Xhibit Telemetry ReceiverRecalls & reported problems
Spacelabs Healthcare (Washington), Inc · ENHANCED TELEMETRY,96280 · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
The SpaceLabs Xhibit Telemetry Receiver keeps hospital beds' vital-sign data flowing to a central monitoring station. Most of the 28 reports describe patients dropping offline — receivers restarting or shutting down on their own, beds going offline in groups, or alarms not escalating when signals drop. No deaths or injuries were reported. Three open recalls from November 2024 cover these software restart and alarm issues, and reports have fallen from 21 to 6 in the last 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 10841522100246), received between Aug 8, 2024 and Sep 21, 2026 — classified as 30 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (16 reports).
3 FDA recall records name this device's identifier, 3 still open.
3 other recalls cover devices cleared under the same FDA submission (K141156); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 25, 2024Open, Classified
Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.
Z-0810-2025 · Root cause: Software change control · FDA record ↗
- Nov 25, 2024Open, Classified
Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
Z-0809-2025 · Root cause: Labeling design · FDA record ↗
- May 31, 2022Open, Classified
Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.
Z-1513-2022 · Root cause: Software Design Change · FDA record ↗
Most reported problems
Device problems
- Communication or Transmission Problem16
- Unexpected Shutdown4
- Application Program Freezes or Fails to Launch2
- Complete Loss of Power2
- Device Alarm System2
- Unintended Application Program Shut Down2
- Inappropriate or Unexpected Reset1
- Incomplete or Inadequate Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions30
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026No Display/Image; Incomplete or Inadequate Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50847374 ↗MalfunctionSep 4, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50623670 ↗MalfunctionJun 15, 2026Unexpected Shutdown
THE CUSTOMER CONTACTED SPACELABS TECHNICAL SUPPORT TO REPORT THAT FIVE TELEMETRY CHANNELS WENT OFFLINE WHILE BEING MONITORED ON AN XHIBIT CENTRAL STATION (XCS). THERE WAS NO REPORT OF PATIENT OR USER HARM ASSOCIATED WITH THE EVENT. THE BIOMED STATED THAT THEY RETUNED ALL THE AFFECTED CHANNELS TO RESTORE MONITORING. THE TECHNICAL SUPPORT REPRESENTATIVE PULLED XHIBIT TELEMETRY RECEIVER (XTR) LOGS. THE LOGS WERE EVALUATED BY A PRODUCT SUPPORT SPECIALIST (PSS). THE PSS FOUND THAT THE HAD BEEN ASSIGNED TO AN XTR THAT WAS NO OFFLINE. THE PSS ADVISED THE CUSTOMER BIOMED TO RESET THE XTR. AFTER THE XTR RESET, THE DEVICE WAS NOW COMMUNICATING ON THE NETWORK. THE PSS REVIEWED LOGS AND FOUND THAT THE XTR HAD CRASHED AFTER JUST OVER SIX MONTHS OF RUN TIME. THE PSS ADVISED THE CUSTOMER OF THE MAINTANCE SCHEDULE, STATING THE XTRS SHOULD BE RESTARTED AT LEAST ONCE EVERY 200 DAYS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25512784 ↗MalfunctionJun 4, 2026Communication or Transmission Problem
