GXL acetabular liners, orthopedic hip implant component.… recallZ-1725-2022 · Exactech, Inc. · initiated Aug 11, 2022
Class II Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.
FDA root cause category: Process control
What the company is doing
The firm sent an "URGENT MEDICAL DEVICE RECALL" DEAR HEALTHCARE PROFESSIONAL COMMUNICATION on 08/11/2022 by email. The purpose of the current letter is to update surgeons regarding the implications of the packaging non-conformity on performance of the Connexion GXL and other conventional UHMWPE acetabular liners. In addition, this letter provides updates to patient management and follow-up recommendations since issuance of the July 2021 DHCP letter. In addition to providing surgeons with lists of all their affected GXL and/or conventional polyethylene patients since 2004, Exactech is providing surgeons with two draft letters directed at patients who have been implanted with Exactech GXL hip devices. The Customers were instructed to take the following actions: CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required. IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customers inventory. EXTEND THE DESCRIPTION OF THE ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession. COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at [email protected]. Reach out to Customer Service at [email protected] to request a Returned Goods (RG) Number for ALL affected product requested to be returned. WE ARE REQUIRING 100% EFFECTIVENESS FOR THIS RECALL. If you have any questions, call (888) 912-0403.
Recall details
- Recall number
- Z-1725-2022
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Aug 11, 2022
- Recalling firm
- Exactech, Inc.
- Quantity
- 1017 devices
- Product code
- KWZ
- FDA submissions
- K040601
- Distribution
- Worldwide distribution: US (nationwide) and OUS (foreign) countries of: ARGENTINA, BRAZIL, CHILE, CHINA, COLOMBIA, ECUADOR, GREECE, GUATEMALA, INDIA, LEBANON PORTUGAL,PUERTO RICO,SPAIN
Product description
GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. 0 DEG ACUMATCH CONSTRAINED LINER SZ E 28MM ID, REF 144-28-05; b. 0 DEG ACUMATCH CONSTRAINED LINER SZ F 28MM ID, REF 144-28-06; c. 0 DEG ACUMATCH CONSTRAINED LINER SZ G 28MM ID, REF 144-28-07; d. 0 DEG ACUMATCH CONSTRAINED LINER SZ H 28MM ID, REF 144-28-08; e. 0 DEG ACUMATCH CONSTRAINED LINER SZ J 28MM ID, REF 144-28-09; f. 0 DEG ACUMATCH CONSTRAINED LINER SZ K 28MM ID, REF 144-28-10; g. 0 DEG MCS CONSTRAINED LINER 50-52 MM, 28 MM ID, REF 146-28-04; h. 0 DEG MCS CONSTRAINED LINER 54-58 MM, 28 MM ID, REF 146-28-05; i. 0 DEG MCS CONSTRAINED LINER 60-64 MM, 32MM ID, REF 146-32-06; j. O DEG ACUMATCH CONSTRAINED LINER SZ E 32MM ID, REF 144-32-05; k. O DEG ACUMATCH CONSTRAINED LINER SZ F 32MM ID, REF 144-32-06; l. O DEG ACUMATCH CONSTRAINED LINER SZ G 32MM ID, REF 144-32-07; m. O DEG ACUMATCH CONSTRAINED LINER SZ H 32MM ID, REF 144-32-08; n. O DEG ACUMATCH CONSTRAINED LINER SZ J 32MM ID, REF 144-32-09; o. O DEG ACUMATCH CONSTRAINED LINER SZ K 32MM ID, REF 144-32-10;
Code information (models, lots, serial numbers)
Show the full code information (10k characters)
