YUKON OCT Spinal System recallZ-1307-2022 · K2M, Inc. · initiated Jun 13, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from K2M, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or inflammatory response; adverse local tissue reactions may occur related to cytotoxicity with the nonconforming rod and surrounding tissue.

FDA root cause category: Under Investigation by firm

What the company is doing

On June 13, 2022, the firm distributed Medical Device Recall letters to affected customers. Customers were informed that local Stryker Sales Representative or Stryker Branch or Agency has already removed any product still located in customer facilities at the time of the initiation of the removal. If you have any questions or concerns, please contact Regulatory Compliance at 201.749.8090.

Recall details

Recall number
Z-1307-2022
Status
Open, Classified
Initiated by the company
Jun 13, 2022
Recalling firm
K2M, Inc.
Quantity
53 devices (19 US; 34 OUS)
Product code
KWP
FDA submissions
K171444
Distribution
Domestic distribution to FL GA NE NV NY OH OK SC Foreign distribution to Canada, EU, and Hong Kong.

Product description

Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal segments.

Code information (models, lots, serial numbers)

UDI-DI: 10888857326804 Lot PCMW

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates