Novation recallZ-1731-2022 · Exactech, Inc. · initiated Aug 11, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.
FDA root cause category: Process control
What the company is doing
The firm sent an "URGENT MEDICAL DEVICE RECALL" DEAR HEALTHCARE PROFESSIONAL COMMUNICATION on 08/11/2022 by email. The purpose of the current letter is to update surgeons regarding the implications of the packaging non-conformity on performance of the Connexion GXL and other conventional UHMWPE acetabular liners. In addition, this letter provides updates to patient management and follow-up recommendations since issuance of the July 2021 DHCP letter. In addition to providing surgeons with lists of all their affected GXL and/or conventional polyethylene patients since 2004, Exactech is providing surgeons with two draft letters directed at patients who have been implanted with Exactech GXL hip devices. The Customers were instructed to take the following actions: CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required. IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customers inventory. EXTEND THE DESCRIPTION OF THE ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession. COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at [email protected]. Reach out to Customer Service at [email protected] to request a Returned Goods (RG) Number for ALL affected product requested to be returned. WE ARE REQUIRING 100% EFFECTIVENESS FOR THIS RECALL. If you have any questions, call (888) 912-0403.
Recall details
- Recall number
- Z-1731-2022
- Status
- Open, Classified
- Initiated by the company
- Aug 11, 2022
- Recalling firm
- Exactech, Inc.
- Quantity
- 3793 devices
- Product code
- LPH
- FDA submissions
- K100269
- Distribution
- Worldwide distribution: US (nationwide) and OUS (foreign) countries of: ARGENTINA, BRAZIL, CHILE, CHINA, COLOMBIA, ECUADOR, GREECE, GUATEMALA, INDIA, LEBANON PORTUGAL,PUERTO RICO,SPAIN
Product description
GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. GXL NEUTRAL LINER, G0 28MM ID, 130-28-50; b. GXL NEUTRAL LINER, G00 22MM ID, 130-22-70; c. NOVATION GXL LIPPED LINER, G0 28MM ID, 132-28-50; d. NOVATION GXL LIPPED LINER, G00 22MM ID, 132-22-70;
Code information (models, lots, serial numbers)
Show the full code information (34k characters)
a. 130-28-50, GUDID 10885862022059, Serial Numbers: 1784995, 1784996, 1784998, 1785000, 1785002, 1785003, 1785006, 1785011, 1785012, 1785013, 1785015, 1785016, 1785017, 1785018, 1785022, 1785024, 1785026, 1785027, 1785030, 1785032, 1785033, 1796637, 1796640, 1796641, 1796642, 1796644, 1796647, 1796652, 1796654, 1796655, 1796656, 1796658, 1810811, 1810813, 1810814, 1810818, 1810819, 1810821, 1810822, 1810825, 1810828, 1810829, 1810831, 1810832, 1859562, 2540144, 2540145, 2540146, 2540147, 2540148, 2540150, 2633432, 2633433, 2691744, 2691746, 2691747, 2691748, 2691749, 2691750, 2691751, 2691752, 2707045, 2707047, 2707048, 2707049, 2707050, 2800307, 2800308, 2812276, 2812277, 2893670, 2893671, 2893672, 2976865, 2976866, 2976867, 2976868, 2976869, 3544420, 3544421, 3544422, 3544423, 3544424, 3544431, 3544432, 3544435, 3544437, 3544439, 3544442, 3544444, 3544445, 3544446, 3544447, 3544450, 3673569, 3673571, 3673572, 3673573, 3673574, 3673575, 3673577, 3673578, 3673579, 3673581, 3673583, 3673586, 3673587, 3673588, 3673589, 3673592, 3673593, 3673594, 3673595, 3673596, 3673597, 3673598, 3673599, 3673600, 3712228, 3712229, 3712230, 3712231, 3712232, 3712234, 3712236, 3712238, 3712239, 3712241, 3712455, 3712457, 3712458, 3712459, 3712460, 3712462, 3712463, 3744701, 3744702, 3744703, 3744704, 3744705, 3744706, 3744707, 3744708, 3744710, 3744713, 3744714, 3744715, 3744716, 3744717, 3744718, 3744720, 3744722, 3744723, 3744724, 3978806, 3978807, 3978808, 3978810, 3978811, 3978815, 3978817, 3978818, 3978819, 3978820, 3978822, 3978823, 3978824, 3978825, 4142832, 4142836, 4142837, 4142838, 4142842, 4142845, 4142846, 4142847, 4142848, 4142851, 4142854, 4383613, 4383614, 4383615, 4383617, 4383618, 4383620, 4383622, 4383623, 4383624, 4383625, 4383626, 4383628, 4383629, 4383630, 4383631, 4383632, 4383633, 4383634, 4383635, 4383637, 4383638, 4383639, 4478739, 4478740, 4478741, 4478742, 4478743, 4478744, 4478746, 4478747, 4478750, 4478753, 4478754, 4478756, 4478757, 4478758, 4478762, 4478763, 4478764, 4673689, 4673690, 4673692, 4673693, 4673694, 4673695, 4673696, 4673699, 4673700, 4673701, 4699019, 4699020, 4699021, 4699022, 4699023, 4727112, 4727113, 4727114, 4727115, 4727116, 4727117, 4770113, 4770116, 4770117, 4770118, 4770119, 4770120, 4770121, 4770122, 4788829, 4788830, 4788831, 4788832, 4788833, 4788835, 4788836, 4788838, 4813711, 4813712, 4813713, 4813716, 4836757, 4836758, 4836759, 4836760, 4836761, 4836762, 4836763, 4836764, 4836765, 4836766, 4836767, 4836768, 4836769, 4836770, 4836771, 4836773, 4845642, 4845643, 4845644, 4845645, 4845646, 4845647, 4845648, 4845649, 4845650, 4845651, 4845652, 4845653, 