ABACUS recallZ-1376-2022 · Baxter Healthcare Corporation · initiated Jun 22, 2022
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.
FDA root cause category: Software design
What the company is doing
Baxter sent an Urgent Medical Device Correction notice on 06/22/2022 by letter. The notice explained the problem and the hazard and recommended the following: customers refrain from making updates to bag label templates and to contact Baxter Technical Support if updates to bag label templates are needed. Please continue to follow the instructions in the Abacus User and Abacus Configuration Guides when creating an order, and have a pharmacist thoroughly review all order outputs including the printed label for accuracy and safety. As mentioned above, please ensure your facility processes include pharmacy checks during the compounding process as well as a nursing check. Baxter will communicate to customers when the software upgrade becomes available. At that time additional instructions to obtain the software upgrade will be provided.
Recall details
- Recall number
- Z-1376-2022
- Status
- Terminated
- Initiated by the company
- Jun 22, 2022
- Posted by the FDA
- Jul 22, 2022
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 9 units
- Product code
- NEP
- Distribution
- United States (including Puerto Rico), Argentina, Bermuda, Chile, Colombia, Dominican Republic, Panama, Canada, Vietnam, Hong Kong, Singapore, and EMEA
Product description
ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE), 8300-0168 (Abacus V3.1 SE), 8300-0169 (Abacus V3.1 ME), 8300-0191 (Abacus V3.2 CE), 8300-0192 (Abacus V3.2 SE), 8300-0193 (Abacus V3.2 ME), 8300-3391 (Abacus V3.3 CE), 8300-3392 (Abacus V3.3 SE), and 8300-3393 (Abacus V3.3 ME). Commonly used with ExactaMix automated compounding devices (ExactaMix 1200 and ExactaMix 2400).
Code information (models, lots, serial numbers)
Product Code 8300-3391, UDI (01)05413765577345(10)3.3.2.1, Release date 20-Sep-2019; Product Code 8300-3392, UDI (01)05413765577352(10)3.3.2.1, Release date 20-Sep-2019; Product Code 8300-3393, UDI (01)05413765577369(10)3.3.2.1, Release date 20-Sep-2019; Product Codes: 8300-0167, V3.1 CE, Release date 20-May-2013, 8300-0168, V3.1 SE, Release date 20-May-2013, 8300-0169, V3.1 ME, Release date 20-May-2013, 8300-0191, V3.2 CE, Release date 17-Dec-2015, 8300-0192, V3.2 SE, Release date 17-Dec-2015, 8300-0193, V3.2 ME, Release date 17-Dec-2015, No UDIs
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- ABACUS 83003391Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING1 recall0 reports
- ABACUS 83003392Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING1 recall0 reports
- ABACUS 83003393Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.