a. REF 144-28-05, GUIDID 10885862027405, Serial Numbers: 646995, 646996, 646997, 646998, 667989, 667990, 667992, 667995, 667996, 668002, 668003, 668004, 770434, 770436, 770440, 770442, 770443, 770444, 863939, 863940, 863943, 863944, 863946, 863948, 1242282, 1242283, 1242284, 1242286, 1242287, 1242288, 1242290, 1242291, 1242292, 1617713, 1617715, 1617717, 1617718, 1617719, 1617720, 1617721, 1869774, 1869775, 1869776, 1869777, 1907611, 1907612, 1907613, 1907614, 1961038, 1961039, 1961040, 1961041, 1961042, 2019292, 2019293, 2019294, 2019295, 2174511, 2174512, 2175232, 2314993, 2314995, 2681417, 2681418, 2681428,; b. REF 144-28-06, GUIDID 10885862027412, Serial Numbers: 647081, 647082, 647083, 647084, 647085, 656277, 656278, 656279, 656281, 667456, 667459, 667466, 667467, 667468, 667470, 667471, 667472, 667474, 667475, 667476, 667477, 736674, 736675, 736677, 736678, 736684, 736685, 736687, 736688, 736689, 736691, 736692, 736695, 736697, 736698, 736704, 736705, 736706, 736707, 736710, 736711, 736712, 736716, 736720, 736722, 1088581, 1088582, 1088583, 1088585, 1088586, 1088587, 1088588, 1088589, 1088590, 1088592, 1088593, 1088594, 1088595, 1088596, 1088597, 1088598, 1088599, 1088601, 1088602, 1088604, 1088605, 1088606, 1088607, 1088608, 1088609, 1088611, 1088612, 1088614, 1088615, 1088616, 1088618, 1088619, 1503039, 1503040, 1503041, 1503043, 1613417, 1613418, 1613419, 1613420, 1613421, 1613422, 1613424, 1613425, 1613426, 1613427, 1748075, 1748076, 1748078, 1748082, 1779388, 1779389, 1779390, 1779391, 1779392, 1779393, 1779394, 1779395, 1779396, 1779397, 1869559, 1869560, 1869561, 1869562, 1869563, 1869564, 1869565, 1869566, 1869567, 1869568, 1869569, 1919876, 1919877, 1919878, 1919879, 1919880, 1958258, 1958259, 1958260, 1958261, 1958262, 1961605, 1961606, 1961607, 1961608, 1961609, 1961610, 1961611, 1961612, 1961613, 1961614, 1961615, 2057734, 2057735, 2057736, 2057737, 2057738, 2057739, 2071335, 2076278, 2102297, 2102298, 2127781, 2127782, 2154421, 2154422, 2154423, 2242513, 2242514, 2242516, 2332952, 2418182, 2418183, 2418185, 2418186, 2418187, 2418189, 2418190, 2681365, 2681366, 2681367, 2681370, 2681371, 2681374, 2681375, 2681378, 2681379, 2681381, 2681384, 2681386, 2681389, 2681391, 5434157, 5448537, 5479670,; c. REF 144-28-07, GUIDID 10885862027429, Serial Numbers: 656453, 656454, 656455, 656456, 667163, 667167, 667173, 667177, 667178, 667179, 667181, 667182, 667184, 667186, 667187, 735322, 735323, 735325, 735326, 735327, 735329, 735330, 735331, 735332, 735334, 735335, 735337, 735339, 735340, 735341, 735342, 735344, 735345, 735346, 735347, 1255739, 1255740, 1255741, 1255742, 1255743, 1255744, 1255745, 1255746, 1255747, 1255748, 1255749, 1255750, 1255751, 1255752, 1255753, 1255754, 1539344, 1539346, 1539347, 1539348, 1552589, 1552590, 1631561, 1631562, 1631563, 1631564, 1631565, 1631566, 1631567, 1631568, 1631570, 1631571, 1631573, 1631574, 1631575, 1631576, 1631578, 1631579, 1631580, 1848625, 1848627, 1848628, 1848629, 1848630, 1848631, 1848632, 1848633, 1848634, 1848635, 1848636, 1848637, 1848638, 1848639, 1848640, 1848641, 1848642, 1848643, 1848644, 1848645, 1995515, 1995516, 1995517, 1995518, 1999705, 2008666, 2020951, 2020952, 2076037, 2116750, 2154208, 2180270, 2180272, 2180273, 2180274, 2180275, 2314476, 2314478, 