EVIS EXERA BRONCHOFIBERVIDEOSCOPE OLYMPUS BF TYPE MP160F (UDI 04953170289064)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA BRONCHOFIBERVIDEOSCOPE (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170289064), received between Mar 11, 2020 and Apr 14, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Crack” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 27, 2023Class IIOpen, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2491-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  2. Mar 8, 2022Class IIOpen, Classified

    Bronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using sterilization

    Z-1048-2022 · Root cause: Other · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Infection1
  • Viral Infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2026Crack

    THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF IMAGE WITH SCRATCH. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, CHIPPED OBJECTIVE LENS WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24872920 ↗
  • InjuryMar 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE SERVICE CENTER WAS INFORMED THAT FOUR PATIENTS CONTRACTED A CYTOMEGALOVIRUS INFECTION AFTER UNDERGOING A PROCEDURE WITH THE FACILITY¿S BRONCHOSCOPES. THE USER FACILITY REPORTED THAT THESE WERE INPATIENTS IN THE FACILITY¿S BURN OR RESPIRATORY CARE UNIT. THE PATIENT SAMPLES WERE SENT AN INDEPENDENT LABORATORY AND CULTURED POSITIVE. AS A RESULT, THE USER FACILITY ACQUIRED A SAMPLE OF THE DISINFECTANT FLUSHED FROM THE SCOPE; HOWEVER, THE CULTURE RESULTS ARE UNKNOWN AT THIS. THE SAMPLES FROM THE SCOPES ARE BEING RETAINED AT THE FACILITY AND REFRIGERATED. THE SCOPES WERE TAKEN OUT SERVICE AT THAT TIME; HOWEVER MAY HAVE BEEN PUT BACK INTO SERVICE DUE TO LIMITED INVENTORY. THE PATIENT¿S COURSE OF TREATMENT IS UNKNOWN. THIS IS 7 OF 8 REPORTS.

    Patient problems: Viral Infection

    MAUDE report 9816434 ↗
  • InjuryMar 11, 2020Adverse Event Without Identified Device or Use Problem

    THE SERVICE CENTER WAS INFORMED THAT FOUR PATIENTS CONTRACTED A CYTOMEGALOVIRUS INFECTION AFTER UNDERGOING A PROCEDURE WITH THE FACILITY¿S BRONCHOSCOPES. THE USER FACILITY REPORTED THAT THESE WERE INPATIENTS IN THE FACILITY¿S BURN OR RESPIRATORY CARE UNIT. THE PATIENT SAMPLES WERE SENT AN INDEPENDENT LABORATORY AND CULTURED POSITIVE. AS A RESULT, THE USER FACILITY ACQUIRED A SAMPLE OF THE DISINFECTANT FLUSHED FROM THE SCOPE; HOWEVER, THE CULTURE RESULTS ARE UNKNOWN AT THIS. THE SAMPLES FROM THE SCOPES ARE BEING RETAINED AT THE FACILITY AND REFRIGERATED. THE SCOPES WERE TAKEN OUT SERVICE AT THAT TIME; HOWEVER MAY HAVE BEEN PUT BACK INTO SERVICE DUE TO LIMITED INVENTORY. THE PATIENT¿S COURSE OF TREATMENT IS UNKNOWN. THIS IS 8 OF 8 REPORTS. AS PART OF OUR INVESTIGATION OF THIS REPORT, AN OLYMPUS ENDOSCOPY SUPPORT SPECIALIST (ESS) WAS DISPATCHED TO THE USER FACILITY TO ASSESS THEIR REPROCESSING PRACTICES AND TO PROVIDE REPROCESSING TRAINING IF NECESSARY. TO DATE, THE ESS VISIT HAS NOT BEEN FINALIZED. THE SCOPE IS BEING RETAINED AT THE USER FACILITY AT THIS TIME. AN INVESTIGATION IS ONGOING TO OBTAIN ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT. THIS REPORT IS TO ACCOUNT FOR PATI

    Patient problems: Unspecified Infection

    MAUDE report 9816412 ↗