4845654, 4845655, 4845656, 4845657, 5027363, 5027364, 5027365, 5027366, 5027367, 5027368, 5027369, 5027371, 5027372, 5092203, 5092207, 5092209, 5092210, 5092211, 5092212, 5133514, 5133515, 5133516, 5133517, 5133518, 5133519, 5133520, 5133521, 5133522, 5133523, 5138402, 5138403, 5138405, 5138406, 5138407, 5138408, 5138409, 5138411, 5192509, 5192510, 5192511, 5192512, 5192513, 5192514, 5192515, 5192518, 5265341, 5265343, 5265346, 5265347, 5265349, 5265350, 5265356, 5265357, 5265359, 5265360, 5265361, 5265362, 5265364, 5469670, 5469671, 5469672, 5469674, 5469677, 5469678, 5469681, 5469682, 5469684, 5469685, 5469687, 5469688, 5525315, 5525316, 5525317, 5525318, 5525319, 5525320, 5525321, 5556531, 5556532, 5556533, 5556534, 5556536, 5556537, 5556539, 5556540, 5556543, 5556544, 5556546, 5556547, 5556550, 5556551, 5556553, 5768901, 5768903, 5768906, 5768908, 5768911, 5768912, 5768915, 5768917, 5768918, 5768919, 5768923, 5768924, 5921794, 5921796, 5921798, 5921802, 5921804, 5921805, 5921807, 5921811, 5921815, 5921816, 5921817, 5921819, 6664657; b. 130-22-70, GUDID 10885862021892, Serial Numbers: 1787743, 1787757, 1787758, 1787764, 2827594, 2827595, 2827596, 2827597, 2827598, 2827599, 2827600, 2827601, 2827602, 2827603, 2827604, 2827605, 2827606, 2827607, 2827608, 2827609, 2827610, 2900032; c. 132-28-50, GUDID 10885862023148, Serial Numbers: 1785082, 1785083, 1785084, 1785088, 1785089, 1785091, 1785092, 1785096, 1785097, 1785098, 1785099, 1785100, 1785101, 1785106, 1785108, 1785110, 1785115, 1785117, 1785119, 1785120, 1785121, 1785122, 1785123, 1797031, 1797032, 1797033, 1797034, 1797035, 1797039, 1797041, 1797042, 1797043, 1797046, 1797047, 1797050, 1811551, 1811552, 1811553, 1811554, 1811556, 1811557, 1811559, 1811560, 1811561, 1811563, 1811564, 1811565, 1811566, 1811567, 1811568, 1811569, 1852506, 1852507, 1852508, 1852509, 1852512, 1852513, 1852515, 1852516, 1852517, 1852518, 1852519, 1852520, 1870759, 1870760, 1870761, 1870762, 1870764, 1870765, 1870766, 1870767, 1870768, 1870769, 1870770, 2133603, 2133604, 2133605, 2133606, 2133607, 2133608, 2133609, 2133610, 2133611, 2133612, 2133613, 2133614, 2133615, 2133616, 2133617, 2133618, 2133619, 2133620, 2133621, 2133622, 2133624, 2133625, 2133626, 2133627, 2133629, 2133630, 2158036, 2158037, 2158038, 2158039, 2158040, 2158041, 2158042, 2158043, 2158044, 2158046, 2158047, 2158048, 2158049, 2158050, 2158051, 2158052, 2186820, 2186822, 2186823, 2186824, 2186825, 2186826, 2186827, 2186828, 2186829, 2193170, 2193171, 2193172, 2193173, 2193175, 2193176, 2193177, 2193178, 2193180, 2193181, 2193182, 2193183, 2193184, 2225603, 2225604, 2225605, 2225606, 2225607, 2225608, 2225609, 2225610, 2225611, 2225612, 2225613, 2225614, 2225615, 2225617, 2225618, 2225619, 2225620, 2225621, 2225622, 2225623, 2225624, 2225625, 2225627, 2225628, 2225629, 2225630, 2229568, 2229569, 2229570, 2229571, 2229572, 2275710, 2275712, 2275714, 2275715, 2275716, 2275719, 2275720, 2275722, 2275723, 2275724, 2275725, 2275726, 2275727, 2275728, 2275729, 2275730, 2275731, 2275733, 2275734, 2275735, 2319138, 2319139, 2319140, 2319142, 2319143, 2319144, 2319145, 2319146, 2319147, 2322304, 2322305, 2322306, 2322307, 2322308, 2322309, 2322310, 2322311, 2322312, 2322313, 2322314, 2322315, 2322316, 2322317, 2322318, 2322319, 2322320, 2322321, 2322322, 2322323, 2322324, 2322325, 2322326, 2322327, 2322329, 2322330, 2322331, 2322332, 2322333, 2359990, 2359991, 2359992, 2359993, 2359994, 2359995, 2359996, 2359997, 2359998, 2359999, 2360000, 2360001, 2360002, 2360003, 2360004, 2360005, 2360006, 2360007, 2360008, 2360009, 2360010, 2360011, 2360012, 2360013, 2360014, 2360015, 2360016, 2360017, 2360018, 2360019, 2360020, 2360021, 2432540, 2432541, 2432542, 2432543, 2432544, 2432545, 2432546, 2432547, 2432548, 2432550, 2432552, 2432553, 2432554, 2432555, 2432556, 2432557, 2432558, 2432559, 2432560, 2432561, 2432562, 2432563, 2432564, 2471859, 2471860, 2471861, 2471862, 2471863, 2471864, 2471865, 2471866, 2471867, 2471868, 2471869, 2471870, 2471871, 2471872, 2471873, 2471874, 2471875, 2471876, 2471877, 2471878, 2471879, 2471880, 2471881, 2471882, 2471883, 2471884, 2479899, 2479900, 2479901, 2479902, 2479903, 2479904, 2479905, 2479906, 2479907, 2479908, 2479909, 2479910, 2479911, 2479912, 2479913, 2479914, 2479915, 2479916, 2479917, 2479918, 2479919, 2479920, 2479921, 2479922, 2479923, 2479924, 2479925, 2479926, 2479927, 2479928, 2479929, 2479930, 2479931, 2479932, 2479933, 2479934, 2479935, 2479936, 2479937, 2479938, 2479939, 2479940, 2479941, 2479942, 2479943, 2506614, 2506615, 2506616, 2506617, 2506618, 2506619, 2506620, 2506621, 2506622, 2506623, 2506624, 2506625, 2506626, 2506627, 2506628, 2506629, 2506630, 2506631, 2506632, 2506633, 2506634, 2506635, 2506636, 2506637, 2506638, 2506639, 2506640, 2506641, 2506642, 2506643, 2506644, 2506645, 2506646, 2506647, 2506648, 2506649, 2506650, 2506651, 2506652, 2506653, 2506654, 2506655, 2506656, 2574470, 2574471, 