Skip to content

URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of URETERO-RENO VIDEOSCOPE (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus URF-V2 videoscope is used to look inside the urinary tract. Many of its 394 adverse event reports describe parts coming loose or breaking, and the reports list 17 injuries and no deaths. There are 2 open recalls, for scopes that may be returned with water left in the channel and for a bending section that may crack. Reports rose from 19 in the prior 12 months to 198 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
397
Deaths reported
0
Injuries reported
17
Recalls naming it
2

Summary

The FDA's MAUDE database contains 397 adverse event reports that list this exact device (DI 04953170343582), received between Apr 23, 2019 and Sep 30, 2026 — classified as 17 injuries, 380 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (80 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 3 reportsJan 2024: 2 reportsFeb 2024: 6 reportsMar 2024: 3 reportsApr 2024: 5 reportsMay 2024: 7 reportsJun 2024: 7 reportsJul 2024: 4 reportsAug 2024: 1 reportsSep 2024: 4 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 6 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 6 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 9 reportsDec 2025: 23 reportsJan 2026: 50 reportsFeb 2026: 21 reportsMar 2026: 7 reportsApr 2026: 19 reportsMay 2026: 24 reportsJun 2026: 13 reportsJul 2026: 14 reportsAug 2026: 15 reportsSep 2026: 6 reportsOct 2026: 0 reports20242025202650
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0348-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Mar 25, 2022Class IIOpen, Classified

    The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.

    Z-1116-2022 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component80
  • Break72
  • Peeled/Delaminated28
  • Material Protrusion/Extrusion25
  • Failure to Clean Adequately24
  • Material Split, Cut or Torn24
  • Material Deformation22
  • Degraded19

Patient problems

  • No Clinical Signs, Symptoms or Conditions361
  • No Patient Involvement9
  • No Consequences Or Impact To Patient5
  • Foreign Body In Patient4
  • Insufficient Information3
  • Perforation3
  • Urinary Tract Infection3
  • Device Embedded In Tissue or Plaque2

Recent adverse event reports

The 25 most recent of 397 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Peeled/Delaminated; Material Split, Cut or Torn

    The customer returned the Uretero-Reno Fiberscope for a separate issue. During the device analysis, the Light Guide tube coating was peeling off. There was no patient involvement, and no harm was reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941892 ↗
  • MalfunctionSep 29, 2026Crack

    The customer returned the Uretero Reno Videoscope for a separate issue. No patient involvement. During the initial evaluation of the device, it was observed that the adhesive was chipped at the bending section.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927993 ↗
  • MalfunctionSep 28, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50924942 ↗
  • MalfunctionSep 23, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870584 ↗
  • MalfunctionSep 3, 2026Crack; Detachment of Device or Device Component

    Patient problems: Insufficient Information

    MAUDE report 50594671 ↗
  • MalfunctionSep 3, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50592022 ↗
  • MalfunctionAug 24, 2026Detachment of Device or Device Component

    The customer returned the Uretero-Reno Videoscope to the service center for evaluation related to a separate issue. During testing, the service repair technician identified the device had chipped bending section adhesive. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50414983 ↗
  • MalfunctionAug 21, 2026Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50372516 ↗
  • MalfunctionAug 19, 2026Degraded; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50333623 ↗
  • MalfunctionAug 18, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50301723 ↗
  • MalfunctionAug 18, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50301532 ↗
  • MalfunctionAug 13, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50240349 ↗
  • MalfunctionAug 12, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50218187 ↗
  • MalfunctionAug 12, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50218013 ↗
  • MalfunctionAug 11, 2026Crack; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50216437 ↗
  • MalfunctionAug 6, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50167283 ↗
  • MalfunctionAug 6, 2026Material Fragmentation

    The customer returned the Uretero-Reno Videoscope to the service center for a separate issue. During the device analysis, it was indicated that a chipped bending section glue was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50161254 ↗
  • MalfunctionAug 5, 2026Loss of or Failure to Bond

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50150027 ↗
  • MalfunctionAug 5, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50143333 ↗
  • MalfunctionAug 5, 2026Contamination; Obstruction of Flow; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50143193 ↗
  • MalfunctionAug 5, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50136662 ↗
  • MalfunctionJul 28, 2026Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50060197 ↗
  • MalfunctionJul 28, 2026Contamination /Decontamination Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50056027 ↗
  • MalfunctionJul 27, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50036950 ↗
  • MalfunctionJul 27, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50012640 ↗

Other Ureteroscope and accessories, flexible/rigid devices

All Ureteroscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page