Acuson 9vE4 Transducer, Model (240)11289564 CE0129, Rx Only recallZ-1010-2022 · Siemens Medical Solutions USA, Inc. · initiated Mar 7, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Transducers (component of ultrasound system) may exceed recommended thermal index limits
FDA root cause category: Process control
What the company is doing
On 4/1/22, Siemens issued a "Urgent Medical Device Safety Correction" via: email to affected consignees. Prior to receiving the email the firm's representatives contacted consignees by phone or by visit to inform them of the recall. In addition, to providing consignees with information on the recalled product, Siemens asked consignees to take the following actions: Immediately stop performing procedures using the ACUSON 9VE4 transducer and segregate the product. Please ensure that all users of the affected transducer within your organization, and others who may need to be informed, receive the safety relevant information provided with this notice and take the actions specified herein. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Medical Device Safety Correction. To acknowledge receipt of this notice, you must DIGITALLY SIGN it by clicking on the link provided in the email where this notice was attached. By digitally signing this document, you are acknowledging that you have read and understand the content of this Urgent Medical Device Safety Correction. If an adverse event or quality problem is experienced with the use of this product, the issue may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Siemens will correct the issue with a free-of-charge replacement ACSUON 9VE4 transducer. If you haven't been contacted yet, siemens representative will call to schedule a time to document and retrieve the transducer that was removed from service
Recall details
- Recall number
- Z-1010-2022
- Status
- Open, Classified
- Initiated by the company
- Mar 7, 2022
- Posted by the FDA
- Apr 27, 2022
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Quantity
- 31 units
- Product code
- IYN
- FDA submissions
- K211859
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of Washington, Alaska, Ohio, South Carolina, Pennsylvania, Arizona, New York, Texas and the countries of Australia, Czech Republic, France, United Kingdom, Indonesia, Mexico, New Zealand.
Product description
Acuson 9vE4 Transducer, Model (240)11289564 CE0129, Rx Only ,
Code information (models, lots, serial numbers)
UDI/DI - (01)4056869243566; Affected Serial Numbers: BT10672 BT10582 BT10617 BT10643 BT10669 BT10580 BT10619 BT10659 BT10611 BT10612 BT10614 BT10673 BT10581 BT10494 BT10610 BT10641 BT10497 BT10615 BT10661 BT10667 BT00248 BT10613 BT10657 BT10640 BT10584 BT10488 BT10481 BT10644
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- 9VE4 Transducer 11289564Siemens Medical Solutions USA, Inc. · Transducer, ultrasonic, diagnostic1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page