Skip to content

BioPlex 2200 recallZ-1096-2022 · Bio-Rad Laboratories, Inc. · initiated Apr 14, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bio-Rad Laboratories, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Performance concerns for the CMV IgM analyte. Analyte has demonstrated variable elevated background depending on the handling of the reagent pack.

FDA root cause category: Under Investigation by firm

What the company is doing

On 04/14/2022, Bio-Rad Laboratories sent an "URGENT: Medical Device Recall" letter via FedEx to customers informing them that Bio-Rad has identified a performance concern for the CMV IgM analyte in the BioPlex 2200 ToRC IgM kit lot 301388. The CMV IgM analyte has demonstrated variable elevated background depending on the handling of the reagent pack. Agitation of the conjugate within the reagent pack has demonstrated a potential for both false negative and false positive results for the CMV IgM assay. QC failures may prevent the release of incorrect results; however, QC may not detect all occurrences of the issue. Customers are instructed to immediately discontinue use and to dispose of any unused product of Lot No. 301388. If the affected lot has been further distributed or transferred the product to other Bio-Rad customers, or labs, these customer/labs should be notified of this field action/recall. For questions or further assistance, contact local Bio-Rad Technical Support.

Recall details

Recall number
Z-1096-2022
Status
Open, Classified
Initiated by the company
Apr 14, 2022
Recalling firm
Bio-Rad Laboratories, Inc.
Quantity
348 units
Product code
LIP
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AR, AZ, CA, FL, HI, IA, IL, MI, MN, NE, NY, TN, TX, and WA. The countries of Czech Republic, France, Germany, and Italy.

Product description

BioPlex 2200 REF 12000670 ToRC IgM Reagent Pack

Code information (models, lots, serial numbers)

Model/Catalog Number: 12000670 UDI-DI (GTIN) Code: 03610520992715 Lot Number: 301388

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page