THE CUSTOMER CONTACTED SPACELABS HEALTHCARE TECHNICAL SUPPORT TO REPORT THAT AFTER A BRIEF POWER OUTAGE TELEMETRY MONITORING DID NOT RETURN AS EXPECTED. THERE WAS NO REPORT OF PATIENT OR USER HARM ASSOCIATED WITH THE EVENT. TECHNICAL SUPPORT REMOTELY CONNECTED AND RESTARTED THE AFFECTED XHIBIT TELEMETRY RECIEVER (XTR) AND CONFIRMED THAT THE ISSUE WAS RESOLVED. CAUSE OF FAILURE WAS NOT DETERMINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25401841 ↗MalfunctionApr 30, 2026Communication or Transmission Problem
THE CUSTOMER CONTACTED SPACELABS HEALTHCARE TECHNICAL SUPPORT TO REPORT THAT THEY LOST SIXTEEN TELEMETRY PATIENTS ACROSS ALL CENTRAL STATIONS. THERE WAS NO REPORT OF PATIENT OR USER HARM ASSOCIATED WITH THE EVENT. A SPACELABS HEALTHCARE TECHNICAL SUPPORT REPRESENTATIVE REMOTELY CONNECTED AND VERIFIED THAT THE TELEMETRY CHANNELS WERE NO LONGER ONLINE. TECHNICAL SUPPORT ADVISED THE CUSTOMER THAT ONE OF THE XHIBIT TELEMETRY RECEIVERS (XTR) APPEAR TO HAVE GONE OFFLINE. TO RESUME MONITORING, TECHNICAL SUPPORT RECOMMENDED THE CUSTOMER READMIT THE PATIENTS. THE CUSTOMER WAS ADVISED TO HAVE A BIOMED INSPECT THE XTR. SEVERAL FOLLOW UP ATTEMPTS WERE MADE FOR ADDITIONAL INFORMATION, HOWEVER NO RESPONSE HAS BEEN RECEIVED AT THIS TIME. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25049237 ↗MalfunctionFeb 4, 2026Communication or Transmission Problem
THE CUSTOMER CONTACTED SPACELABS HEALTHCARE TO REPORT THAT SEVERAL TELEMETRY PATIENTS WERE UNABLE TO BE MONITORED ON A XHIBIT CENTRAL STATION. NO PATIENT HARM WAS REPORTED. THE ISSUE RESOLVED WITHOUT INTERVENTION. THE CAUSE OF THE ISSUE COULD NOT BE DETERMINED. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24264934 ↗MalfunctionJan 15, 2026Application Program Freezes or Fails to Launch
THE CUSTOMER CONTACTED SPACELABS HEALTHCARE TECHNICAL SUPPORT TO REPORT THAT THEIR XHIBIT TELEMETRY RECEIVER (XTR) RESTARTED AND WAS OFFLINE FOR ABOUT FIVE MINUTES BEFORE RETURNING ON ITS OWN. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. A SPACELABS PRODUCT SUPPORT SPECIALIST (PSS) REMOTELY CONNECTED AND GATHERED XTR LOGS. PER THE LOG REVIEW, THE XTR WENT THROUGH A SERVICE RESTART. FOLLOWING THE RESTART A QUAD RECEIVER CARD (QRC) FAILED TO INITIALIZE PROPERLY, CAUSING A DELAY IN THE WAVEFORMS TO RETURN TO THE CENTRAL STATION. CAUSE OF FAILURE COULD NOT BE DETERMINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24090448 ↗MalfunctionDec 19, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT ALL TELEMETRY BEDS WENT OFFLINE AT THE XHIBIT CENTRAL STATION. SPACELABS HEALTHCARE TECHNICAL SUPPORT REMOTELY CONNECTED AND CONFIRMED THAT THE TELEMETRY RECEIVERS WERE NOT SENDING ANY DATA. THE ISSUE WAS ESCALATED TO THE CUSTOMER'S INTERNAL IT TEAM. IN A FOLLOW UP CALL WITH THE CUSTOMER, IT WAS CONFIRMED THAT IT WAS ABLE TO RESOLVE THE FAILURE BY RESTARTING THE XHIBIT TELEMETRY RECEIVER. WHILE CAUSE OF FAILURE COULD NOT BE ESTABLISHED, THE CUSTOMER CONFIRMED NO FURTHER ISSUES FOLLOWING THE RESTART.