EVIS EXERA BRONCHOFIBERVIDEOSCOPE recallZ-1048-2022 · Olympus Corporation of the Americas · initiated Mar 8, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Bronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using sterilization

FDA root cause category: Other

What the company is doing

Olympus issued Urgent Medical Device Correction Letters on 3/8/22. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory for the referenced devices and identify any device with the model names specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Carefully read the content of this Medical Device Correction Action as well as the attached Addendum . The addendum provides updated information or instructions on the chemicals to be used for reprocessing, the sterilization conditions, rinsing after high-level disinfection, and a recommendation on using sterilization. 3. Ensure all personnel are completely knowledgeable and thoroughly trained on the new reprocessing steps. 4. Additional copies of the new Operation Manuals can be obtained by accessing our OlympusConnect site and downloading a copy of the new Manuals. Please visit our OlympusConnect customer website: at https://www.OlympusConnect.com. New users will need to register. Once registered select [Product Support] on the left navigation bar, then select [Instruction Manuals], locate the BF-XT160, BF-XP160F, BF-MP160F and BF-3C160 and select the [Download] option. 5. Access the Olympus portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the reference number 0405 and provide your contact information as indicated in the portal. 6. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. Contact our Technical Assistance Center (TAC) at 1-800-848-9024, option 1. The TAC department can assist in answering questions on BF-XT160, BF-XP160F, BF-MP160F and BF-3C160 bronchoscope cleaning and reprocessing. " Olympus America has dedicated field personnel, called Endoscopy Support Specialists, who visit customer sites to assess and observe customer reprocessing methods and to

Recall details

Recall number
Z-1048-2022
Status
Open, Classified
Initiated by the company
Mar 8, 2022
Recalling firm
Olympus Corporation of the Americas
Quantity
292 units
Product code
EOQ
FDA submissions
K033225
Distribution
US Nationwide distribution.

Product description

BF-MP160F: EVIS EXERA Bronchofibervideoscope

Code information (models, lots, serial numbers)

All serial numbers. UDI: 04953170289064

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 89942).