TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx recallZ-1130-2022 · Invacare Corporation · initiated Mar 30, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Invacare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to the wheelchair. The disconnection of the battery harness results in loss of power to the wheelchair.
FDA root cause category: Device Design
What the company is doing
A preliminary notification was sent to providers on March 30, 2022, introducing the product issue and providing instructions regarding how to handle impacted units before field correction kits are available. A notification letter dated April 13, 2022, was sent to impacted consignees via traceable mail. This mailing includes a Provider Letter which will introduces the field correction requirements and provides instructions for acknowledging awareness and understanding of the requested field correction activities. Providers should inform affected customers to inform them that their wheelchair is subject to recall and requires repair. The letter provides user precautions that must be taken until the wheelchair is repaired. If you have any questions concerning this recall, please call Invacare s field correction support line at (877) 413-6008
Recall details
- Recall number
- Z-1130-2022
- Status
- Open, Classified
- Initiated by the company
- Mar 30, 2022
- Recalling firm
- Invacare Corporation
- Quantity
- 133 devices
- Product code
- ITI
- FDA submissions
- K170507
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AR, CA, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA, and WI. The country of Canada.
Product description
Invacare TDX SP2, Model Nos. TDXSP2 TDXSP2-MCG TDXSP2V
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
UDI (DI): Model Number Device Identifier TDXSP2 00841447100911 TDXSP2-MCG 00841447100966 TDXSP2V 00841447101512 Serial Numbers: 22CE000031 22BE004288 22BE002669 22BE003672 22BE004782 22BE004151 22CE000984 22BE004286 22BE004171 22BE004540 22CE001086 22BE003661 22CE000184 22CE000878 22CE001449 22CE000032 22CE001421 22BE004933 22CE001002 22CE000180 22CE000725 22BE004544 22BE005009 22BE003671 22BE004284 22BE004159 22BE004285 22BE002652 22BE004282 22BE004798 22BE004281 22BE004170 22CE000182 22CE000183 22CE000876 22CE000873 22CE000981 22CE000870 22CE000871 22CE001248 22CE001004 22CE001084 22BE004776 22BE004165 22BE004931 22BE005010 22CE000875 22CE000728 22CE000727 22BE004910 22BE004904 22BE005007 22BE004160 22BE004169 22BE004175 22BE004166 22BE004905 22CE001420 22CE000496 22BE003645 22BE002651 22CE000999 22CE001078 22BE004545 22BE004763 22BE003662 22BE004929 22BE004937 22BE004176 22BE004903 22CE000181 22BE004167 22BE005006 22BE002666 22BE004177 22BE003962 22BE003957 22BE004173 22BE003664 22BE003659 22BE004767 22BE004940 22BE004777 22BE004538 22CE000186 22CE000867 22BE002935 22BE003959 22CE000493 22CE001077 22BE002934 22CE000726 22BE005004 22CE001251 22BE003961 22BE003673 22BE004902 22BE002936 22BE004801 22BE004287 22BE004995 22BE004539 22CE001075 22CE000490 22CE000879 22BE004168 22CE000847 22CE000033 22CE001005 22CE001006 22BE004765 22BE004768 22BE004769 22BE004770 22BE004911 22BE002932 22BE003960 22CE000874 22CE001253 22BE004930 22CE000178 22CE000179 22CE001080 22BE004174 22BE003963 22BE004935 22BE004775 22BE004172 22CE000177 22BE004543 22BE004764 22BE004766 22CE001076
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx TDXSP2-MCGInvacare Corporation · Wheelchair, powered2 recalls2 reports
- TDX SP2 with LiNX Electronics TDXSP2Invacare Corporation · Wheelchair, powered2 recalls0 reports
- Invacare® TDX SP2 with Captain's Seat TDXSP2VInvacare Corporation · Wheelchair, powered2 recalls0 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K170507 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- TDX SP2 Base for Single Act. System w/Ultra Low Maxx TDXSP2-CGInvacare Corporation · Wheelchair, powered1 recall0 reports
- TDX SP2 Base for Single Act. System w/ Ultra Low Maxx TDXSP2X-CGInvacare Corporation · Wheelchair, powered1 recall0 reports
- TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx TDXSP2X-MCGInvacare Corporation · Wheelchair, powered1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page