2574472, 2574473, 2574474, 2574475, 2574476, 2574477, 2574478, 2574479, 2574480, 2574481, 2574482, 2574483, 2574484, 2574485, 2655103, 2655104, 2655105, 2655107, 2655108, 2655109, 2655110, 2667683, 2667685, 2667686, 2680108, 2680109, 2680110, 2680111, 2691767, 2691768, 2691769, 2691770, 2691771, 2691772, 2691773, 2691774, 2691775, 2691777, 2691778, 2691779, 2691780, 2691781, 2707017, 2707018, 2707019, 2707020, 2707021.2707022, 2707023, 2707024, 2707025, 2707026, 2707027, 2707028, 2707029, 2707030, 2707031, 2707032, 2707034, 2707035, 2707036, 2707037, 2707038, 2707039, 2707040, 2707041, 2707042, 2707043, 2707044, 2745483, 2745484, 2745485, 2745486, 2745487, 2745488, 2745489, 2745490, 2745491, 2745492, 2745493, 2745494, 2745495, 2745496, 2745497, 2745498, 2745499, 2745500, 2745501, 2745502, 2745503, 2770664, 2770665, 2770666, 2770667, 2770668, 2770669, 2770670, 2770671, 2770672, 2770673, 2770674, 2770675, 2770676, 2770677, 2770678, 2770679, 2770680, 2770681, 2770682, 2770683, 2770684, 2770685, 2770686, 2770687, 2770688, 2770689, 2770690, 2770691, 2770692, 2770693, 2770694, 2770695, 2796050, 2796051, 2796052, 2796053, 2796054, 2796055, 2796056, 2796057, 2796058, 2796059, 2796060, 2796061, 2796062, 2796063, 2796064, 2796065, 2796066, 2796067, 2796068, 2796069, 2796070, 2796071, 2796072, 2796073, 2796074, 2796075, 2796076, 2796077, 2796078, 2796079, 2796080, 2815983, 2815985, 2815986, 2815987, 2815989, 2815990, 2815991, 2815992, 2815995, 2815996, 2815997, 2815998, 2815999, 2818660, 2818661, 2818662, 2818663, 2818664, 2818665, 2818666, 2818667, 2818668, 2818669, 2818670, 2818671, 2818672, 2818673, 2818674, 2818675, 2818676, 2818677, 2871793, 2871794, 2871795, 2871796, 2871797, 2871798, 2871799, 2871800, 2871801, 2871802, 2871803, 2871804, 2871805, 2871806, 2871807, 2871808, 2871809, 2871810, 2871811, 2871812, 2871813, 2871814, 2871815, 2871817, 2871818, 2871819, 2871820, 2871821, 2871822, 2923594, 2923595, 2923596, 2923597, 2923598, 2923599, 2923600, 2923601, 2923602, 2923603, 2923604, 2923605, 2923606, 2923607, 2923608, 2923609, 2923610, 2923611, 2923612, 2923613, 2923614, 2923615, 2923616, 2923617, 2923618, 2923619, 2923620, 2923621, 2923622, 2923623, 2956982, 2956983, 2956984, 2956985, 2956986, 2956987, 2956988, 2956989, 2956990, 2956991, 2956992, 2956993, 2956994, 2956995, 2956996, 2956997, 2956998, 2956999, 2957000, 2957001, 2957002, 2957003, 2957004, 2957005, 2957006, 2957007, 2957008, 2957009, 2957010, 2957011, 2957012, 2957013, 2957014, 2976663, 2976664, 2976665, 2976666, 2976667, 2976668, 2976669, 2976670, 2976671, 2976672, 2976673, 2976674, 2976676, 2976677, 2976678, 2976679, 2976680, 2976681, 2976682, 2976683, 2976684, 2976685, 2976686, 2976687, 2976688, 2976689, 2976690, 2976691, 2976692, 2976693, 2976694, 2976695, 2976696, 2976697, 2976698, 2976699, 2976700, 2976701, 2976702, 2976704, 2976706, 2976707, 2976708, 2976709, 2976710, 2976711, 2976712, 2976713, 2976714, 2976715, 2976716, 2976717, 2976718, 2976719, 2976720, 2976722, 2976723, 2976724, 2976725, 2976726, 3514249, 3514250, 3514251, 3514252, 3514253, 3514255, 3514256, 3514257, 3514259, 3514260, 3514261, 3514262, 3514263, 3514264, 3514265, 3514266, 3514267, 3514268, 3514269, 3514270, 3514271, 3514272, 3514273, 3514274, 3514275, 3514276, 3514277, 3514278, 3514279, 3514280, 3514281, 3579174, 3579175, 3579177, 3579178, 3579179, 3579180, 3579181, 3579182, 3579183, 3579184, 3579185, 3579186, 3579187, 3579188, 3579189, 3579190, 3579191, 3579192, 3579193, 3579194, 3579195, 3579196, 3579197, 3579198, 3579199, 3579200, 3579201, 3579202, 3579203, 3579204, 3579205, 3579206, 3590161, 3590162, 3590163, 3590164, 3590165, 3590166, 3590167, 3590168, 3590169, 3590170, 3590171, 3590172, 3590173, 3590174, 3590175, 3590177, 3590179, 3590180, 3590181, 3590183, 3590184, 3590185, 3590186, 3590187, 3590188, 3590189, 3590190, 3590191, 3590193, 3590194, 3590195, 3590198, 3590200, 3590201, 3590202, 3640729, 3640730, 3640731, 3640732, 3640734, 3640735, 3640736, 3640737, 3640738, 3640739, 3640740, 3640741, 3640742, 3640743, 3640744, 3640745, 3640746, 3640747, 3640748, 3640749, 3640750, 3640752, 3640753, 3640756, 3640758, 3640760, 3658110, 3658111, 3658112, 3658113, 3658114, 3658115, 3658116, 3658117, 3658118, 3658119, 3658120, 3658121, 3658122, 3658123, 3658124, 3658125, 3658126, 3658127, 3658128, 3658129, 3658130, 3658131, 3658132, 3658133, 3658134, 3658135, 3658136, 3658137, 3658138, 3658139, 3658140, 3658141, 3658142, 3658143, 3658144, 3658145, 3658146, 3658147, 3658148, 3658149, 3658150, 3658151, 3658152, 3658153, 3668657, 3668658, 3668659, 3668660, 3668661, 3668662, 3668663, 3668664, 3668665, 3668666, 3668667, 3668668, 3668669, 3668670, 3668671, 3668672, 3668673, 3668674, 3668675, 3668676, 3668677, 3668678, 3668679, 3668680, 3668681, 3668682, 3668683, 3668684, 3668685, 3668686, 3668687, 3668688, 3668689, 3668690, 3668691, 3668692, 3668694, 3668695, 3668696, 3668697, 3668698, 3668699, 3668700, 3668701, 