Skip to content

URETERO-RENO FIBERSCOPE OLYMPUS URF-P6Recalls & reported problems

Olympus Medical Systems Corp. · "URF-P6 SUPER-SLIM FLEXIBLE FIBEROPTIC UR" · Class 2

All models of URETERO-RENO FIBERSCOPE (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus URF-P6 ureteroscope is a flexible fiberscope used to examine the urinary tract. Of its 171 reports, 25 are filed as injuries, and many describe breaks, detached parts, and peeling covers. The scope has 2 open recalls, covering wet channels after repair and an adhesive error that can blur vision. The reports mention no deaths. Reports rose over the past 12 months, with 52 compared with 17 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
170
Deaths reported
0
Injuries reported
25
Recalls naming it
2

Summary

The FDA's MAUDE database contains 170 adverse event reports that list this exact device (DI 04953170340802), received between Oct 18, 2018 and Aug 12, 2026 — classified as 25 injuries, 145 malfunctions.

The most frequently reported specific device problem is “Break” (29 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 5 reportsJan 2026: 11 reportsFeb 2026: 5 reportsMar 2026: 2 reportsApr 2026: 10 reportsMay 2026: 8 reportsJun 2026: 5 reportsJul 2026: 1 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0347-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Apr 20, 2022Class IIOpen, Classified

    Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

    Z-1263-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break29
  • Detachment of Device or Device Component23
  • Adverse Event Without Identified Device or Use Problem20
  • Scratched Material11
  • Microbial Contamination of Device9
  • Peeled/Delaminated9
  • Degraded8
  • Failure to Clean Adequately8

Patient problems

  • No Clinical Signs, Symptoms or Conditions138
  • No Patient Involvement6
  • Bacterial Infection4
  • Fever4
  • Hemorrhage/Blood Loss/Bleeding4
  • Pain4
  • Unspecified Kidney or Urinary Problem4
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 25 most recent of 170 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2026Corroded

    The videoscope was returned to the service center with a report of a leak. This does not indicate an MDR reportable event. However, MDR reportable failure was found during testing at the service center. During inspection of the device, it was found that the distal end plastic cover was rusty. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50223011 ↗
  • MalfunctionAug 12, 2026Material Fragmentation; Peeled/Delaminated

    The customer returned the Uretero-Reno Fiberscope to the service center for a leak and image issue which did not indicate an MDR reportable event. During the evaluation of the device, it was found that the adhesive was lifted and chipped. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50217444 ↗
  • MalfunctionAug 10, 2026Corroded

    The customer returned the Uretero-Reno Videoscope to the service center for a separate issue. During the device analysis, it was indicated that the distal end was rusty. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50200630 ↗
  • MalfunctionJul 21, 2026Loss of or Failure to Bond

    THE CUSTOMER RETURNED THE URETERO-RENO FIBERSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING TESTING, THE SERVICE REPAIR TECHNICIAN IDENTIFIED THE A-RUBBER GLUE WAS CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25888031 ↗
  • MalfunctionJun 30, 2026Detachment of Device or Device Component

    THE CUSTOMER RETURNED THE URETERO-RENO FIBERSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THAT A PEELED BENDING SECTION COVER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25670529 ↗
  • MalfunctionJun 26, 2026Detachment of Device or Device Component

    THE CUSTOMER RETURNED THE URETERO-RENO FIBERSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT THE BENDING SECTION COVER GLUE IS CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25635369 ↗
  • MalfunctionJun 18, 2026Failure to Clean Adequately

    THE FIBERSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF A SMASHED DISTAL TIP. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, IT WAS FOUND THAT THE DISTAL END CONTAINED RESIDUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25553289 ↗
  • MalfunctionJun 5, 2026Detachment of Device or Device Component

    THE CUSTOMER RETURNED THE URETERO-RENO FIBERSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT CHIPPED BENDING SECTION COVER GLUE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25410752 ↗
  • MalfunctionJun 2, 2026Detachment of Device or Device Component

    THE CUSTOMER RETURNED THE URETERO-RENO FIBERSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE REPAIR TECHNICIAN INDICATES THAT CHIPPED RUBBER GLUE WAS FOUND. IT WAS DETERMINED THAT THE RUBBER GLUE NEEDED TO BE REPLACED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25365831 ↗
  • MalfunctionMay 29, 2026Loss of or Failure to Bond