2314481, 2314482, 2314483, 2681343, 2681346, 2681362, 2681363, 2681364, 5620814; d. REF 144-28-08, GUIDID 10885862027436, Serial Numbers: 646958, 656496, 667567, 667571, 667575, 667578, 667581, 735816, 735818, 735819, 735821, 735822, 735824, 735825, 735826, 735829, 735831, 735832, 735833, 735835, 735836, 735838, 735839, 735841, 735842, 1496501, 1496502, 1496503, 1496504, 1496505, 1497983, 1497985, 1497986, 1628671, 1628672, 1628673, 1628676, 1628680, 1628681, 1780295, 1780296, 1780297, 1780298, 1780299, 1780300, 1780301, 1780302, 1780303, 1780304, 1886863, 1886865, 1886866, 1886867, 1886869, 1886870, 1886871, 1886872, 1886873, 1978121, 1978122, 1978123, 1978124, 1978125, 1978126, 1978127, 1978128, 1978129, 1978130, 1978131, 2154167, 2154169, 2155086, 2155088, 2173630, 2173632, 2173636, 2173637, 2203733, 2203735, 2203736, 2203737, 2675713, 2741976, 2741978, 2741979, 2741980, 3877265, 3877269,; e. REF 144-28-09, GUIDID 10885862027443, Serial Numbers: 646927, 646928, 656808, 667271, 667274, 667278, 667280, 667286, 735649, 735650, 735652, 735657, 735658, 735659, 735660, 735661, 735663, 735665, 735667, 1501517, 1501518, 1501519, 1619068, 1619071, 1758479, 1758480, 1758482, 1758483, 1758484, 1758485, 1758487, 1758488, 1907668, 1907669, 1907670, 1907671, 1907673, 1907674, 1907675, 1907676, 2015630, 2015631, 2015632, 2015633, 2173625, 2173626, 2173627, 2173628, 2173629, 2315681, 2315686, 2315688, 2681432, 2681433, 2681435, 2681436, 2681438, 2681444, 2681449, 5448848; f. REF 144-28-10, GUIDID 10885862027450, Serial Numbers: 732949, 732957, 732958, 1501571, 1626559, 1626560, 1626564, 1766032, 1766033, 1766034, 1766036, 1766037, 1766038, 1766039, 1813495, 2314200, 2314201, 2522283, 2522284, 2675864, 2675866, 2675867, 2741969, 2741971, 2741972, 2741973; g. REF 146-28-04, GUIDID 10885862027719, Serial Numbers: 673521, 673522, 673523, 673524, 673526, 673527, 673528, 673529, 1071025, 1071031, 1071036, 1071037, 1071040, 2430266, 5092898, 5092901, 5987556; h. REF 146-28-05, GUIDID 10885862027726, Serial Numbers: 672814, 672817, 672821, 672822, 672823, 672824, 672825, 672828, 672831, 672832, 1440928, 1440929, 1440930, 1440932, 1597246, 3645387, 3731190; i. REF 146-32-06, GUIDID 10885862027771, Serial Numbers: 652352, 652353, 652355, 652359, 652364, 786425, 786432, 786438, 786447, 786456, 1501636, 1501638, 1501639, 1501640, 1788763, 1902369, 1902370, 1902371, 1902372, 1902373, 1939712, 1939713, 1939714, 1939715, 1939716, 2155789, 2205295, 2205296, 2205297, 2205298, 2205299, 2205300, 2205301, 2205302, 2205303, 2591479, 2591480, 2744018, 2744019, 2744020, 2744021, 2744022, 2744023, 2833717, 2833718, 2833719, 2833720, 2833721, 2853034, 2853035, 2853036, 2853037, 2853038, 2853039, 2853040, 2853041, 2853042, 2853043, 2853044, 2853045, 2960474, 2960476, 2960477, 2998200, 2998201, 2998202, 4116872, 4125594, 4125599, 4125600, 5983742, 6616862; j. REF 144-32-05, GUIDID 10885862027467, Serial Numbers: 667367, 750499, 750500, 750510, 750511, 1466301, 1468509, 1468510, 1760868, 1760869, 1760870, 1760871, 1760872, 1761157, 2014622, 2116742, 2178819, 2178820, 2178821, 2178822, 2408155, 2408156, 2408157, 2408158, 2408159, 2408160, 2408161, 3877259, 3952744, 4040044, 4040046; k. REF 