Skip to content

URETERO-RENO VIDEOSCOPE recallZ-1116-2022 · Olympus Corporation of the Americas · initiated Mar 25, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.

FDA root cause category: Process change control

What the company is doing

On March 25, 2022, the firm contacted the affected consignee to inform them that certain manufacturing changes by a supplier led to the possibility that a part within the bending section may crack. Olympus repaired the affected URF-V2 unit free of charge.

Recall details

Recall number
Z-1116-2022
Status
Open, Classified
Initiated by the company
Mar 25, 2022
Recalling firm
Olympus Corporation of the Americas
Quantity
1
Product code
NWB
FDA submissions
K172246
Distribution
US Nationwide distribution in the state of AZ.

Product description

Olympus Uretero-reno videoscope, Model No. URF-V2

Code information (models, lots, serial numbers)

1 unit, serial number 2825505F UDI 04953170343582

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates