Skip to content

ARIA Radiation Therapy Management (RTM) 16.1Recalls & reported problems

Varian Medical Systems, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00856100006165), received between Feb 16, 2022 and Jan 2, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Compatibility Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K173838, K221408); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Compatibility Problem1
  • Computer Software Problem1
  • Improper or Incorrect Procedure or Method1
  • No Apparent Adverse Event1

Patient problems

  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 2, 2024No Apparent Adverse Event; No Apparent Adverse Event

    ARIA SERVER AND BACKUP TAPE DRIVE SUFFERED WATER DAMAGE AND ARE DEFECTIVE. IT WAS SEEN CONDENSATION INSIDE THE SMALL LCD SCREEN OF THE SERVER'S FRONT PANEL, SERVER WAS DISCONNECTED. IT WAS NOTICED, THAT IN CD DRIVE, SSDS AND TAPE BACKUPS WAS WATER INSIDE AS WELL. FROM THE PHYSICAL INSPECTION OF THE SERVER, IT WAS CONCLUDED THAT IT HAD SHORT CIRCUITED DUE TO MOISTURE FROM THE FAULTY AC UNIT AND IS UNRECOVERABLE, AND THE SITE WOULD THEREFORE NEED A NEW REPLACEMENT SERVER. SYSTEM WENT DOWN ON 4TH DEC 2023 DUE TO WATER ON THE SERVER, THIS CAUSED TREATMENT DELAY FOR AT LEAST 2 WEEKS. THE CUSTOMER REFUSED TO GIVE US ANY PATIENT DATA.

    Patient problems: Insufficient Information

    MAUDE report 18426624 ↗
  • MalfunctionFeb 16, 2022Computer Software Problem; Improper or Incorrect Procedure or Method; Compatibility Problem; Computer Software Problem; Improper or Incorrect Procedure or Method; Compatibility Problem

    PER INFORMATION PROVIDED BY THE CUSTOMER, A PATIENT WAS BEING TREATED FOR LUNG CANCER TO A DOSE OF 60 GY IN 30 FRACTIONS USING PROTONS ON A MEVION HYPERSCAN 250I, TREATMENT PLANNING ON RAYSTATION AND WITH ARIA 16.1 FOR RECORDING AND VERIFICATION. ON FRACTION 4, THE TABLE POSITION OF THE SETUP FIELD WAS ERRONEOUSLY CHANGED BY THE USER AFTER TREATMENT APPROVAL WITHOUT ALSO UPDATING THE TREATMENT POSITION. A TOTAL OF 3 FRACTIONS WERE TREATED CORRECTLY BUT 20 WERE TREATED INCORRECTLY. THIS RESULTED IN A TREATMENT SIGNIFICANTLY INFERIOR TO THE TARGET WITH 20 FRACTIONS BEING DELIVERED TO NORMAL LUNG, MEDIASTINUM, HEART, ETC. WHEN THE MISADMINISTRATION WAS NOTICED, THE PATIENT'S PLAN WAS CONVERTED TO A PHOTON PLAN AND WAS DELIVERED USING ADAPTIVE RADIATION FOR 27 FRACTIONS SINCE THE 20 MIS-ADMINISTERED FRACTIONS DID NOT TOUCH THE TUMOR.

    Patient problems: Insufficient Information

    MAUDE report 13543300 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K173838, K221408. One submission often covers many models, so check the product and lot details in each record.

  • Z-0942-2022 · Mar 21, 2022 · Completed
    ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0, 16.1

Other Accelerator, linear, medical devices

All Accelerator, linear, medical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page