Skip to content

YUKON OCT Spinal SystemRecalls & reported problems across all models

VB Spine LLC · 54 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names at least one YUKON OCT Spinal System model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
238
Deaths reported
0
Injuries reported
99
Recalls
3

We found 54 models of YUKON OCT Spinal System by VB Spine LLC in the FDA's device registry (GUDID). Together they are named in 238 adverse event reports in the FDA's MAUDE database, received between Jul 17, 2019 and Apr 23, 2026 — classified as 99 injuries, 139 malfunctions.

The most frequently reported specific device problem is “Fracture” (125 reports).

3 FDA recall records name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 8 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 9 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 3 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 4 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 4 reportsNov 2025: 9 reportsDec 2025: 8 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

YUKON OCT Spinal System recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 13, 2022Class IIOpen, Classified

    Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or inflammatory response; adverse local tissue reactions may occur related to cytotoxicity with the nonconforming rod and surrounding tissue.

    Z-1307-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 14, 2020Class IITerminated

    The Yukon polyaxial screw used in the Yukon Spinal System was incorrectly laser marked with size 5.0 mm (diameter) & 24 mm (length).

    Z-0931-2021 · Root cause: Employee error · Recall details · FDA record ↗

  3. Nov 3, 2020Class IITerminated

    Shafts do not pass through the inner cannula of the Anti-Torque Alignment Tube from Lot KTGG due to a manufacturing nonconformance. This could result in a delay of surgery or result in surgical intervention later on.

    Z-0520-2021 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported YUKON OCT Spinal System problems

Device problems

  • Fracture125
  • Migration69
  • Insufficient Device Problem Information14
  • Mechanical Jam13
  • Device Dislodged or Dislocated8
  • Nonstandard Device8
  • Break6
  • Migration or Expulsion of Device6

Patient problems

  • No Clinical Signs, Symptoms or Conditions145
  • No Consequences Or Impact To Patient34
  • Injury24
  • Insufficient Information24
  • Pain18
  • Neck Pain8
  • No Code Available5
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4

YUKON OCT Spinal System models

54 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent YUKON OCT Spinal System adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other VB Spine LLC products

All VB Spine LLC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page