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23854470 ↗MalfunctionOct 12, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT WHILE MONITORING TELEMETRY PATIENTS AT TWO XHIBIT CENTRAL STATIONS (XCS), MULTIPLE TELEMETRY PATIENTS WENT OFFLINE. A SPACELABS HEALTHCARE PRODUCT SUPPORT SPECIALIST (PSS) PERFORMED TROUBLESHOOTING AND A LOG REVIEW AND IDENTIFIED THAT THE XHIBIT TELEMETRY RECEIVER (XTR) FARM WAS EXCEEDING CONNECTIONS. THE PSS INFORMED THE CUSTOMER THAT THEY WOULD NEED TO EVALUATE THEIR CONNECTIONS TO REDUCE ANY FURTHER IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23273769 ↗MalfunctionOct 12, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down
THE CUSTOMER REPORTED THAT SEVERAL TELEMETRY BEDS WENT OFFLINE FOR A BRIEF PERIOD OF TIME UNEXPECTANTLY. A PRODUCT SUPPORT SPECIALIST (PSS) GATHERED DEVICE LOGS AND WAS ABLE TO CONFIRM THE PATIENTS HAD GONE OFFLINE. THE LOGS INDICATED THAT THE XHIBIT TELEMETRY RECEIVER (XTR) HAD RESTARTED, BUT THERE WAS NOT ENOUGH INFORMATION IN THE LOGS TO INDICATE THAT THE RESTART WAS UNPROMPTED. THE PSS CONTACTED THE CUSTOMER TO CONFIRM IF THEY HAD RESTARTED THE RECEIVER DURING THE TIME THE PATIENTS WERE OFFLINE, BUT NO RESPONSE HAS BEEN RECEIVED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW UP MDR WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23273765 ↗MalfunctionSep 4, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down
THE CUSTOMER REPORTED THAT A XHIBIT TELEMETRY RECEIVER (XTR) HAD UNEXPECTEDLY RESTARTED RESULTING IN A LOSS OF TELEMETRY MONITORING. A SPACELABS HEALTHCARE PRODUCT SUPPORT SPECIALIST (PSS) REVIEWED XTR LOGS AND CONFIRMED THE CRASH, HOWEVER, CAUSE OF THE CRASH WAS NOT DETERMINED. THE PSS IS CURRENTLY WORKING WITH THE CUSTOMER TO GATHER ADDITIONAL DATA TO DETERMINE CAUSE OF FAILURE. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22974508 ↗MalfunctionAug 29, 2025Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER REPORTED THAT WHILE MONITORING TELEMETRY PATIENTS, SEVEN BEDS WENT OFFLINE UNEXPECTANTLY. A SPACELABS HEALTHCARE PRODUCT SUPPORT SPECIALIST GATHERED THE EVENT DATA AND DEVICE LOGS TO INVESTIGATE THE ISSUE. IT WAS FOUND THAT THE SEVEN PATIENTS WERE ADMITTED TO THREE DIFFERENT XHIBIT TELEMETRY RECEIVERS (XTR). A REVIEW OF THE DATA SHOWED THAT THE PATIENTS ADMITTED TO TWO OF THE THREE XTRS DID NOT LOSE MONITORING AT ANYTIME DURING THE EVENT. HOWEVER IT WAS FOUND THAT THE PATIENTS ADMITTED TO THE THIRD XTR DID LOSE MONITORING FOR APPROXIMATELY 30 SECONDS BEFORE RETURNING ON THEIR OWN. FURTHER REVIEW OF THE DATA SHOWED THAT THE XTR HAD RESTARTED UNEXPECTANTLY, WHICH RESULTED IN THE GAP IN MONITORING. AS A PRECAUTION, THE DEVICE WAS REMOVED FROM SERVICE AND THE CUSTOMER WILL RETURN IT TO SPACELABS HEALTHCARE FOR ADDITIONAL INVESTIGATION. AT THIS TIME, THE DEVICE HAS NOT BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22929520 ↗MalfunctionAug 27, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT DURING A STORM, TELEMETRY PATIENTS STOPPED DISPLAYING