Xhibit Telemetry Receiver, Model: 96280 recallZ-1513-2022 · Spacelabs Healthcare, Inc. · initiated May 31, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spacelabs Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.
FDA root cause category: Software Design Change
What the company is doing
On 05/31/22, correction notices were mailed to customers who were asked to take the following actions that can be taken to minimize risk to patients or users: Weigh the benefits versus the risks when deciding whether to continue to use the alarm escalation feature within XTR 1.4. Suggestions from Spacelabs to mitigate the risk: - Set the priority to the same level as the escalation max on the product. This will ensure that the highest priority alarm is triggered for the technical alarm. - Do not run the XTR 1.4 for more than 24 days consecutively. Restarting the XTR after 24 days will mitigate the issue for another 24 days. - The recalling firm is working on a 1.4.1 release of XTR which will address this defect. For additional information or technical assistance, please contact: Technical Support, 1-800-522-7025 and select 2
Recall details
- Recall number
- Z-1513-2022
- Status
- Open, Classified
- Initiated by the company
- May 31, 2022
- Recalling firm
- Spacelabs Healthcare, Inc.
- Quantity
- 529
- Product code
- MHX
- FDA submissions
- K141156
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of NY, AR, CO, FL, NJ, AL, KY, TN, SC, WA, OK, MN, PA, GA, VA, CA, OH, IA, KS, MI, IN, TX, MT, MO, ME, LA, IL, SD, MA, OR, ID, MD, AZ, AK and the countries of MB, ON, FR, QC, BC, IS, CT, RM.
Product description
Xhibit Telemetry Receiver, Model: 96280
Code information (models, lots, serial numbers)
Show the full code information (7k characters)
UDI-DI: 10841522100246, Software version 1.4.0. Serial Numbers: 6280-001538, 6280-000590, 6280-000589, 6280-000339, 9280-000711, 6280-001103, 6280-001102, 6280-001469, 6280-001467, 6280-001686, 6280-001399, 6280-001323, 6280-000457, 6280-000456, 6280-000455, 6280-000440, 6280-000439, 6280-000438, 6280-000429, 6280-000476, 6280-000475, 6280-001693, 6280-000703, 6280-000702, 6280-000701, 6280-000700, 6280-000314, 6280-000258, 6280-000157, 6280-001223, 6280-001222, 6280-001221, 6280-001220, 6280-001219, 6280-001218, 6280-001217, 6280-001216, 6280-001215, 6280-001214, 6280-001207, 6280-001127, 6280-001125, 6280-001124, 6280-001123, 6280-001056, 6280-001055, 6280-001054, 6280-001053, 6280-001052, 6280-001051, 6280-001050, 6280-001288, 6280-001287, 6280-001490, 6280-001692, 6280-000588, 6280-000517, 6280-000516, 6280-000515, 6280-001668, 6280-000206, 6280-000162, 6280-000210, 6280-000663, 6280-000377, 6280-001589, 6280-001385, 6280-001384, 6280-001382, 6280-001405, 6280-000594, 6280-000593, 6280-000592, 6280-000596, 6280-000274, 6280-001844, 6280-001843, 6280-001721, 6280-001720, 6280-000136, 6280-001784, 6280-001783, 6280-000359, 6280-000340, 6280-000336, 6280-001718, 6280-001706, 6280-001705, 6280-001704, 6280-001703, 6280-001702, 6280-001699, 6280-001698, 6280-001672, 6280-001669, 6280-000622, 6280-001797, 6280-000097, 6280-001558, 6280-001551, 6280-001649, 6280-001648, 6280-001647, 6280-001646, 6280-001645, 6280-001594, 6280-000093, 6280-001168, 6280-001531, 6280-001670, 6280-000219, 6280-000550, 6280-000878, 6280-000877, 6280-000872, 6280-000862, 6280-000858, 6280-000857, 6280-000221, 6280-000089, 6280-000204, 6280-000212, 6280-000510, 6280-000509, 6280-000209, 6280-000320, 6280-000538, 6280-000528, 6280-000205, 6280-000361, 6280-000360, 6280-000223, 6280-000963, 6280-000529, 6280-001575, 6280-001574, 6280-001573, 6280-001572, 6280-001571, 6280-001570, 6280-000467, 6280-000466, 6280-000465, 6280-000464, 6280-000463, 6280-000462, 6280-000461, 6280-000460, 6280-000459, 6280-000458, 6280-000222, 6280-000264, 6280-000275, 6280-000354, 6280-001677, 6280-001048, 6280-001043, 6280-001041, 6280-001038, 6280-001037, 6280-001036, 6280-001023, 6280-001022, 6280-001016, 6280-000787, 6280-001418, 6280-001419, 6280-001416, 6280-000894, 6280-000866, 6280-000865, 6280-000863, 6280-000902, 6280-000898, 6280-000724, 6280-000420, 6280-000419, 6280-000798, 6280-000325, 6280-000532, 6280-000521, 6280-000213, 6280-000254, 6280-000856, 6280-000855, 6280-000854, 6280-000853, 6280-000852, 6280-000851, 6280-001143, 6280-001142, 6280-001141, 6280-000472, 6280-000375, 6280-000368, 6280-000421, 6280-000519, 6280-001213, 6280-000836, 6280-000835, 6280-000834, 6280-000833, 6280-000832, 6280-000831, 6280-000586, 6280-001473, 6280-000319, 6280-000263, 6280-000312, 6280-000579, 6280-000092, 