3668702, 3668703, 3668704, 3668705, 3668706, 3679509, 3679510, 3679511, 3679512, 3679513, 3679514, 3679515, 3679516, 3679517, 3679518, 3679519, 3679520, 3679521, 3719327, 3719328, 3719329, 3719330, 3719331, 3719332, 3719333, 3719334, 3719335, 3719336, 3719337, 3719338, 3719339, 3719340, 3719341, 3719342, 3719343, 3719344, 3719345, 3719346, 3719347, 3719348, 3719349, 3719350, 3719351, 3719352, 3719353, 3719354, 3719355, 3719356, 3719357, 3719358, 3719359, 3719360, 3719361, 3719362, 3719363, 3719364, 3719365, 3719366, 3719367, 3719368, 3719369, 3719370, 3719371, 3719372, 3719373, 3719374, 3719375, 3719376, 3719377, 3719378, 3719379, 3719380, 3719381, 3766107, 3766108, 3766109, 3766110, 3766111, 3766112, 3766113, 3766114, 3766115, 3766116, 3766117, 3766118, 3766119, 3766120, 3766121, 3766122, 3766123, 3766124, 3766125, 3766126, 3766127, 3766128, 3766129, 3766130, 3766131, 3766132, 3766133, 3766134, 3766135, 3766136, 3792204, 3792205, 3792206, 3792207, 3792208, 3792209, 3792210, 3792211, 3792212, 3792213, 3792214, 3792215, 3792216, 3792217, 3792218, 3792219, 3792220, 3792221, 3792222, 3792223, 3792224, 3792225, 3792226, 3792227, 3792228, 3792229, 3792230, 3792231, 3792232, 3792233, 3844316, 3844317, 3844318, 3844319, 3844320, 3844321, 3844322, 3844323, 3844324, 3844325, 3844326, 3844327, 3844328, 3844329, 3844330, 3844331, 3844332, 3844333, 3844334, 3844335, 3844336, 3844337, 3844338, 3844339, 3844340, 3844341, 3844342, 3844343, 3844344, 3844345, 3847011, 3847012, 3847013, 3847014, 3847015, 3847016, 3877041, 3877042, 3877043, 3877044, 3877045, 3877046, 3877047, 3877048, 3877049, 3877050, 3958850, 3958851, 3958852, 3958853, 3958854, 3958855, 3958856, 3958857, 3958858, 3958859, 3958860, 3958861, 3958862, 3958863, 3958864, 3958865, 3958866, 3958867, 3958868, 3958869, 3958870, 3958871, 3958873, 3958874, 3958875, 3958876, 3958877, 3958878, 3958879, 3958880, 3958881, 3958882, 3958883, 3958884, 3958885, 3958886, 3958887, 3958888, 3958889, 3958890, 3958891, 3958892, 3958893, 3958894, 3958895, 3958896, 3958897, 3958898, 3958899, 3958900, 3958901, 3958902, 3958903, 3958904, 3958905, 3958906, 3958907, 3958908, 3958909, 3964184, 3964185, 3964186, 3964187, 3964188, 3964189, 3964190, 3964191, 3964192, 3964193, 3964194, 3964195, 3964196, 3964197, 3964198, 3964199, 3964200, 3964201, 3964202, 3964203, 3964204, 3964205, 3964206, 3964207, 3964208, 3964209, 3964210, 3964211, 3964212, 3964213, 4038155, 4038156, 4038157, 4038158, 4038159, 4038160, 4038161, 4038162, 4038163, 4038164, 4038165, 4038166, 4038169, 4038170, 4038171, 4038173, 4038174, 4038175, 4038176, 4038177, 4038178, 4038179, 4038180, 4038181, 4038182, 4038183, 4038184, 4038185, 4038186, 4038187, 4038188, 4038189, 4038190, 4038191, 4038192, 4038193, 4038194, 4038222, 4038223, 4038224, 4038225, 4038226, 4038227, 4038228, 4038229, 4038230, 4038231, 4038232, 4038233, 4038234, 4038235, 4038236, 4038237, 4038238, 4038239, 4038240, 4038241, 4038242, 4038243, 4038244, 4038245, 4038246, 4038247, 4038248, 4038249, 4038250, 4038251, 4038252, 4038253, 4056512, 4056513, 4056514, 4056515, 4056516, 4056518, 4056519, 4056520, 4056521, 4056522, 4056523, 4056524, 4056525, 4056526, 4056528, 4056529, 4056530, 4056531, 4056532, 4056533, 4056534, 4056535, 4056536, 4056537, 4056538, 4120703, 4120704, 4120705, 4120706, 4120707, 4120709, 4120710, 4120711, 4120713, 4120714, 4120715, 4120716, 4120717, 4120718, 4120719, 4120720, 4120721, 4120722, 4120723, 4120724, 4120725, 4120726, 4120727, 4120728, 4120729, 4120731, 4120732, 4120733, 4120734, 4120736, 4120737, 4120738, 4120739, 4120740, 4120741, 4120742, 4120743, 4120744, 4120745, 4120746, 4120747, 4120748, 4120749, 4120750, 4120751, 4120752, 4120753, 4120754, 4148971, 4148972, 4148973, 4148974, 4148975, 4148976, 4148977, 4148978, 4148980, 4148981, 4148982, 4148983, 4148984, 4148985, 4148986, 4148987, 4148988, 4148989, 4148990, 4148991, 4148992, 4148993, 4148994, 4148995, 4148996, 4148997, 4148998, 4148999, 4149000, 4149002, 4149003, 4149055, 4149056, 4149057, 4149058, 4149059, 4149060, 4149061, 4149062, 4149063, 4149064, 4149065, 4149066, 4149067, 4149068, 4149069, 4149070, 4149071, 4149072, 4149073, 4149074, 4149075, 4149076, 4149077, 4149078, 4149079, 4149080, 4149081, 4149082, 4149083, 4149084, 4149085, 4173006, 4173007, 4173008, 4173009, 4173010, 4173011, 4173012, 4173013, 4173014, 4173015, 4173016, 4173017, 4173018, 4173019, 4173020, 4173021, 4173022, 4173023, 4173024, 4173025, 4173026, 4173027, 4173028, 4173029, 4173030, 4173031, 4173032, 4173033, 4173034, 4173035, 4173036, 4173037, 4173038, 4215079, 4215080, 4215082, 4215083, 4215085, 4215086, 4215087, 4215088, 4215089, 4215090, 4215091, 4215092, 4215093, 4215094, 4215095, 4215096, 4215097, 4215098, 4215099, 4215100, 4215101, 4215102, 4215103, 4215104, 4215105, 4215106, 4215107, 4215108, 4242703, 4242704, 4242705, 4242706, 4242707, 4242708, 4242709, 4242710, 4242711, 4242712, 4242713, 4242714, 4242715, 4242716, 4242717, 