WAVEFORM. INITIALLY, IT WAS REPORTED THAT TELEMETRY WAS SHOWING SQUELCH, HOWEVER WHEN TECH SUPPORT REMOTELY CONNECTED ALL EXHIBIT TELEMETRY RECEIVERS (XTR) WENT OFFLINE. A FIELD SERVICE ENGINEER (FSE) CONTACTED THE CUSTOMER AND ADVISED THE CUSTOMER RESTART ALL XTRS, WHICH DID NOT RESOLVE THE OUTAGE. THE TELEMETRY PATIENTS WERE THEN DISCHARGED AND READMITTED, WHICH ULTIMATELY RESOLVED THE ISSUE. CAUSE OF FAILURE WAS NOT IDENTIFIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22909687 ↗MalfunctionAug 1, 2025Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
THE CUSTOMER REPORTED FREQUENT TELEMETRY SYSTEM ISSUES WHERE PATIENT DATA WOULD STOP DISPLAYING ON A CENTRAL STATION UNTIL THE XHIBIT TELEMETRY RECEIVER (XTR) WAS RESTARTED. A SPACELABS HEALTHCARE FIELD SERVICE ENGINEER WENT ON SITE AND COULD NOT VERIFY THE REPORTED ISSUE AND COULD NOT FIND ANY FAULTS WITH ANY XTR. XTR LOGS WERE PULLED AND PROVIDED TO A PRODUCT SUPPORT SPECALIST (PSS). THE PSS REVIEWED THE LOGS AND FOUND THAT THE BIOS SETTINGS WERE INCORRECT ALL FIVE OF THE XTRS. THE FSE ATTEMPTED TO ADDRESS THE BIOS ISSUE VIA SOFTWARE RELOAD, HOWEVER THE FIRST XTR WOULD NOT LOAD THE SOFTWARE. THE FSE STATED THAT HE WOULD ORDER PARTS TO REPLACE HARDWARE. A FOLLOW UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22692960 ↗MalfunctionAug 1, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT WHILE MONITORING TELEMETRY PATIENTS, TWO BEDS WENT OFFLINE. THE USER MOVED THE PATIENTS TO NEW TRANSMITTERS AND THE BIOMED REPORTED THAT THE ISSUE WAS RESOLVED. THE CUSTOMER CALLED BACK ON 3/JULY STATING THE ISSUE REAPPEARED WITH TWO DIFFERENT TRANSMITTERS. THE PATIENTS WERE ONCE AGAIN MOVED TO NEW TRANSMITTERS TO RESOLVE THE FAILURE. THE BIOMED STATED THAT HE HAD TESTED THE TRANSMITTERS AND COULD NOT FIND ANY ISSUES. SPACELABS HEALTHCARE REMOTELY CONNECTED AND FOUND FOUR RECEIVER SLOTS WERE DISABLED ON A EXHIBIT TELEMETRY RECEIVER (XTR) QUAD RECEIVER CARD (QRC). A FIELD SERVICE ENGINEER (FSE) WAS PAGED TO GO ON SITE, NO BEDS WERE FOUND OFFLINE. ALL XTRS WERE RESTARTED WHILE THE FSE WAS ON SITE AS PART OF PREVENTIVE MAINTENANCE. CAUSE OF FAILURE WAS NOT IDENTIFIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22692957 ↗MalfunctionJul 11, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT AFTER A NETWORK DROP AT THE FACILITY, FOUR PATIENTS HAD NOT RESUMED MONITORING ON THEIR OWN. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. A SPACELABS HEALTHCARE TECHNICAL SUPPORT REPRESENTATIVE CONFIRMED THAT ALL FOUR PATIENTS WERE ON ONE XHIBIT TELEMETRY RECEIVER (XTR). IT WAS CONFIRMED THAT THE XTR WAS RECEIVING THE TELEMETRY DATA, HOWEVER, IT APPEARED THE DATA WAS NOT BEING TRANSMITTED TO THE XHIBIT CENTRAL STATION OR ICS. THE TECHNICAL SUPPORT REPRESENTATIVE MOVED ALL AFFECTED CHANNELS TO ANOTHER SYSTEM AND RESTARTED THE XTR TO RESTORE NETWORK COMMUNICATION. THE CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22487798 ↗MalfunctionJul 10, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED A FEW TELEMETRY BOXES WOULD DROP OFF THE