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, IT WAS INDICATED THAT A CHIPPED ADHESIVE ON THE BENDING SECTION COVER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25340026 ↗
  • MalfunctionMay 28, 2026Stretched

    THE URETERO-RENO FIBERSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A SHEATH THAT WAS PEELING OFF. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, A STRETCHED BENDING SECTION COVER WAS FOUND TO BE MOST LIKELY DUE TO DEFORMATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25317869 ↗
  • MalfunctionMay 27, 2026Peeled/Delaminated

    THE URETERO-RENO FIBERSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF FAILED A LEAK TEST. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, THE BENDING SECTION HAS LIFTED PIN WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25302378 ↗
  • MalfunctionMay 22, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERO-RENO FIBERSCOPE EXHIBITED CHIP ON A-RUBBER GLUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25256395 ↗
  • MalfunctionMay 20, 2026Detachment of Device or Device Component

    THE URETERO-RENO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF THERE'S BLACK SPOTS IN THE IMAGE THINK THERE ARE BROKEN FIBERS. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, CHIPPED ADHESIVE OF THE BENDING RUBBER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25230454 ↗
  • MalfunctionMay 19, 2026Break

    THE URETERO-RENO FIBERSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF FULL OF BLACK SPOTS. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, BENDING SECTION HAD BROKEN SKELETON WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25219871 ↗
  • MalfunctionMay 14, 2026Detachment of Device or Device Component

    THE URETERO-RENO FIBERSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF A FAILED LEAK TEST. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, PEELED ADHESIVE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25173800 ↗
  • MalfunctionMay 5, 2026Detachment of Device or Device Component

    THE URETERO-RENO FIBERSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF A CHIPPED BENDING SECTION COVER GLUE. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, THE CHIPPED GLUE WAS FOUND TO BE MOST LIKELY DUE TO STRESS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25086243 ↗
  • MalfunctionApr 29, 2026Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERO-RENO FIBERSCOPE BENDING SECTION HAD A LIFTED PIN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25027691 ↗
  • MalfunctionApr 28, 2026Contamination

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERO-RENO FIBERSCOPE HAD AN UNKNOWN SUBSTANCE ON THE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25016160 ↗
  • MalfunctionApr 23, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERO-RENO FIBERSCOPE HAD A CHIPPED ADHESIVE ON THE BENDING SECTION COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24966144 ↗
  • MalfunctionApr 17, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE UETERO-RENO FIBERSCOPE HAD CHIPPED RUBBER GLUES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912257 ↗
  • MalfunctionApr 15, 2026Material Fragmentation; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERO RENO FIBERSCOPE EXHIBITED CHIPPED AND LIFTING A RUBBER GLUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24887325 ↗
  • MalfunctionApr 10, 2026Degraded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERO RENO FIBERSCOPE EXHIBITED PEELED ADHESIVE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24839587 ↗
  • MalfunctionApr 9, 2026Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FIBERSCOPE'S ADHESIVE HAD PEELED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24835723 ↗
  • MalfunctionApr 8, 2026Degraded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERO-RENO FIBERSCOPE EXHIBITED A RUST AROUND THE SEAM OF THE DISTAL END PLASTIC COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24818418 ↗

FDA clearance and classification

The FDA regulates product code FGB (ureteroscope and accessories, flexible/rigid) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the URETERO-RENO FIBERSCOPE OLYMPUS URF-P6

Has the URETERO-RENO FIBERSCOPE OLYMPUS URF-P6 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0347-2024 (Class II), began on Sep 25, 2023: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the URETERO-RENO FIBERSCOPE OLYMPUS URF-P6?

The FDA's MAUDE database lists 170 adverse event reports for this device, received between Oct 18, 2018 and Aug 12, 2026. In those reports, the most common specific device problems are “Break” (29), “Detachment of Device or Device Component” (23) and “Scratched Material” (11); the most common patient problems are “Bacterial Infection” (4), “Fever” (4) and “Hemorrhage/Blood Loss/Bleeding” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the URETERO-RENO FIBERSCOPE OLYMPUS URF-P6?

None of the reports are classified as deaths, 25 as injuries and 145 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the URETERO-RENO FIBERSCOPE OLYMPUS URF-P6, and is it still sold?

The FDA's device database lists Olympus Medical Systems Corp. as the company responsible for it (brand URETERO-RENO FIBERSCOPE, model OLYMPUS URF-P6). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Ureteroscope and accessories, flexible/rigid devices

All Ureteroscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page