Centricity recallZ-1229-2022 · GE Healthcare, LLC · initiated Apr 28, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on lossy images that are scaled down from their original resolution.
FDA root cause category: Software design
What the company is doing
GE Healthcare notified consignees on about 04/28/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION." The letter outlined two safety issues and instructed customers to continue to use the device in accordance with the User Manuals, but for issue #1, do not rely on measurements displayed in the viewer. They should manually calibrate the image to create a measurement calibration reference and then perform necessary measurements. For issue #2, do not perform measurements on lossy images. In both instances, customers were requested to complete and return the acknowledgment form to [email protected]. A second letter was sent to customers on about 02/24/2023, to provide instructions for downloading the user manual addendum for CWeb.
Recall details
- Recall number
- Z-1229-2022
- Status
- Open, Classified
- Initiated by the company
- Apr 28, 2022
- Posted by the FDA
- Jun 3, 2022
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 488 units
- Product code
- LLZ
- FDA submissions
- K131977
- Distribution
- Worldwide distribution. Nationwide distribution to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International distribution to Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Bolivia, Plurinational State of, Brazil, Canada, Chile, China, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Korea, South, Kuwait, Lebanon, Malaysia, Malta, Mexico, Namibia, Netherlands, Nigeria, Norway, Pakistan, Panama, Peru, Poland, Portugal, Puerto Rico, Qatar, Russia, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Bolivarian Republic of, Vietnam
Product description
Centricity Universal Viewer Zero Footprint Client
Code information (models, lots, serial numbers)
Software Version 6.0 SP11 through 6.0 SP11.4; GTIN/DI: 00840682102988; Codes: 2089507-1XX
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 90136).
- Z-1231-2022 · Open, ClassifiedCentricity Radiology RA600
- Z-1232-2022 · Open, ClassifiedCentricity Cardiology CA1000
- Z-1233-2022 · Open, ClassifiedCentricity PACS RA1000
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page