XHIBIT CENTRAL STATIONS (XCS) INTERMITTENTLY. THE TELEMETRY PATIENTS WOULD RETURN ON THEIR OWN, HOWEVER CUSTOMER SETTINGS AND NOTES WERE SET TO DEFAULT. A FIELD SERVICE ENGINEER (FSE) WENT ON SITE AND CONFIRMED THE SYSTEM WAS OPERATIONAL. AN ERROR MESSAGE WAS SEEN IN THE XCS MESSAGE BAR INDICATES THE XHIBIT TELEMETRY RECEIVER (XTR) HAD DISCONNECTED AND RECONNECTED. AS A PRECAUTION THE FSE GATHERED DEVICE LOGS AND FORWARDED THEM TO THE PRODUCT SUPPORT SPECIALIST (PSS) FOR ADDITIONAL INVESTIGATION. APSS REVIEWED THE XCS LOGS AND CONFIRMED THAT A XTR HAD DISCONNECTED ON 9/JUN/2025 FOR ROUGHLY 30 SECONDS. XTR LOGS WERE UNAVAILABLE FOR REVIEW. A PSS MADE SEVERAL ATTEMPTS TO REACH OUT TO THE CUSTOMER TO GATHER ADDITIONAL DATA REGARDING THE LOSS OF TELEMETRY MONITORING AND ACCESS TO THE XTR LOGS, HOWEVER, THE CUSTOMER HAS NOT RESPONDED. SHOULD ADDITIONAL INFORMATION BE MADE AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22476005 ↗MalfunctionJun 30, 2025Device Alarm System; Device Alarm System
THE CUSTOMER REPORTED THAT THEY HAD A FAILURE TO ALARM FOR A VENTRICULAR TACHYCARDIA. DUE TO THE DELAY IN THE CUSTOMER REPORTING THE EVENT TO SPACELABS HEALTHCARE, WAVEFORM DATA WAS NO LONGER ABLE TO BE REVIEWED AS THE DATA HAD BEEN PURGED. ADDITIONALLY, IT WAS UNCLEAR WHICH XHIBIT TELEMETRY RECEIVER (XTR) WAS INVOLVED WITH THE EVENT. A SPACELABS HEALTHCARE FIELD SERVICE ENGINEER (FSE) WENT ON SITE TO EVALUATE ALL THE XTRS AT THE CUSTOMERS SITE THAT COULD HAVE BEEN ASSOCIATED WITH THE EVENT. ALL OF THE XTRS PASSED FUNCTIONAL AND PERFORMANCE TESTING. THE FSE GATHERED XTR LOGS AND FORWARDED TO A SPACELABS HEALTHCARE PRODUCT SUPPORT SPECIALIST FOR ADDITIONAL REVIEW. WHEN THE REVIEW OF DATA IS COMPLETE, A FOLLOW UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22352731 ↗MalfunctionJun 20, 2025Complete Loss of Power; Complete Loss of Power
THE CUSTOMER REPORTED THAT WHILE MONITORING TELEMETRY PATIENTS ON A XHIBIT CENTRAL STATION (XCS), ALL PATIENTS WENT OFFLINE. SPACELABS HEALTHCARE TECHNICAL SUPPORT WALKED THE USER THROUGH RESTARTING THE XCS AND THE CUSTOMER STATED THAT BEDSIDE PATIENTS RETURNED, HOWEVER, TELEMETRY PATIENTS CONTINUED TO SHOW OFFLINE. THE CUSTOMER LATER IDENTIFIED THAT THE XHIBIT TELEMETRY RECEIVER (XTR) WAS FOUND TO BE OFF. ONCE THE DEVICE WAS POWERED BACK ON, THE MISSING TELEMETRY PATIENTS REAPPEARED AT THE XCS. LOGS WERE COLLECTED FROM TWO DIFFERENT XTRS. AT THIS TIME, THE LOGS HAVE NOT BEEN REVIEWED TO DETERMINE CAUSE OF FAILURE. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22294659 ↗MalfunctionMay 30, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT THEY EXPERIENCED POWER DISRUPTION AT THEIR FACILITY. WHEN THE POWER WAS RESTORED, MOST PATIENT MONITORING RESUMED, HOWEVER 18 TELEMETRY PATIENTS WERE SHOWING OFFLINE. A SPACELABS HEALTHCARE FIELD SERVICE ENGINEER WORKED WITH THE CUSTOMER TO PERFORM TROUBLESHOOTING OF THE XHIBIT TELEMETRY RECEIVERS (XTR). TWO DIFFERENT FAULTY XTRS WERE IDENTIFIED, ALL PATIENTS ASSOCIATED WITH THE TWO XTRS WERE SHOWING OFFLINE AT THE CENTRAL