4242718, 4242719, 4242720, 4242721, 4242722, 4242723, 4242724, 4279797, 4279798, 4279799, 4279800, 4279801, 4279802, 4279803, 4279804, 4279805, 4279806, 4279807, 4279808, 4279809, 4279810, 4279811, 4279812, 4279813, 4279814, 4279815, 4279816, 4279817, 4279818, 4279819, 4279820, 4283927, 4283928, 4283929, 4283930, 4283931, 4283932, 4283933, 4283934, 4283935, 4283936, 4283937, 4283938, 4283939, 4283940, 4283941, 4283942, 4283943, 4283944, 4283945, 4283946, 4283947, 4283948, 4283949, 4283950, 4300502, 4300503, 4300504, 4300505, 4300506, 4300507, 4300508, 4300509, 4300510, 4300511, 4300512, 4300513, 4300514, 4300515, 4300516, 4300517, 4300518, 4300519, 4300520, 4300521, 4300522, 4300523, 4300524, 4300525, 4300526, 4300527, 4300528, 4300529, 4300530, 4300531, 4304556, 4304557, 4304559, 4304560, 4304561, 4304562, 4304563, 4304564, 4304565, 4304566, 4304567, 4304568, 4304569, 4304570, 4304571, 4304572, 4304573, 4304574, 4304575, 4304576, 4304577, 4304578, 4304579, 4304580, 4304581, 4304582, 4304583, 4304584, 4304585, 4354848, 4354849, 4354850, 4354851, 4354852, 4354853, 4354854, 4354855, 4354856, 4354857, 4354858, 4354859, 4354860, 4354861, 4354862, 4354863, 4354864, 4354865, 4354866, 4354867, 4354868, 4354870, 4354871, 4354872, 4354873, 4354874, 4354875, 4354876, 4354877, 4354878, 4354879, 4354880, 4354881, 4354882, 4354883, 4354884, 4354885, 4354886, 4354887, 4354888, 4354889, 4354890, 4354891, 4354892, 4354893, 4354894, 4354895, 4354896, 4354897, 4354898, 4354899, 4354900, 4354901, 4354902, 4354903, 4354904, 4354905, 4354906, 4354907, 4378087, 4378088, 4378089, 4378090, 4378091, 4378092, 4378093, 4378094, 4378095, 4378096, 4378097, 4378098, 4378099, 4378100, 4378101, 4378102, 4378103, 4378104, 4378105, 4378106, 4378107, 4378108, 4378109, 4378110, 4378111, 4378112, 4378113, 4378114, 4378115, 4378116, 4378117, 4378118, 4378119, 4378120, 4378121, 4378122, 4378123, 4378124, 4378125, 4378126, 4378127, 4378128, 4378129, 4378130, 4378131, 4378132, 4378133, 4378134, 4378135, 4378136, 4378137, 4378138, 4378139, 4378140, 4378141, 4378142, 4378143, 4378144, 4378145, 4378146, 4407089, 4407090, 4407091, 4407092, 4407093, 4407094, 4407095, 4407096, 4407097, 4407098, 4407099, 4407100, 4407101, 4407102, 4407103, 4407104, 4407105, 4407106, 4407107, 4407108, 4407109, 4407110, 4407111, 4407112, 4407113, 4407114, 4407115, 4409560, 4409561, 4409564, 4409565, 4409566, 4409567, 4409568, 4409569, 4409571, 4409573, 4409574, 4409575, 4409576, 4409577, 4409578, 4409579, 4409581, 4409582, 4465720, 4465721, 4465722, 4465723, 4465724, 4465725, 4465726, 4465727, 4465728, 4465729, 4465730, 4465731, 4465732, 4465733, 4465734, 4465735, 4465736, 4465737, 4465738, 4465739, 4465740, 4465741, 4465742, 4465743, 4465744, 4465745, 4465746, 4465747, 4465748, 4465749, 4465750, 4465752, 4465754, 4465755, 4465756, 4465757, 4465758, 4465759, 4465760, 4465761, 4465762, 4465763, 4465764, 4465765, 4465766, 4465767, 4465768, 4465770, 4465771, 4465772, 4465773, 4465774, 4465775, 4465777, 4465778, 4465779, 4483569, 4483570, 4483571, 4483572, 4483573, 4483574, 4483575, 4483576, 4483577, 4483578, 4483579, 4483580, 4483581, 4483582, 4483583, 4483584, 4483585, 4483586, 4483587, 4483588, 4483589, 4483590, 4483591, 4483592, 4483593, 4483594, 4483595, 4483596, 4483597, 4483598, 4537717, 4537718, 4537719, 4537720, 4537721, 4537722, 4537723, 4537724, 4537725, 4537726, 4537727, 4537728, 4537729, 4537730, 4537731, 4537732, 4537733, 4537734, 4537735, 4537736, 4537737, 4537738, 4537739, 4537740, 4537741, 4537742, 4537743, 4537744, 4537745, 4537746, 4537747, 4537748, 4537749, 4537750, 4537751, 4537752, 4537753, 4537754, 4537755, 4537756, 4537757, 4537758, 4537760, 4537761, 4537762, 4537763, 4537764, 4537765, 4537766, 4537767, 4537768, 4537769, 4537770, 4537771, 4537772, 4537773, 4537774, 4537775, 4537776, 4537777, 4537778, 4537780, 4537781, 4537782, 4537783, 4537784, 4537785, 4537786, 4537787, 4537788, 4537789, 4537790, 4537791, 4537792, 4537793, 4537794, 4537795, 4537797, 4537798, 4537799, 4537800, 4537802, 4537803, 4537804, 4537805, 4537806, 4594296, 4594297, 4594298, 4594299, 4594300, 4594301, 4594302, 4594303, 4594304, 4594305, 4594306, 4594307, 4594308, 4594309, 4594310, 4594311, 4594312, 4594313, 4594314, 4594315, 4594316, 4594317, 4594318, 4594319, 4594320, 4594321, 4594322, 4594323, 4594324, 4634867, 4634868, 4634869, 4634871, 4634872, 4634873, 4634876, 4634877, 4634878, 4634879, 4634880, 4634881, 4634882, 4634883, 4634884, 4634885, 4634886, 4634888, 4634889, 4634890, 4634891, 4634892, 4634893, 4634894, 4634895, 4634896, 4648523, 4648524, 4648525, 4648526, 4648528, 4648529, 4648530, 4648531, 4648532, 4648533, 4648534, 4648535, 4648536, 4648537, 4648538, 4648539, 4648540, 4648541, 4648542, 4648543, 4648544, 4648545, 4648546, 4648547, 4648548, 4648549, 4648551, 4648552, 4672693, 4672694, 4672695, 4672697, 4672698, 4672699, 4672700, 4672701, 4672702, 4672703, 4672704, 4672705, 4672706, 4672707, 4672708, 