Alinity recallZ-1455-2022 · Abbott Laboratories · initiated Jun 9, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
FDA root cause category: Under Investigation by firm
What the company is doing
On June 9, 2022, Abbott issued a "Urgent Medical Device Correction" to affected consignees via FedEx. 1. Immediately install the updated assay file versions listed below (as applicable for your laboratory), and manually configure the Low-Linearity value. The assay files can be obtained from www.corelaboratory.abbott 2. For the whole blood application, manually configure the THbWB Low-Linearity to 53.8278. For the hemolysate application, manually configure the THbH Low-Linearity to 1250. For detailed information on editing assay parameters, refer to Changing assay configuration settings, Change a linearity range in the ARCHITECT System Operations Manual, Section 2. 3. Please review this letter with your Medical Director or Laboratory Management and follow your laboratory protocol regarding the need for reviewing previously reported patient results. 4. Complete and return the Customer Reply Form. 5. Please retain this letter for your laboratory records. 6. If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter.
Recall details
- Recall number
- Z-1455-2022
- Status
- Open, Classified
- Initiated by the company
- Jun 9, 2022
- Recalling firm
- Abbott Laboratories
- Quantity
- 7,391 units
- Product code
- PDJ
- FDA submissions
- K130255
- Distribution
- US: AK AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS MT NC ND NE NJ NV NY OH OK OR PA PR SC SD TN TX UT VA WA WI WV OUS: ALBANIA ALGERIA ANDORRA ANGOLA ARGENTINA ARMENIA AUSTRALIA AUSTRIA AZERBAIJAN BAHAMAS BANGLADESH BELARUS BELGIUM BERMUDA BOSNIA AND HERZEGOVI BRAZIL BRUNEI BULGARIA BURKINA FASO CANADA CHILE CHINA COLOMBIA COSTA RICA Country CROATIA CYPRUS CZECH REPUBLIC ECUADOR EGYPT EL SALVADOR ESTONIA ETHIOPIA FAROE ISLANDS FINLAND FRANCE GABON GERMANY GREECE GUATEMALA HONG KONG HUNGARY INDIA INDONESIA IRAQ IRELAND ISRAEL ITALY JAMAICA JORDAN KAZAKHSTAN KENYA KUWAIT KYRGYZSTAN LEBANON LIBYA LITHUANIA MACEDONIA MALAYSIA MALDIVES MALI MEXICO MOLDOVA MOROCCO MOZAMBIQUE NAMIBIA NETHERLANDS NIGERIA NORWAY OMAN PAKISTAN PALESTINIAN TERRITOR PANAMA PARAGUAY PERU PHILIPPINES POLAND PORTUGAL QATAR REPUBLIC OF KOSOVO REPUBLIC OF YEMEN ROMANIA RUSSIA RWANDA SAUDI ARABIA SENEGAL SERBIA SINGAPORE SLOVAKIA SLOVENIA SOUTH AFRICA SPAIN ST. LUCIA ST. VINCENT SWEDEN SWITZERLAND TANZANIA THAILAND THE GAMBIA TRINIDAD & TOBAGO TUNISIA TURKEY TURKS & CAICOS UAE UGANDA UNITED KINGDOM URUGUAY UZBEKISTAN VIETNAM ZIMBABWE
Product description
REF: 08P4320, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL
Code information (models, lots, serial numbers)
UDI/DI: 00380740135591/ Lots: 59561UQ04, 59779UQ05, 60229UQ07, 60540UQ09, 61011UQ10, 61349UQ01, 61658UQ01
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Alinity 08P4320Abbott GmbH · Hemoglobin A1c test system1 recall7 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 90499).
- Z-1454-2022 · Open, ClassifiedREF: 4P52-20 Hemoglobin A1c for use with Architect, IVD, CE, R1 1 x 52 mL, R2 1 x 20 mL, A1cDL 2 x 35 mL
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page