Skip to content

URETERO-RENO FIBERSCOPE recallZ-1263-2022 · Olympus Corporation of the Americas · initiated Apr 20, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Olympus notified Urgent Medical Device Correction letter on 4/20/22 via email. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. The model and serial number can be found on the device as illustrated. 2. An Olympus representative will contact you to make arrangements for return of your affected device. Olympus will repair your unit free of charge. In the meantime, follow the instruction manual for inspection of the device before use. Should any irregularities be identified, including a loose parts, do not use the device and contact Olympus for repair. 3. Olympus will arrange for a service loaner in an effort not to disrupt clinical care. Contact me directly at 647-999-3203 or at [email protected] for any additional information or support concerning this matter.

Recall details

Recall number
Z-1263-2022
Status
Open, Classified
Initiated by the company
Apr 20, 2022
Recalling firm
Olympus Corporation of the Americas
Quantity
2 units
Product code
FGB
FDA submissions
K172298
Distribution
CA FL MA MD MI NJ OH PR WI

Product description

Olympus URETERO-RENO FIBERSCOPE Model: URF-P6

Code information (models, lots, serial numbers)

UDI-DI: 04953170340802 Serial Number: 2836210 2834747I

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 90195).

Sources and updates