Artis icono biplane- in combination with a Siemens… recallZ-0985-2022 · Siemens Medical Solutions USA, Inc · initiated Feb 25, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed
FDA root cause category: Software Design Change
What the company is doing
Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice to all affected customers via AX005/22/S on 2/28/22. Siemens will provide a software update for all affected systems via Update Instruction AX003/22/S and AX004/22/S. Letter states reason for recall, health risk and action to take: We do not consider it necessary to re-examine any patients in relation to the issues (issue 1 to 5) described above. If measurements have already been performed in the past for diagnostics, please verify the results and diagnostic evaluation if applicable. Siemens will correct the software error via Update Instruction AX003/22/S and Update Instruction AX004/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein. If you have received this notification by email via Adobe EchoSign, kindly sign with your digital signature to acknowledge that you have read and understand the content provided. If you have received this notification via Certified Mail, please provide at least two contact names and email addresses of the appropriate personnel within your facility who should receive these notifications electronically. This information will assist in providing you with these important notifications in a more expedient and compliant manner. Please include your Siemens system names, material numbers, serial numbers, and Siemens Functional Location numbers and email this information to: recallsandrefusals.team@siemens-hea
Recall details
- Recall number
- Z-0985-2022
- Status
- Open, Classified
- Initiated by the company
- Feb 25, 2022
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 94 U.S.
- Product code
- OWB
- FDA submissions
- K193326
- Distribution
- US Nationwide distribution.
Product description
Artis icono biplane- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327600
Code information (models, lots, serial numbers)
S/N: 180062 180054 180019 180065 180056 180057 180041 180421 180317 180034 180072 180349 180367 180382 180060 180337 180070 180048 180083 180369 180370 180371 180372 180345 180315 180343 180080 180301 180305 180067 180330 180040 180379 180386 180360 180333 180029 180352 180336 180407 180412 180400 180397 180316 180068 180363 180084 180418 180334 180018 180350 180376 180354 180046 180047 180335 180085 180035 180339 180357 180358 180361 180362 180081 180423 180331 180411 180414 180381 180021 180342 180321 180309 180356 180378 180393 180368 180413 180341 180323 180329 180043 180049 180377 180077 180086 180053 180055 180015 180014 180066 180082 180037 180373 UDI: 04056869063317
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 89870).
- Z-0986-2022 · Open, ClassifiedArtis icono floor in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. …
- Z-0987-2022 · Open, ClassifiedArtis pheno- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use-single and biplane diagnostic imaging and interventional procedures. Model…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page