ARIA Radiation Therapy Management Software recallZ-0942-2022 · Varian Medical Systems Imaging Laboratory GmbH · initiated Mar 21, 2022

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Varian Medical Systems Imaging Laboratory GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software issue for treatment plan and image management application may result in mismatch values which could result in treatment in the wrong location

FDA root cause category: Software Design Change

What the company is doing

On 3/21/22, Varian Medical Systems issued a "Urgent Medical Device Correction" notification to affected consignees. In addition to informing consignees about the recall, Varian asked consignees to take the following actions: Please retain a copy of this document along with your most current product labeling. Advise the appropriate personnel working in your radiotherapy department of the content of this letter. This document will be posted to the Varian Medical Systems customer support website: http://www.MyVarian.com. Please complete the attached Return Response form provided. Kindly return the completed form to [email protected].

Recall details

Recall number
Z-0942-2022
Status
Completed
Initiated by the company
Mar 21, 2022
Recalling firm
Varian Medical Systems Imaging Laboratory GmbH
Quantity
14
Product code
IYE
FDA submissions
K133572, K173838
Distribution
Worldwide distribution - US Nationwide distribution in the states of MI, KS, OH, VA, DC, FL, PA, OK, MN, AZ and the countries of South Korea, Netherlands, Poland, China.

Product description

ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0, 16.1

Code information (models, lots, serial numbers)

DI Numbers: 00856100006066, 00856100006080, 00856100006097, 00856100006110, 00856100006127,00856100006165

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K133572, K173838 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.