Artis pheno- in combination with a Siemens Healthineers… recallZ-0987-2022 · Siemens Medical Solutions USA, Inc · initiated Feb 25, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed
FDA root cause category: Software Design Change
What the company is doing
Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice to all affected customers via AX005/22/S on 2/28/22. Siemens will provide a software update for all affected systems via Update Instruction AX003/22/S and AX004/22/S. Letter states reason for recall, health risk and action to take: We do not consider it necessary to re-examine any patients in relation to the issues (issue 1 to 5) described above. If measurements have already been performed in the past for diagnostics, please verify the results and diagnostic evaluation if applicable. Siemens will correct the software error via Update Instruction AX003/22/S and Update Instruction AX004/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein. If you have received this notification by email via Adobe EchoSign, kindly sign with your digital signature to acknowledge that you have read and understand the content provided. If you have received this notification via Certified Mail, please provide at least two contact names and email addresses of the appropriate personnel within your facility who should receive these notifications electronically. This information will assist in providing you with these important notifications in a more expedient and compliant manner. Please include your Siemens system names, material numbers, serial numbers, and Siemens Functional Location numbers and email this information to: recallsandrefusals.team@siemens-hea
Recall details
- Recall number
- Z-0987-2022
- Status
- Open, Classified
- Initiated by the company
- Feb 25, 2022
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 20 units
- Product code
- OWB
- FDA submissions
- K201156
- Distribution
- US Nationwide distribution.
Product description
Artis pheno- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use-single and biplane diagnostic imaging and interventional procedures. Model: 10849000
Code information (models, lots, serial numbers)
Serial Number: 164719 164712 164740 164722 164763 164735 164723 164743 164741 164728 164711 164718 164739 164703 164721 164726 164724 164725 164760 UDI: 04056869046877
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ARTIS pheno 10849000Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system14 recalls64 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 89870).
- Z-0985-2022 · Open, ClassifiedArtis icono biplane- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedure…
- Z-0986-2022 · Open, ClassifiedArtis icono floor in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page