STATION WHILE SHOWING WAVEFORMS AT THE RECEIVER. THE PATIENTS WERE TRANSFERRED TO DIFFERENT XTRS AND THE FAULTY XTRS WERE RESTARTED. FOLLOWING THE RESTART, IT WAS CONFIRMED THE NETWORK COMMUNICATION WAS RESTORED, CHANNELS WERE ABLE TO BE TUNED AND MONITORED AND NO FURTHER ISSUES WERE OBSERVED. THE LOGS FROM THE XTRS AND PROVIDED TO A SPACELABS HEALTHCARE PRODUCT SUPPORT SPECIALIST (PSS). THE PSS IDENTIFIED THAT THE TWO FAULTY XTRS LOST NETWORK COMMUNICATION RELATING TO THE POWER DISRUPTION AND RESTARTED, HOWEVER UPON RESTART, THE XTRS FAILED TO ESTABLISH CONNECTIONS DUE TO A FAILURE TO PROPERLY RESTART. ONCE THE XTRS WERE RESTARTED AGAIN, THE ERRORS PREVENTING THE COMMUNICATION WERE RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22117188 ↗MalfunctionApr 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT WHILE MONITORING TELEMETRY PATIENTS, SIX PATIENTS DROPPED OFF THE CENTRAL FOR A BRIEF PERIOD OF TIME BEFORE RETURNING ON ITS OWN. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21770407 ↗MalfunctionJan 31, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT MID-MONITORING SEVERAL TELEMETRY PATIENTS WENT OFFLINE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21284192 ↗MalfunctionJan 24, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT WHILE MONITORING PATIENTS ON A EXHIBIT CENTRAL STATION, THE CENTRAL LOST ALL TELEMETRY CHANNELS. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21230822 ↗MalfunctionJan 10, 2025Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED THAT WHILE MONITORING TELEMETRY PATIENTS THEY LOST NETWORK CONNECTIVITY TO THE XHIBIT TELEMETRY RECEIVER, RESUTING A LOSS OF MONITORING. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21126969 ↗MalfunctionJan 10, 2025Device Alarm System; Device Alarm System
THE CUSTOMER REPORTED THAT WHILE MONITORING TELEMETRY PATIENTS, ALARM MESSAGES FAILED TO ESCALATE TO HIGHER PRIORITY LEVELS. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21126961 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141156. One submission often covers many models, so check the product and lot details in each record.
- Z-0532-2018 · Jan 17, 2018 · TerminatedSpacelabs Healthcare Xhibit Telemetry Receiver, Model 96280, software version 1.1, physiological patient monitor (with arrhythmia detection or alarms) The Spacelabs Healthcare Tel
- Z-2886-2016 · Aug 25, 2016 · TerminatedSpacelabs Healthcare Xhibit Telemetry Receiver (XTR), Model 96280. Xhibit Telemetry Receivers (XTR) must be connected to an Xhibit Central Station for the display of patient vital
- Z-1502-2016 · Mar 28, 2016 · TerminatedThe Spacelabs Healthcare Xhibit Telemetry Receiver (XTR), Model 96280 is used to transmit patient(s) physiological measurements and data to a monitor.
Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,541 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,493 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,172 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths941 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page