4672710, 4672711, 4672712, 4672713, 4672714, 4672716, 4672717, 4672718, 4672719, 4672720, 4672721, 4672722, 4679309, 4679310, 4679311, 4679312, 4679314, 4679315, 4679316, 4679317, 4679318, 4679319, 4679320, 4679321, 4679322, 4679323, 4679324, 4679325, 4679326, 4679327, 4679328, 4679329, 4679330, 4679332, 4679333, 4679335, 4679336, 4679337, 4679338, 4767585, 4767586, 4767587, 4767588, 4767589, 4767591, 4767592, 4767593, 4767594, 4767595, 4767596, 4767597, 4767598, 4767599, 4767600, 4767601, 4767602, 4767603, 4767604, 4767605, 4767606, 4767607, 4767608, 4767609, 4767610, 4767611, 4767612, 4767613, 4767614, 4788869, 4788870, 4788871, 4788872, 4788874, 4788875, 4788876, 4788877, 4788878, 4788879, 4788880, 4788881, 4788882, 4788883, 4788884, 4788885, 4788886, 4788887, 4788889, 4788890, 4788891, 4788892, 4788893, 4788895, 4788896, 4788897, 4788898, 4817484, 4817485, 4817486, 4817487, 4817488, 4817489, 4817490, 4817492, 4817493, 4817495, 4817496, 4817497, 4817499, 4817500, 4817501, 4817502, 4817503, 4817505, 4817506, 4817507, 4817509, 4817510, 4817511, 4817512, 4817513, 4817514, 4817515, 4817516, 4817517, 4817519, 4817520, 4817521, 4817522, 4817523, 4817524, 4817525, 4817527, 4817528, 4817529, 4817530, 4817531, 4817532, 4817533, 4817534, 4817535, 4817536, 4817537, 4817538, 4817539, 4817540, 4817541, 4817542, 4817543, 4845577, 4845578, 4845579, 4845580, 4845581, 4845582, 4845583, 4845584, 4845585, 4845586, 4845587, 4845588, 4845589, 4845590, 4845591, 4845592, 4845593, 4845594, 4845595, 4845596, 4845597, 4845598, 4845599, 4845600, 4845601, 4845602, 4845603, 4845604, 4845605, 4845606, 4879257, 4879258, 4879259, 4879260, 4879261, 4879262, 4879263, 4879264, 4879265, 4879266, 4879267, 4879268, 4879269, 4879270, 4879271, 4879272, 4879273, 4879274, 4879275, 4879276, 4879277, 4879278, 4879279, 4879280, 4879281, 4879282, 4879283, 4879284, 4879285, 4879286, 4975416, 4975417, 4975418, 4975419, 4975420, 4975421, 4975422, 4975424, 4975425, 4975426, 4975427, 4975428, 4975429, 4975430, 4975431, 4975432, 4986993, 4986994, 4986995, 4986996, 4986997, 4986998, 4986999, 4987000, 4987001, 4987002, 4987003, 4987004, 4987005, 4987006, 4987007, 4987008, 4987009, 4987010, 4987011, 4987012, 4987013, 4987014, 4987015, 4987016, 4987017, 4987018, 4987019, 4987020, 4987021, 4987022, 5139027, 5139028, 5139029, 5139030, 5139031, 5139032, 5139034, 5139035, 5139036, 5139037, 5139038, 5139039, 5139040, 5139041, 5139042, 5139043, 5139044, 5139045, 5139046, 5139047, 5139048, 5139049, 5139050, 5139051, 5139052, 5139053, 5139054, 5139055, 5139056, 5162809, 5162810, 5162811, 5162812, 5162813, 5162815, 5162816, 5162817, 5162818, 5162819, 5162820, 5162821, 5162822, 5162823, 5162824, 5162825, 5162826, 5162827, 5162828, 5162829, 5162830, 5162831, 5162832, 5162833, 5162834, 5162835, 5162836, 5162837, 5162838, 5187151, 5187152, 5187153, 5187154, 5187155, 5187156, 5187157, 5187158, 5187159, 5187160, 5187161, 5187162, 5187163, 5187164, 5187165, 5187166, 5187167, 5187168, 5187169, 5187170, 5187171, 5187172, 5187173, 5187174, 5187175, 5187176, 5187178, 5187179, 5187180, 5219326, 5219327, 5219328, 5219329, 5219330, 5219331, 5219332, 5219333, 5219334, 5219335, 5219336, 5219339, 5219340, 5219341, 5219342, 5219343, 5219344, 5219345, 5219346, 5219347, 5219348, 5219349, 5219350, 5219351, 5219352, 5219353, 5219354, 5219355, 5270347, 5270348, 5270349, 5270350, 5270351, 5270352, 5270353, 5270354, 5270355, 5270356, 5270357, 5270358, 5270359, 5270360, 5270361, 5270362, 5270363, 5270364, 5270365, 5270366, 5270367, 5270368, 5270369, 5270370, 5270371, 5270372, 5270373, 5270374, 5270375, 5270376, 5326140, 5326141, 5326142, 5326143, 5326145, 5326146, 5340128, 5340129, 5340130, 5340131, 5340132, 5340133, 5340134, 5340135, 5340136, 5340137, 5340138, 5340139, 5340140, 5340141, 5340142, 5340143, 5340144, 5340145, 5340146, 5340147, 5340148, 5340149, 5340150, 5340151, 5340152, 5369095, 5369096, 5369097, 5369098, 5369099, 5369100, 5369101, 5369102, 5369104, 5369105, 5369106, 5369107, 5369108, 5369109, 5369110, 5369111, 5369114, 5369115, 5369116, 5369117, 5369118, 5369119, 5369120, 5369121, 5369122, 5369123, 5369124, 5369125, 5369126, 5369127, 5403258, 5403260, 5403261, 5403262, 5403263, 5403264, 5403265, 5403266, 5403267, 5403268, 5403269, 5403270, 5403271, 5403272, 5403273, 5403275, 5403276, 5403277, 5403278, 5403279, 5403280, 5403281, 5403282, 5403283, 5403284, 5403285, 5403286, 5403287, 5446414, 5446415, 5446416, 5446417, 5446418, 5446419, 5446420, 5446421, 5446422, 5446423, 5446424, 5446425, 5446426, 5446427, 5446428, 5446429, 5446430, 5446431, 5446432, 5446433, 5446434, 5446435, 5446436, 5446437, 5446438, 5446439, 5446440, 5446441, 5446442, 5446443, 5451231, 5451232, 5451233, 5451234, 5451235, 5451236, 5451237, 5451238, 5451239, 5451240, 5451241, 5451242, 5451243, 5451244, 5451245, 5451246, 5451247, 5451248, 5451249, 5451250, 5451251, 5451252, 5451253, 5451254, 5451255, 5451256, 5451257, 5451258, 5451259, 5451260, 