6280-000090, 6280-000084, 6280-000083, 6280-000079, 6280-001380, 6280-001466, 6280-001458, 6280-001611, 6280-001592, 6280-001591, 6280-000388, 6280-000161, 6280-000160, 6280-000159, 6280-000158, 6280-000066, 6280-000053, 6280-000174, 6280-000173, 6280-000172, 6280-000170, 6280-000168, 6280-000167, 6280-000077, 6280-000598, 6280-000355, 6280-000344, 6280-000341, 6280-000338, 6280-000468, 6280-000423, 6280-000478, 6280-000477, 6280-000454, 6280-000428, 6280-000261, 6280-001614, 6280-001642, 6280-001641, 6280-001640, 6280-001639, 6280-001638, 6280-001637, 6280-001636, 6280-001635, 6280-001634, 6280-001633, 6280-001632, 6280-001631, 6280-001630, 6280-001629, 6280-001913, 6280-000138, 6280-000437, 6280-000436, 6280-000177, 6280-000176, 6280-000337, 6280-001857, 6280-001853, 6280-001749, 6280-001359, 6280-001160, 6280-001159, 6280-001158, 6280-001157, 6280-001156, 6280-001154, 6280-001132, 6280-001112, 6280-001109, 6280-000991, 6280-000983, 6280-001189, 6280-001188, 6280-001205, 6280-001204, 6280-001203, 6280-001202, 6280-001200, 6280-001194, 6280-001193, 6280-001184, 6280-001166, 6280-000358, 6280-000357, 6280-000345, 6280-000432, 6280-000431, 6280-000811, 6280-000693, 6280-000690, 6280-000929, 6280-001063, 6280-001062, 6280-001061, 6280-001060, 6280-000391, 6280-001273, 6280-001554, 6280-001553, 6280-001261, 6280-000426, 6280-001525, 6280-000229, 6280-000876, 6280-000875, 6280-000745, 6280-000208, 6280-000574, 6280-000218, 6280-001087, 6280-000045, 6280-000041, 6280-000374, 6280-000373, 6280-001257, 6280-001256, 6280-001255, 6280-001254, 6280-001253, 6280-001252, 6280-001251, 6280-001250, 6280-001249, 6280-000840, 6280-000486, 6280-001825, 6280-000543, 6280-000542, 6280-000839, 6280-000879, 6280-001325, 6280-001071, 6280-001070, 6280-001069, 6280-000900, 6280-000897, 6280-001822, 6280-001874, 6280-001872, 6280-001871, 6280-000695, 6280-000540, 6280-000541, 6280-000539, 6280-000166, 6280-000624, 6280-000614, 6280-000613, 6280-000612, 6280-000611, 6280-000610, 6280-000609, 6280-000608, 6280-000607, 6280-000630, 6280-000629, 6280-000628, 6280-000627, 6280-000626, 6280-001644, 6280-001619, 6280-001617, 6280-000587, 6280-000490, 6280-000489, 6280-000488, 6280-000487, 6280-000974, 6280-000973, 6280-000972, 6280-000971, 6280-000970, 6280-000969, 6280-000968, 6280-000967, 6280-000966, 6280-000965, 6280-000329, 6280-000326, 6280-000318, 6280-001621, 6280-001609, 6280-001596, 6280-001536, 6280-001381, 6280-001340, 6280-001339, 6280-001338, 6280-001337, 6280-001336, 6280-001335, 6280-001334, 6280-001315, 6280-001312, 6280-001311, 6280-000547, 6280-000546, 6280-000545, 6280-000544, 6280-000518, 6280-001764, 6280-001752, 6280-001751, 6280-001746, 6280-001548, 6280-001408, 6280-001431, 6280-001729, 6280-001046, 6280-000601, 6280-001350, 6280-000689, 6280-001487, 6280-001475, 6280-000928, 6280-000927, 6280-000926, 6280-000925, 6280-000901, 6280-001329, 6280-001331, 6280-001330, 6280-001342, 6280-001341, 6280-000514, 6280-000513, 6280-000512, 6280-000511, 6280-000530, 6280-000108, 6280-000107, 6280-000106, 6280-000105, 6280-000104, 6280-000103, 6280-000102, 6280-000101, 6280-000100, 6280-000600, 6280-000276, 6280-001374, 6280-000195, 6280-000192, 6280-000189, 6280-000187, 6280-000178, 6280-000194, 6280-001820, 6280-001805, 6280-001444, 6280-001225, 6280-000850, 6280-000725, 6280-000694, 6280-000553, 6280-000549, 6280-000548, 6280-000330, 6280-000317, 6280-000188, 6280-000182, 6280-000290, 6280-001818, 6280-001765, 6280-001602, 6280-001601, 6280-001450, 6280-001449, 6280-001377, 6280-001376, 6280-001178, 6280-001049, 6280-000796, 6280-000743, 6280-000491, 6280-000356, 6280-000295, 6280-000293, 6280-000292, 6280-000291, 6280-000231, 6280-000196, 6280-000061, 6280-000060, 6280-000473, 6280-000232, 6280-000062, 6280-001319, 6280-001313, 6280-000384, 6280-000199, 6280-000198, 6280-000065, 6280-000064, 6280-000237, 6280-000406, 6280-000401, 6280-000362, 6280-000270, 6280-000400, 6280-001245, 6280-001101, 6280-000873, 6280-000871, 6280-000870, 6280-000861, 6280-000860, 6280-000815, 6280-000554, 6280-000527, 6280-000499, 6280-000416, 6280-000371, 6280-000197, 6280-000523, 6280-000063
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Xhibit Telemetry ReceiverSpacelabs Healthcare (Washington), Inc · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)3 recalls30 reports
Other devices under the same FDA submission
1 device cleared under K141156 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- XTR 96280Spacelabs Healthcare (Washington), Inc · SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page