Cut · Class 2
All models of Nextra Hammertoe Correction System (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00817701020059), received between Mar 14, 2019 and Feb 10, 2021 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Feb 10, 2022Class IITerminated
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
Z-0913-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- No Apparent Adverse Event4
- Device Dislodged or Dislocated1
- Mechanical Problem1
- Unintended Movement1
Patient problems
- No Known Impact Or Consequence To Patient5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2021No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A SURGERY WITH THE NEXTRA HAMMERTOE CORRECTION SYSTEM, THE PROXIMAL AND MIDDLE IMPLANTS DID NOT SIT FLUSH ON THE NEXTRA DRIVER. THIS CONDITION DID NOT AFFECT THE SURGERY. THE SURGERY WAS COMPLETED AND THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11309456 ↗MalfunctionMar 4, 2020No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A SURGERY ON (B)(6) 2020 USING THE NEXTRA HAMMERTOE CORRECTION SYSTEM, THE PROXIMAL IMPLANT COMPONENT COULD NOT BE DISENGAGED FROM THE DRIVER. A SECOND KIT WAS OPENED TO COMPLETE THE SURGERY. THERE WERE NO REPORTED PATIENT COMPLICATIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9786594 ↗MalfunctionJan 10, 2020No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A SURGERY WITH THE NEXTRA HAMMERTOE CORRECTION SYSTEM THE PROXIMAL IMPLANT WOULD NOT SIT FLAT ON THE SCREWDRIVER. A SECOND SCREWDRIVER WAS OPENED TO COMPLETE THE SURGERY. THERE WERE NO REPORTED PATIENT COMPLICATIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9578284 ↗MalfunctionJan 10, 2020No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT DURING AN OPERATION PERFORMED WITH THE NEXTRA HAMMERTOE CORRECTION SYSTEM, THE SURGEON HAD DIFFICULTY MATING THE DISTAL END OF THE NEXTRA DRIVER WITH THE DISTAL IMPLANT. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH AN ALTERNATE DRIVER. THERE WERE NO REPORTED PATIENT COMPLICATIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9576550 ↗MalfunctionMar 14, 2019Mechanical Problem; Device Dislodged or Dislocated; Unintended Movement; Mechanical Problem; Device Dislodged or Dislocated; Unintended Movement
THE PATIENT UNDERWENT HAMMERTOE SURGERY ON (B)(6) 2019. AFTER SURGERY, X-RAYS SHOWED THE IMPLANT HAD DISLOCATED. THE PATIENT CAME BACK INTO THE DOCTOR'S OFFICE THE FOLLOWING DAY FOR THE SURGEON TO ATTEMPT TO FIX THE DISLOCATION. THIS WAS NOT SUCCESSFUL. THE PATIENT WAS SCHEDULED FOR SURGERY ON (B)(6) 2019, AT WHICH POINT THE SURGEON REMOVED THE DISLOCATED 3.5MM MIDDLE IMPLANT. THE SURGEON IMPLANTED A 4.5 MIDDLE IMPLANT TO REPLACE THE IMPLANT. THIS WAS SUCCESSFUL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8422053 ↗
FDA clearance and classification
The FDA regulates product code HWC (screw, fixation, bone) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Nextra Hammertoe Correction System NX-3532K-SC
Has the Nextra Hammertoe Correction System NX-3532K-SC been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0913-2022 (Class II), began on Feb 10, 2022: Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Nextra Hammertoe Correction System NX-3532K-SC?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Mar 14, 2019 and Feb 10, 2021. In those reports, the most common specific device problems are “Device Dislodged or Dislocated” (1), “Mechanical Problem” (1) and “Unintended Movement” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Nextra Hammertoe Correction System NX-3532K-SC?
None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Nextra Hammertoe Correction System NX-3532K-SC, and is it still sold?
The FDA's device database lists Medartis Inc. as the company responsible for it (brand Nextra Hammertoe Correction System, model NX-3532K-SC). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Screw, Fixation, Bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone498 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone248 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-325Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-550Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-425Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page