5451261, 5451262, 5451263, 5451264, 5548402, 5548403, 5548404, 5548406, 5548407, 5548408, 5548409, 5548410, 5548411, 5548412, 5548413, 5548414, 5548415, 5548416, 5548417, 5548418, 5548419, 5548420, 5548421, 5548422, 5548423, 5548424, 5548425, 5548426, 5548427, 5548428, 5548429, 5548430, 5548431, 5548433, 5567332, 5567333, 5567334, 5567335, 5567336, 5567337, 5567339, 5567340, 5567341, 5567342, 5567343, 5567344, 5567346, 5567347, 5567348, 5567349, 5567350, 5567351, 5567352, 5567355, 5567356, 5567357, 5567358, 5567359, 5567360, 5567361, 5567362, 5567363, 5567364, 5567365, 5567366, 5567367, 5653813, 5653814, 5653817, 5653818, 5653819, 5653820, 5653821, 5653823, 5653824, 5653825, 5653826, 5653827, 5653829, 5653830, 5653831, 5653832, 5653833, 5653834, 5653835, 5653836, 5653837, 5653838, 5653839, 5653840, 5653841, 5653842, 5653843, 5653844, 5653845, 5653846, 5653847, 5653848, 5653849, 5653850, 5653851, 5653852, 5653853, 5653854, 5653855, 5653856, 5653857, 5653858, 5653859, 5653860, 5657775, 5657777, 5657778, 5657779, 5657780, 5657781, 5657782, 5657783, 5657784, 5657785, 5657786, 5657787, 5657788, 5657789, 5657791, 5657792, 5657793, 5657794, 5657795, 5657796, 5657797, 5657798, 5658915, 5658916, 5658918, 5658919, 5658920, 5658921, 5658922, 5658923, 5658924, 5658926, 5658927, 5658929, 5658931, 5658932, 5658933, 5658935, 5658936, 5658937, 5658938, 5754358, 5754359, 5754360, 5754361, 5754362, 5754363, 5754365, 5754366, 5754367, 5754370, 5754371, 5754372, 5754373, 5754375, 5754376, 5754377, 5754378, 5754380, 5754381, 5754382, 5754384, 5754385, 5754386, 5781370, 5781371, 5781372, 5781373, 5781374, 5781375, 5781376, 5781377, 5781378, 5781379, 5781380, 5781381, 5781382, 5781383, 5781384, 5781385, 5781387, 5781388, 5781389, 5781390, 5781391, 5781392, 5781393, 5781394, 5781395, 5781396, 5781397, 5781398, 5781399, 5863136, 5863137, 5863138, 5863139, 5863140, 5863142, 5863143, 5863144, 5863145, 5863146, 5863147, 5863148, 5863149, 5863150, 5863151, 5863152, 5863153, 5863154, 5863155, 5863156, 5863157, 5863158, 5863159, 5863160, 5863161, 5863162, 5863163, 5863164, 5863165, 5863421, 5863423, 5863424, 5863425, 5863426, 5863427, 5863428, 5863429, 5863430, 5863432, 5863433, 5863434, 5863435, 5863436, 5863437, 5863438, 5863439, 5863440, 5863441, 5863442, 5863443, 5863444, 5863445, 5863446, 5863447, 5863448, 5863450, 5882888, 5882889, 5882890, 5882891, 5969917, 5969918, 5969919, 5969920, 5969921, 5969922, 5969923, 5969924, 5969925, 5969926, 5969927, 5969928, 5969929, 5969930, 5969931, 5969932, 5969933, 5969935, 5969936, 5969937, 5969938, 5969940, 5969941, 5969942, 5969944, 5969946, 5969947, 5970352, 5970353, 5970355, 5970356, 5970357, 5970358, 5970359, 5970360, 5970361, 5970363, 5970364, 5970366, 5970367, 5970368, 5970369, 5970370, 5970371, 5970372, 5970373, 5970374, 5970375, 5970377, 5970378, 5970379, 5970380, 5970381, 5970382, 6073340, 6073341, 6073342, 6073343, 6073344, 6073345, 6073346, 6073348, 6073349, 6073350, 6073351, 6073352, 6073353, 6073354, 6073355, 6073356, 6073357, 6073358, 6073359, 6073360, 6073361, 6073362, 6073363, 6073364, 6073365, 6073366, 6073367, 6073368, 6073369, 6073370, 6073371, 6073374, 6073375, 6073376, 6073377, 6073378, 6073379, 6073380, 6073381, 6073382, 6073383, 6073384, 6077256, 6077258, 6077259, 6077260, 6077261, 6077262, 6077266, 6077267, 6077268, 6077269, 6077270, 6077271, 6077273, 6077274, 6077275, 6077276, 6077278, 6079254, 6079255, 6079256, 6079257, 6079258, 6079259, 6079260, 6079261, 6079262, 6079263, 6079264, 6079265, 6079266, 6079267, 6079268, 6079269, 6079270, 6079271, 6079272, 6079273, 6079274, 6079275, 6079276, 6079277, 6142849, 6142850, 6142851, 6142852, 6142853, 6142854, 6142855, 6142856, 6142857, 6142858, 6142860, 6142861, 6142862, 6142863, 6142864, 6142865, 6142866, 6142867, 6142868, 6142869, 6142892, 6142893, 6142895, 6142896, 6142897, 6142898, 6142899, 6142900, 6142901, 6142902, 6142903, 6142905, 6142906, 6142908, 6142909, 6142910, 6142911, 6142912, 6142913, 6142915, 6142916, 6142918, 6142924, 6142926, 6142927, 6142929, 6142930, 6142931, 6142932, 6142934, 6323878, 6323879, 6323880, 6323881, 6323882, 6323883, 6323884, 6323885, 6323886, 6323888, 6323889, 6323891, 6323892, 6323894, 6323897, 6323898, 6323900, 6323901, 6323902, 6323903, 6323904, 6323905, 6323906, 6323907, 6323908, 6323909, 6323910, 6323912, 6323914, 6323915, 6323916, 6323918, 6323919, 6323920, 6323921, 6323922, 6323923, 6323925, 6323926, 6323927, 6323928, 6323930, 6323931, 6323932, 6323933, 6323934, 6323935, 6323936, 6323937, 6323938, 6323939, 6323940, 6551560, 6551561, 6551562, 6551563, 6551564, 6551565, 6551566, 6551567, 6551568, 6551569, 6551570, 6551571, 6551572, 6551573, 6551574, 6551575, 6551576, 6551577, 6551578, 6551579, 6551580, 6551581, 6551582, 6551583, 6551584, 6551585, 6551586, 6551587, 6551588, 6551589, 6551590, 6551591, 6551592, 6551593, 6551594, 6551595, 6551596, 6551597, 6551598, 6551599, 6551600, 6551601, 6551602, 6551603, 6551604, 6551605, 