144-32-06, GUIDID 10885862027474, Serial Numbers: 652352, 652353, 652355, 652359, 652364, 786425, 786432, 786438, 786447, 786456, 1501636, 1501638, 1501639, 1501640, 1788763, 1902369, 1902370, 1902371, 1902372, 1902373, 1939712, 1939713, 1939714, 1939715, 1939716, 2155789, 2205295, 2205296, 2205297, 2205298, 2205299, 2205300, 2205301, 2205302, 2205303, 2591479, 2591480, 2744018, 2744019, 2744020, 2744021, 2744022, 2744023, 2833717, 2833718, 2833719, 2833720, 2833721, 2853034, 2853035, 2853036, 2853037, 2853038, 2853039, 2853040, 2853041, 2853042, 2853043, 2853044, 2853045, 2960474, 2960476, 2960477, 2998200, 2998201, 2998202, 4116872, 4125594, 4125599, 4125600, 5983742, 6616862; l. REF 144-32-07, GUIDID 10885862027481, Serial Numbers: 646979, 646980, 646981, 646982, 656282, 656283, 656285, 667544, 667545, 667547, 667548, 667549, 667550, 667551, 667552, 667553, 667555, 667557, 667558, 667559, 667560, 667561, 667562, 667563, 733057, 733058, 733059, 733060, 733061, 733062, 733063, 733065, 733067, 733068, 733069, 733070, 733071, 733072, 733073, 733074, 733075, 733076, 733077, 733079, 733080, 733081, 733082, 733083, 733084, 733085, 733086, 733087, 1083038, 1083039, 1083041, 1083043, 1083044, 1083045, 1083046, 1083047, 1083048, 1083049, 1083050, 1083051, 1083052, 1083053, 1240363, 1240364, 1240365, 1240366, 1240367, 1240368, 1240369, 1240370, 1240371, 1240372, 1240373, 1637247, 1637248, 1637249, 1637250, 1637251, 1637252, 1637253, 1637255, 1637256, 1960914, 1960915, 1960916, 1960917, 1960918, 1978108, 1978109, 1978110, 1978111, 1978112, 2199477, 2199478, 2199479, 2199480, 2199481, 2209286, 2209287, 2209288, 2209289, 2209290, 2316057, 2316058, 2316060, 2316061, 2316062, 2316063, 2316065, 2316066, 2534039, 2534040, 2534042, 2534043, 2534044, 2534045, 2534046, 2534047, 2709177, 2741974, 2741975, 2760031, 2825163, 2998199, 3695205, 4033071, 4057982, 4057983, 4057984, 4057985, 4057986, 4269948, 4269949, 4269950, 4269953, 4269954, 4269955, 4269959, 4269960, 4269961, 4269963, 4269964, 4269965, 4269966, 6035701, 6035704; m. REF 144-32-08, GUIDID 10885862027498, Serial Numbers: 646984, 646985, 646987, 646988, 656490, 656491, 656492, 656493, 656494, 667530, 667531, 667533, 667534, 667535, 667536, 667537, 667538, 667539, 667542, 667543, 733870, 733872, 733873, 733874, 733875, 733877, 733879, 733881, 733882, 733883, 733884, 733887, 733888, 733889, 733891, 733892, 733893, 733894, 733895, 1099238, 1099239, 1099241, 1099242, 1099243, 1099246, 1099247, 1099248, 1597853, 1597854, 1597855, 1613660, 1613662, 1613663, 1821432, 1821433, 1821434, 1821435, 1821436, 1821437, 1821438, 2315089, 2315090, 2400213, 2400214, 2400215, 2400216, 2400217, 2400218, 2400219, 2795185, 2795186, 2795188, 2795192, 2795193, 2795194, 2795197, 2795198, 2795199, 2795200, 2795201, 2795202, 2795204, 2795209, 5448809, 5448810, 5448811, 5501762; m. REF 144-32-09, GUIDID 10885862027504, Serial Numbers: 646970, 646971, 646972, 656764, 656765, 656767, 656768, 667686, 667687, 667689, 667690, 667691, 667695, 732988, 732989, 732991, 732992, 732993, 732995, 732997, 732998, 732999, 733000, 733002, 733004, 733006, 733009, 733010, 733011, 1098128, 1098130, 1098131, 1098132, 1098133, 1098135, 1098137, 1758159, 1758161, 