Medical device recalls: July 2022
The FDA posted 35 medical device recall records in July 2022, from 17 recall events by 16 companies. 15 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- GE Healthcare, LLC · 2 events, 2 records, 1 Class I
- Covidien Llc · 1 event, 9 records, 9 Class I
- Philips Medical Systems DMC GmbH · 1 event, 5 records
- American Contract Systems, Inc. · 1 event, 2 records, 2 Class I
- Arcoma AB · 1 event, 2 records
- Fresenius Medical Care Holdings, Inc. · 1 event, 2 records
- LumiraDx · 1 event, 2 records
- ROi CPS LLC · 1 event, 2 records
FDA root cause
- Process control · 14
- Under Investigation by firm · 7
- Radiation Control for Health and Safety Act · 5
- Device Design · 4
- Process change control · 3
- Software design · 2
Class I recalls (15)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1376-2022 · Baxter Healthcare Corporation · posted Jul 22, 2022 · Terminated
There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.
- CereLink ICP Monitor ↗Class IZ-1395-2022 · Integra LifeSciences Corp. · posted Jul 22, 2022 · Open, Classified
Integra has received complaints associated with ICP readings drifting to -50 mmHg (out-of-range) and may manifest an error message "sensor or extension cable failure!" The firm has found that this is caused by electrical interference from a component of…
- Z-1397-2022 · Bard Access Systems, Inc. · posted Jul 22, 2022 · Open, Classified
BD Intraosseous Needle Kits may exhibit: 1) Increased force required to remove the stylet from intraosseous needle during placement may cause removal of the entire needle assembly and loss of intraosseous access; 2) Stylet safety mechanism may not engage…
- Z-1389-2022 · GE Healthcare, LLC · posted Jul 19, 2022 · Terminated
There is a potential reversal of the O2 and air cylinder pressure transducer connections.
- Z-1312-2022 · Covidien Llc · posted Jul 7, 2022 · Open, Classified
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
- Z-1313-2022 · Covidien Llc · posted Jul 7, 2022 · Open, Classified
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
- Z-1314-2022 · Covidien Llc · posted Jul 7, 2022 · Open, Classified
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
- Z-1315-2022 · Covidien Llc · posted Jul 7, 2022 · Open, Classified
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
- Z-1316-2022 · Covidien Llc · posted Jul 7, 2022 · Open, Classified
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
- Z-1317-2022 · Covidien Llc · posted Jul 7, 2022 · Open, Classified
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
- Z-1318-2022 · Covidien Llc · posted Jul 7, 2022 · Open, Classified
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
- Z-1319-2022 · Covidien Llc · posted Jul 7, 2022 · Open, Classified
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
- Z-1320-2022 · Covidien Llc · posted Jul 7, 2022 · Open, Classified
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
- Z-1276-2022 · American Contract Systems, Inc. · posted Jul 1, 2022 · Open, Classified
Affected products were assembled off-site by individuals who may not have been properly trained.
- Z-1277-2022 · American Contract Systems, Inc. · posted Jul 1, 2022 · Open, Classified
Affected products were assembled off-site by individuals who may not have been properly trained.
Class II recalls (19)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D)Class IIZ-1480-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Jul 28, 2022 · Open, Classified
Manufacturing error that may have resulted in a cathode component being out of specification. All devices met final functional testing requirements.
- AquaC UNO H [120V, 60Hz, 1700VA,Class IIZ-1463-2022 · Fresenius Medical Care Holdings, Inc. · posted Jul 25, 2022 · Open, Classified
Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device),…
- AquaC UNO H [120V, 60Hz, 1000VA,Class IIZ-1464-2022 · Fresenius Medical Care Holdings, Inc. · posted Jul 25, 2022 · Open, Classified
Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device),…
- Z-1450-2022 · LumiraDx · posted Jul 22, 2022 · Terminated
Modification to the LumiraDx SARS-CoV-2 Antibody (Ab) test that was made outside the scope of the approved EUA.
- Z-1451-2022 · LumiraDx · posted Jul 22, 2022 · Terminated
Modification to the LumiraDx SARS-CoV-2 Antibody (Ab) test that was made outside the scope of the approved EUA.
- Nonsterile convenience kits:Class IIZ-1407-2022 · ROi CPS LLC · posted Jul 19, 2022 · Terminated
Some nonsterile components that were supposed to be sterile were packaged in nonsterile kits.
- Z-1408-2022 · ROi CPS LLC · posted Jul 19, 2022 · Terminated
Some nonsterile components that were supposed to be sterile were packaged in nonsterile kits.
- Z-1325-2022 · Philips Medical Systems DMC GmbH · posted Jul 12, 2022 · Open, Classified
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board.…
- Z-1326-2022 · Philips Medical Systems DMC GmbH · posted Jul 12, 2022 · Open, Classified
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board.…
- Z-1327-2022 · Philips Medical Systems DMC GmbH · posted Jul 12, 2022 · Open, Classified
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board.…
- Z-1328-2022 · Philips Medical Systems DMC GmbH · posted Jul 12, 2022 · Open, Classified
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board.…
- Z-1329-2022 · Philips Medical Systems DMC GmbH · posted Jul 12, 2022 · Open, Classified
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board.…
- Z-1374-2022 · Ventana Medical Systems Inc · posted Jul 12, 2022 · Open, Classified
There is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire
- Z-1344-2022 · Abbott · posted Jul 7, 2022 · Terminated
When connected to the EnSite Precision Navigation System, an affected TactiCath Contact Force Ablation Catheter, Sensor Enabled may present the error message: "invalid catheter,, or "expired catheter."
- Z-1340-2022 · Siemens Healthcare Diagnostics, Inc. · posted Jul 6, 2022 · Terminated
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 g/L], may lead to inappropriate intervention for myocardial infarction
- Z-1341-2022 · Siemens Healthcare Diagnostics, Inc. · posted Jul 6, 2022 · Terminated
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 g/L], may lead to inappropriate intervention for myocardial infarction
- Z-1331-2022 · Arcoma AB · posted Jul 5, 2022 · Open, Classified
There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC).
- Z-1332-2022 · Arcoma AB · posted Jul 5, 2022 · Open, Classified
There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC).
- Z-1333-2022 · GE Healthcare, LLC · posted Jul 5, 2022 · Open, Classified
Potential to display inaccurate measurements on images in Centricity Universal Viewer Zero Footprint Client (ZFP)
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1390-2022 · R & D Systems, Inc. · posted Jul 15, 2022 · Terminated
The Instructions for Use (IFU, Part Number 750095) provided with the affected lots by the firm contains an error associated with product s CE marking. The CE mark contained in the IFU is followed by the Notified Body designation, British Standards…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.