6551606, 6551607, 6551608, 6551609, 6551610, 6551611, 6551612, 6551613, 6551614, 6551616, 6551617, 6551619, 6551620, 6551622, 6823051, 6823052, 6823053, 6823054, 6823055, 6823056, 6823057, 6823058, 6823059, 6823062, 6823063, 6823064, 6823065, 6823066, 6823068, 6823069, 6823070, 6823072, 6823073, 6823074, 6823076, 6823077, 6823078, 6823079, 6823080, 6823082, 6823083, 6823085, 6823086, 6823087, 6823088, 6823089, 6886624, 6886626, 6886627, 6886628, 6886629, 6886630, 6886633, 6886634, 6886638, 6886639, 6886641, 6886642, 6886643, 6886644, 6886646, 6886647, 6886653, 6886654, 6886656, 6886657, 6886659, 6886661, 6886662, 6925933, 6925935, 6925937, 6925938, 6925940, 6925941, 6925943, 6925944, 6925945, 6925946, 6925947, 6925948, 6925949, 6925951, 7036197, 7036199; d. 132-22-70, GUDID 10885862022967, Serial Numbers: 1785049, 1785056, 1785058, 1785059, 1785061, 1785064, 1785067, 1785071, 1785077, 1785078, 1785079, 2131590, 2131595, 2202604, 2620392, 2620397, 2620398, 2620399, 2620400, 2620401, 2620403, 3650010, 3650012, 3780297, 3780298, 3780299, 3780301, 3780302, 3780304, 3780305, 3863468, 3863470, 3863471, 3863473, 3863475, 3863476, 3863477, 3863479, 3863480, 3863481, 3863482, 3863483, 3863484, 3863485, 3863486, 3863487, 3863489, 3863490, 3863491, 3863494, 3909677, 3909678, 3909679, 3909680, 4245383, 4245385, 4245389, 4245390, 4245392, 4245393, 4245395, 4245396, 4245397, 4245398, 4245399, 5988300, 5988301, 5988302, 5988303, 5988304, 5988305, 5988306, 6100230, 6100231, 6100232, 6100233, 6100234, 6100236, 6100237, 6100238, 6100239, 6100240, 6100242, 6100243, 6100244, 6100246, 6100247, 6100248, 6100249, 6100250, 6100251, 6796472, 6796473, 6796474, 6796475, 6796476, 6796477, 6796478, 6796479, 6796480, 6796481, 6796482, 6796483, 6796484, 6796485, 6796486, 6796487, 6796488, 6796489, 6796490, 6796491, 6796492, 6796493, 6796494, 6796495, 6796496, 6796497, 6851098, 6851099, 6851100, 6851101, 6851102, 6851103, 6851104, 6851105, 6851106, 6851107, 6851108, 6851109, 6851110, 6851111, 6851112, 6851113, 6851114, 6851115, 6851116, 6851117, 6851118, 6851119, 6851120, 6851121, 6851122, 6851123, 6851124.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Novation 130-28-50Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls19 reports
- Novation 132-28-50Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls2 reports
- Novation 130-22-70Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls0 reports
- Novation 132-22-70Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls0 reports
Product lines
Other devices under the same FDA submission
9 devices cleared under K100269 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NA 181-02-01Exactech, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate1 recall1 report
- Novation 136-22-70Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate1 recall0 reports
- Novation 136-28-50Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate1 recall0 reports
- Novation 138-22-70Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate1 recall0 reports
- Novation 138-28-50Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate1 recall0 reports
- NA 121-01-02Exactech, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented4 reports
- Novation 180-01-44Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 reports
- NA 181-02-00Exactech, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- Novation 180-01-46Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 90279).
- Z-1725-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. 0 DEG ACUMATCH CONSTRAINED LINER SZ E 28MM ID, REF 144-28-05; b. 0 DEG ACUMATCH CONSTRAINED LI…
- Z-1726-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH 15 DEGREE LINER 22MM SZ C, REF 132-22-03 b. ACUMATCH 15 DEGREE LINER 28MM SZ D, REF…
- Z-1727-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH +5MM LAT LINER 15 DEG 36MM ID, SIZE H, REF 138-36-08 b. ACUMATCH +5MM LAT LINER 15 DE…
- Z-1728-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH GXL 0 DEGREE LINER 28MM SZ E, REF 130-28-25 b. ACUMATCH GXL 0 DEGREE LINER 28MM SZ …
- Z-1729-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. NOVATION GXL LINER, NEUTRAL, 28MM ID, GROUP 1 CUPS, REF 130-28-51 b. NOVATION GXL LINER, NEUT…
- Z-1730-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. NOVATION CONSTRAINED LINER 28MM ID GROUP 1, REF 134-28-41 b. NOVATION CONSTRAINED LINER 32M…
- Z-1732-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. NOVATION GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS, REF 130-32-51 b. NOVATION GXL LINER NEUTRAL,…
- Z-1733-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ALL POLYETHYLENE ACETABULAR CUP, CEMENTED 22MM SZ 42, REF 106-22-42; b. ALL POLYETHYLENE ACE…
- Z-1734-2022 · Open, ClassifiedGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH 0 DEGREE LINER 28MM 46MM, REF 130-28-05; b. ACUMATCH 0 DEGREE LINER 28MM 48MM, REF 13…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page