1758163, 1758164, 1758165, 1758166, 1995520, 1995522, 1995523, 1995524, 2205545, 2205546, 2205547, 2316311, 2400723, 2408785, 2484275, 2505368, 2505369, 2505370, 2505371, 2533442, 2675868, 2675869, 2675870, 2675871, 2675872, 2675873, 4061421, 4061422, 4249796, 4249803, 5577615; o. REF 144-32-10, GUIDID 10885862027511, Serial Numbers: 646959, 667640, 733100, 733108, 1501641, 1700462, 1758490, 1836563, 1836564, 1836565, 1836567, 1836568, 1836569, 1836570, 1836571, 1836572, 2199475, 2199476, 2314645, 2314646, 2400076, 2400077, 2400078, 4040047.
Devices named in this recall
15 devices in the FDA device registry whose identifier appears in the recall's code information.
- Acumatch 144-28-05Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch 144-28-06Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch 144-28-07Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch 144-28-08Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch 144-28-09Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch 144-28-10Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch A Series 144-32-05Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch A Series 144-32-06Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch A Series 144-32-07Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch A Series 144-32-08Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch A Series 144-32-09Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- Acumatch A Series 144-32-10Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- MCS 146-28-04Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- MCS 146-28-05Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
- MCS 146-32-06Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 90279).
- Z-1726-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH 15 DEGREE LINER 22MM SZ C, REF 132-22-03 b. ACUMATCH 15 DEGREE LINER 28MM SZ D, REF 1…
- Z-1727-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH +5MM LAT LINER 15 DEG 36MM ID, SIZE H, REF 138-36-08 b. ACUMATCH +5MM LAT LINER 15 DE…
- Z-1728-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH GXL 0 DEGREE LINER 28MM SZ E, REF 130-28-25 b. ACUMATCH GXL 0 DEGREE LINER 28MM SZ F,…
- Z-1729-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. NOVATION GXL LINER, NEUTRAL, 28MM ID, GROUP 1 CUPS, REF 130-28-51 b. NOVATION GXL LINER, NEUTRA…
- Z-1730-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. NOVATION CONSTRAINED LINER 28MM ID GROUP 1, REF 134-28-41 b. NOVATION CONSTRAINED LINER 32M…
- Z-1731-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. GXL NEUTRAL LINER, G0 28MM ID, 130-28-50; b. GXL NEUTRAL LINER, G00 22MM ID, 130-22-70; c. N…
- Z-1732-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. NOVATION GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS, REF 130-32-51 b. NOVATION GXL LINER NEUTRAL,…
- Z-1733-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ALL POLYETHYLENE ACETABULAR CUP, CEMENTED 22MM SZ 42, REF 106-22-42; b. ALL POLYETHYLENE ACE…
- Z-1734-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH 0 DEGREE LINER 28MM 46MM, REF 130-28-05; b. ACUMATCH 0 DEGREE LINER 28MM 48MM, REF 13…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page