TactiCath recallZ-1344-2022 · Abbott · initiated Jun 9, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When connected to the EnSite Precision Navigation System, an affected TactiCath Contact Force Ablation Catheter, Sensor Enabled may present the error message: "invalid catheter,, or "expired catheter."

FDA root cause category: Device Design

What the company is doing

The firm notified its consignees by hand delivered Medical Device Recall letter on 06/09/2022. The letter explained the problem and provided the following instructions: "Next Steps To help reduce risk, we recommend: " Do not use any remaining inventory from the affected lots listed in Appendix A. " Complete and return the accompanying Acknowledgement Form to Abbott. " Return all remaining unused affected devices to Abbott. Your Abbott representative can assist you in returning these devices and obtaining replacements. Please forward this notice to anyone within your organization who may need to be notified and maintain a record of this notice along with a copy of the completed Acknowledgement Form for your records."

Recall details

Recall number
Z-1344-2022
Status
Terminated
Initiated by the company
Jun 9, 2022
Posted by the FDA
Jul 7, 2022
Recalling firm
Abbott
Quantity
191 devices
Product code
OAE
FDA submissions
P130026S036
Distribution
US Nationwide distribution.

Product description

Abbott TactiCath Sensor Enabled, Contact Force Ablation Catheter, 8F 115cm FJ, REF A-TCSE-FJ. For mapping of the heart chambers during ablation.

Code information (models, lots, serial numbers)

UDI 01)05415067027665(17)230831(10)xxxxxxx, Batch Numbers 8409416, 8409675, 8410811

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates