Skip to content

SOLTIVE recallZ-1296-2022 · Olympus Corporation of the Americas · initiated May 26, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device model and size for affected lots of Laser Fibers may be mismatched between the product's carton, immediate pouch packaging, and/or the device.

FDA root cause category: Labeling mix-ups

What the company is doing

Firm notified customers by letter on May 26, 2022. Customers were instructed to return on hand inventory to Olympus for credit.

Recall details

Recall number
Z-1296-2022
Status
Terminated
Initiated by the company
May 26, 2022
Recalling firm
Olympus Corporation of the Americas
Quantity
190 cartons (950 pieces total)
Product code
GEX
FDA submissions
K183647
Distribution
Nationwide distribution in the US. Foreign distribution worldwide.

Product description

SOLTIVE SuperPulsed Laser Fibers, Model# Labeled TFL-FBX150BS (Carton) - 150 Micron Ball Tip Single Use Fiber Model # Labeled TFL-FBX200S (Pouch) - 200 Micron Single Use Fiber Model # Labeled TFL-FBX200BS (Carton) - 200 Micron Ball Tip Single Use Fiber Model # Labeled TFL-FBX550S (Carton) - 550 Micron Single Use Fiber Model # Labeled TFL-FBX200S (Carton) - 200 Micron Single Use Fiber Model # Labeled TFL-FBX365S (Pouch) - 200 Micron Single Use Fiber

Code information (models, lots, serial numbers)

UDI-DI Lot Model# Labeled  00821925043923 KR149856 TFL-FBX150BS (Carton) 00821925043923 KR149856 TFL-FBX200S (Pouch) 00821925043985 KR195775 TFL-FBX200BS (Carton)  00821925044043 KR197789 TFL-FBX550S (Carton)  00821925043930 KR206254 TFL-FBX200S (Carton)  00821925043930 KR206254 TFL-FBX365S (Pouch)

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page